Correcting scaling errors in tomographic images using a nine degree of freedom registration algorithm

被引:35
|
作者
Hill, DLG
Maurer, CR
Studholme, C
Fitzpatrick, JM
Hawkes, DJ
机构
[1] United Med & Dent Sch Guys & St Thomas Hosp, Dept Radiol Sci, London SE1 9RT, England
[2] Vanderbilt Univ, Dept Comp Sci, Nashville, TN 37235 USA
[3] Vanderbilt Univ, Dept Neurol Surg, Nashville, TN USA
关键词
image registration; magnetic resonance imaging; computed tomography; physics and instrumentation; physics;
D O I
10.1097/00004728-199803000-00031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Clinical imaging systems, especially MR scanners, frequently have errors of a few percent in their voxel dimensions. We evaluate a nine degree of freedom registration algorithm that maximizes mutual information for determining scaling errors. We evaluate it by registering MR and CT images for each of five patients (patient scaling) and by registering MR images of a phantom to a computer model of the phantom (phantom scaling). Method: Each scaling method was validated using bone-implanted markers localized in the patient images and also intraoperatively. The root mean square residual in the alignment of the fiducial markers [fiducial registration error (FRE)] was determined without scale correction, with patient scaling, and with phantom scaling. Results: Each scaling method significantly reduced the average FRE (p < 0.05) for MR to CT registration and for MR to physical space registration, indicating that voxel scaling errors were reduced. The greater reduction in scaling errors was achieved using the phantom scaling method. Conclusion: We have demonstrated that a nine degree of freedom registration algorithm that maximizes mutual information can significantly reduce scaling errors in MR.
引用
收藏
页码:317 / 323
页数:7
相关论文
共 50 条
  • [1] Correcting scaling errors in tomographic images using a nine degree of freedom registration algorithm
    Department of Radiological Sciences, UMDS, Guy's and St. Thomas' Hospitals, London, United Kingdom
    不详
    不详
    J. Comput. Assisted Tomogr., 2 (317-323):
  • [2] Registration of SAR images using the chirp scaling algorithm
    Fernandes, D
    Waller, G
    Moreira, JR
    IGARSS '96 - 1996 INTERNATIONAL GEOSCIENCE AND REMOTE SENSING SYMPOSIUM: REMOTE SENSING FOR A SUSTAINABLE FUTURE, VOLS I - IV, 1996, : 799 - 801
  • [3] An improved algorithm for intraoperative registration of computed tomographic left atrial images
    Zhong, Hua
    Schwartzman, David
    EUROPACE, 2011, 13 (03): : 383 - 388
  • [4] A new DSP architecture for correcting errors using Viterbi algorithm
    Yoon, S
    Kim, S
    Oh, J
    Kang, S
    ADVANCED INTERNET SERVICES AND APPPLICATIONS, PROCEEDINGS, 2002, 2402 : 95 - 102
  • [5] Registration of video images to tomographic images by optimising mutual information using texture mapping
    Clarkson, MJ
    Rueckert, D
    King, AP
    Edwards, PJ
    Hill, DLG
    Hawkes, DJ
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION, MICCAI'99, PROCEEDINGS, 1999, 1679 : 579 - 588
  • [6] Correcting Image Placement Errors Using Registration Control (RegC®) Technology In The Photomask Periphery
    Cohen, Avi
    Lange, Falk
    Ben-Zvi, Guy
    Graitzer, Erez
    Vladimir, Dmitriev
    PHOTOMASK TECHNOLOGY 2012, 2012, 8522
  • [7] Correcting Image Placement Errors Using Registration Control (RegC®) Technology In The Photomask Periphery
    Cohen, Avi
    Lange, Falk
    Ben-Zvi, Guy
    Graitzer, Erez
    Vladimir, Dmitriev
    28TH EUROPEAN MASK AND LITHOGRAPHY CONFERENCE, 2012, 8352
  • [8] A novel cross-range scaling algorithm for ISAR images based on feature registration
    Xu, Zhi-Wei
    Zhang, Lei
    Xing, Meng-Dao
    Dianzi Yu Xinxi Xuebao/Journal of Electronics and Information Technology, 2014, 36 (09): : 2173 - 2179
  • [9] Registration of three-dimensional LA computed tomographic images with projection images obtained using fluoroscopy
    Chen, PS
    HEART RHYTHM, 2006, 3 (03) : 375 - 376
  • [10] An Unbiased Groupwise Registration Algorithm for Correcting Motion in Dynamic Contrast-Enhanced Magnetic Resonance Images
    Mojica, Mia
    Ebrahimi, Mehran
    IMAGE ANALYSIS FOR MOVING ORGAN, BREAST, AND THORACIC IMAGES, 2018, 11040 : 42 - 52