Temporary vascular access for extracorporeal renal replacement therapies in acute renal failure patients

被引:0
|
作者
Canaud, B [1 ]
Leray-Moragues, H
Leblanc, M
Klouche, K
Vela, C
Béraud, JJ
机构
[1] CHU Montpellier, Hop Lapeyronie, Dept Nephrol, Nephrol & Metab Intens Care Unit, F-34295 Montpellier, France
[2] Nephrol Ctr, Montreal, PQ, Canada
关键词
vascular access; dialysis; catheter; acute renal failure;
D O I
暂无
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Temporary vascular access is an essential component to perform any extracorporeal renal replacement therapy (RRT) in the acute renal failure patient. RRT used in the acute setting may be categorized in two groups: intermittent (IRRT) and continuous (CRRT). Therapeutic indications are based on clinical and technical considerations. Continuous modalities are mainly utilized in intensive care units for hemodynamically compromised patient. Initially performed spontaneously via an arteriovenous circuit, CRRT modalities have progressively become venovenous with the circulatory assistance of a blood pump. Since both intermittent and continuous RRT modalities are now performed almost exclusively by venovenous modalities, this article deals exclusively with temporary venous catheters. At present, double-lumen catheters represent the most common vascular access for RRT modalities. Semi-rigid polyurethane catheters currently used in case of emergency are limited to short term use. Hemocompatible, flexible silicone catheters, less aggressive for the vessels, seem better suited for the medium end long term run. The tunneled silicone cathers (DualCath type) meet the short and long term needs, and allows for blood flow rates up to 400 ml/min. The internal jugular vein, particularly the right one, stems to warrant the proper functioning of catheters while reducing the risk of stenotic complications. Subclavian access should be limited in time and reserved fur silicons catheters in order to limit the risk of stenosis and/or thrombosis. Femoral access, very useful in eases of emergency and respiratory problems, greatly impairs the patient's mobility and should be limited by time to prevent thrombosis and/or infection. Late and!or delayed dysfunctioning of catheters are indicative of a thrombosis. Performance standards of catheters are less of a limiting factor in continuous low flow RRT modalities than in the intermittent ones. Finally, careful handling of the catheter essential to prevent infectious complications.
引用
收藏
页码:S142 / S150
页数:9
相关论文
共 50 条
  • [1] Vascular access for acute extracorporeal renal replacement therapies
    Granata, A
    D'Intini, V
    Bellomo, R
    Ronco, C
    HEMODIALYSIS VASCULAR ACCESS AND PERITONEAL DIALYSIS ACCESS, 2004, 142 : 159 - 177
  • [2] Vascular Access for Extracorporeal Renal Replacement Therapy in Veterinary Patients
    Chalhoub, Serge
    Langston, Cathy E.
    Poeppel, Karen
    VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 2011, 41 (01) : 147 - +
  • [3] Vascular access for extracorporeal renal replacement therapies in the intensive care unit in clinical practice
    Canaud, B
    Martin, K
    Nguessan, C
    Klouche, K
    Leray-Loragues, H
    Béraud, JJ
    BLOOD PURIFICATION IN INTENSIVE CARE, 2001, 132 : 266 - 282
  • [4] Solute control by extracorporeal therapies in acute renal failure
    Clark, WR
    Mueller, BA
    Kraus, MA
    Macias, WL
    AMERICAN JOURNAL OF KIDNEY DISEASES, 1996, 28 (05) : S21 - S27
  • [5] Dose capabilities of renal replacement therapies in acute renal failure (ARF).
    Liao, ZJ
    Huang, ZP
    Cui, X
    Gao, DY
    Kraus, MA
    Clark, WR
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2002, 13 : 237A - 237A
  • [7] Extracorporeal therapies in acute renal failure:: Different therapeutic options
    Gretz, N
    Quintel, M
    Kränzlin, B
    KIDNEY INTERNATIONAL, 1998, : S57 - S60
  • [8] Is there a role for continuous renal replacement therapies in patients with liver and renal failure?
    Davenport, A
    KIDNEY INTERNATIONAL, 1999, 56 : S62 - S66
  • [9] Vascular access for extracorporeal renal replacement therapy in the intensive care unit
    Canaud, B
    Formet, C
    Raynal, N
    Amigues, L
    Klouche, K
    Leray-Moragues, H
    Béraud, JJ
    SEPSIS, KIDNEY AND MULTIPLE ORGAN DYSFUNCTION, 2004, 144 : 291 - 307
  • [10] Vascular access and extracorporeal circuit patency in continuous renal replacement therapy
    Gonzalez de Molina, F. J.
    Galindo, M.
    Gonzalez, C.
    Broch, M. J.
    del Bano, L.
    Roglan, A.
    MEDICINA INTENSIVA, 2016, 40 (09) : 572 - 585