Left ventricular free wall perforation by a right ventricular pacemaker lead: a case report

被引:7
|
作者
Satomi, Natsuko [1 ]
Enta, Kenji [1 ]
Otsuka, Masato [1 ]
Ishii, Yasuhiro [1 ]
Asano, Ryota [2 ]
Sawa, Shigeharu [2 ]
机构
[1] Ogikubo Hosp, Cardiovasc Ctr, Dept Cardiol, Suginami Ku, 3-1-24 Imagawa, Tokyo 1670035, Japan
[2] Ogikubo Hosp, Cardiovasc Ctr, Dept Cardiovasc Surg, Suginami Ku, 3-1-24 Imagawa, Tokyo 1670035, Japan
关键词
Case report; Pacemaker; Complication; Lead perforation; Left ventricular perforation; Lead extraction;
D O I
10.1093/ehjcr/ytab125
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Lead perforation is one of the major complications of pacemaker implantation, but cases of right ventricular (RV) lead perforation through the septum and left ventricle are rarely reported. We described a rare case of left ventricular (LV) free wall perforation by an RV lead and the management of this complication. Case summary An 84-year-old man was admitted with a dual-chamber pacemaker due to pacing failure caused by an RV lead fracture. New lead implantation was performed on the next day, but pacing failure occurred again on the second postoperative day (POD). We found the lead perforation on the fluoroscopy during temporary pacemaker insertion. Computed tomography scan and transthoracic echocardiogram showed that the added lead perforated through both the septum and LV free wall. A new lead was inserted on the fourth POD, and an off-pump open chest surgery for extraction of the penetrating lead was performed uneventfully on the 20th POD. Discussion We considered that some features of the lead (SelectSecure 3830-69, Medtronic) may be related to this complication, as the lead was very thin, had a non-retractable bare screw and was inserted with a dedicated delivery catheter. We have to be careful when performing implantation of this kind of lead to avoid such a rare complication.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Case report: Left ventricular perforation caused by right ventricular pacemaker lead
    Huang, Xiang
    Zhou, Heng
    Li, Xiao-Mei
    Li, Xiao-Lan
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 9
  • [2] Right ventricular perforation, pneumothorax, and a pneumatocele by a pacemaker lead: a case report
    Shihoko Iwata
    Ayana Hirose
    Ikue Furui
    Takako Matsumoto
    Makoto Ozaki
    Yasuko Nagasaka
    JA Clinical Reports, 7
  • [3] Right ventricular perforation, pneumothorax, and a pneumatocele by a pacemaker lead: a case report
    Iwata, Shihoko
    Hirose, Ayana
    Furui, Ikue
    Matsumoto, Takako
    Ozaki, Makoto
    Nagasaka, Yasuko
    JA CLINICAL REPORTS, 2021, 7 (01)
  • [4] Perforation of right ventricular free wall by pacemaker lead detected by multidetector computed tomography
    Yavari, Arash
    Khawaja, Zeeshan O.
    Krishnamoorthy, Sanjay
    McWilliams, Eric T.
    EUROPACE, 2009, 11 (02): : 252 - 254
  • [5] An Unusual Case of Migrating Pacemaker Lead: Right Ventricular Perforation
    Chowdhury, Shamim
    Nesaruddin, Mir
    Malik, Fazila-Tun-Nesa
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 : S52 - S52
  • [6] Late Left Ventricular Perforation by Active Fixation Pacemaker Lead Implanted in the Right Ventricular Septum
    Kondoh, Haruhiko
    Funatsu, Toshihiro
    Taniguchi, Kazuhiro
    JOURNAL OF CARDIAC SURGERY, 2012, 27 (04) : 530 - 531
  • [7] An unusual case of a subacute right ventricular perforation from a pacemaker lead with subsequent left hemothorax
    Dean, Ryan K.
    Subedi, Rogin
    Gill, Dalvir
    Nat, Amitpal
    COR ET VASA, 2018, 60 (05) : E479 - E481
  • [8] SUBACUTE RIGHT VENTRICULAR PERFORATION BY PACEMAKER LEAD PRESENTING WITH LEFT HEMOTHORAX AND SHOCK
    Nichols, Julianne
    Berger, Natalie
    Joseph, Praveen
    Datta, D.
    CRITICAL CARE MEDICINE, 2014, 42 (12)
  • [9] Delayed Right-Ventricular Perforation of a Pacemaker Lead
    Jürgen Hörer
    Albrecht Will
    Christian Schreiber
    Pediatric Cardiology, 2011, 32
  • [10] Rib perforation from a right ventricular pacemaker lead
    Singhal, Sunil
    Cooper, Joshua M.
    Cheung, Albert T.
    Acker, Michael A.
    CIRCULATION, 2007, 115 (14) : E391 - E392