Combined Guarded-Hot-Plate and Heat Flow Meter Method for absolute thermal conductivity tests excluding thermal contact resistance

被引:0
|
作者
Tleoubaev, A. [1 ]
Brzezinski, A. [1 ]
机构
[1] LaserComp Inc, Saugus, MA 01906 USA
关键词
D O I
暂无
中图分类号
O414.1 [热力学];
学科分类号
摘要
A Combined Guarded Hot Plate and Heat Flow Meter Method was developed and tested for absolute thermal conductivity tests of moderate thermal conductivity (up to similar to 10 W/mK) materials. A thin flat guarded heater of known area is placed between two flat-parallel samples of the same material and of different thicknesses. The stack is clamped between two isothermal plates each having a heat flow meter. Heat flux across each of the two samples is inversely proportional to its total thermal resistance - sum of sample's thermal resistance (thickness divided by thermal conductivity) and its two surface contact resistances, which are assumed to be equal for the two samples. After reaching thermal equilibrium the measured amount of electric power of the heater's central part, it's and plates' temperatures, samples' thicknesses and both heat flow meters' readings are used to calculate the material's absolute thermal conductivity excluding the thermal contact resistance. Measurements without taking into account the thermal contact resistance would cause very large errors (as much as hundreds percent in some cases). This combination of the two traditional steady-state methods provides significantly increased accuracy of the absolute thermal conductivity measurements of many very important materials such as ceramics, glasses, plastics, rocks, polymers, composites, fireproof materials, etc. Both theoretical aspects of the combined method and its experimental check using some reference materials (Pyrex, Pyroceram, Vespe (R) 1) are presented.
引用
收藏
页码:502 / 510
页数:9
相关论文
共 50 条
  • [1] Correct Use of the Guarded-Hot-Plate Method for Thermal Conductivity Measurements on Solids
    Hans-Peter Ebert
    Stephan Vidi
    International Journal of Thermophysics, 2024, 45
  • [2] Correct Use of the Guarded-Hot-Plate Method for Thermal Conductivity Measurements on Solids
    Ebert, Hans-Peter
    Vidi, Stephan
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2024, 45 (02)
  • [3] Verification of thermal conductivity measurements using guarded hot plate and heat flow meter methods
    Daeho Kim
    Sanghyun Lee
    Inseok Yang
    Journal of the Korean Physical Society, 2021, 78 : 1196 - 1202
  • [4] Verification of thermal conductivity measurements using guarded hot plate and heat flow meter methods
    Kim, Daeho
    Lee, Sanghyun
    Yang, Inseok
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2021, 78 (12) : 1196 - 1202
  • [5] Regression analysis estimation of thermal conductivity using guarded-hot-plate apparatus
    Sanjaya, Christian Suryono
    Wee, Tiong-Huan
    Tamilselvan, T.
    APPLIED THERMAL ENGINEERING, 2011, 31 (10) : 1566 - 1575
  • [6] MEASUREMENT OF THERMAL-DIFFUSIVITY WITH A GUARDED-HOT-PLATE DEVICE USING A DYNAMIC METHOD
    ZOBEL, W
    HETFLEISCH, J
    FRICKE, J
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (07) : 842 - 846
  • [7] Fast measurements of absolute thermal conductivity excluding thermal contact resistance errors
    Tleoubaev, A.
    Brzezinski, A.
    THERMAL CONDUCTIVITY 28: THERMAL EXPANSION 16, 2006, 28 : 339 - +
  • [8] Influence of Thermal Contact Resistance on Thermal Conductivity Measurement with a High-Temperature Guarded Hot Plate Apparatus
    Yao, Kai
    Liu, Yunchuan
    Zhou, Yanping
    Wang, Xuerong
    Wang, Qianqian
    Ma, Yandong
    Yao, Cui
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2020, 41 (02)
  • [9] Influence of Thermal Contact Resistance on Thermal Conductivity Measurement with a High-Temperature Guarded Hot Plate Apparatus
    Kai Yao
    Yunchuan Liu
    Yanping Zhou
    Xuerong Wang
    Qianqian Wang
    Yandong Ma
    Cui Yao
    International Journal of Thermophysics, 2020, 41
  • [10] CIRCULAR AND SQUARE EDGE EFFECT STUDY FOR GUARDED-HOT-PLATE AND HEAT-FLOW-METER APPARATUSES.
    Peavy, Brad
    Rennex, Brian
    Journal of thermal insulation, 1986, 9 : 254 - 300