Design and evaluation of a novel triaxial isometric trunk muscle strength measurement system

被引:25
|
作者
Azghani, M. R. [1 ]
Farahmand, F. [1 ,2 ]
Meghdari, A. [1 ]
Vossoughi, G. [1 ]
Parnianpour, M. [1 ,3 ]
机构
[1] Sharif Univ Technol, Dept Mech Engn, Tehran, Iran
[2] Univ Tehran Med Sci, RCSTIM, Tehran, Iran
[3] Hanyang Univ, Seoul 133791, South Korea
基金
美国国家科学基金会;
关键词
trunk; triaxial strength measurement; design; reliability; maximum voluntary exertion; LOW-BACK-PAIN; PERFORMANCE; LEVEL; RECRUITMENT; GENERATION; EXTENSION; PATTERNS; ROTATION; FLEXION; FORCES;
D O I
10.1243/09544119JEIM537
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Maximal strength measurements of the trunk have been used to evaluate the maximum functional capacity of muscles and the potential mechanical overload or overuse of the lumbar spine tissues in order to estimate the risk of developing musculoskeletal injuries. A new triaxial isometric trunk strength measurement system was designed and developed in the present study, and its reliability and performance was investigated. The system consisted of three main revolute joints, equipped with torque sensors, which intersect at L5-S1 and adjustment facilities to fit the body anthropometry and to accommodate both symmetric and asymmetric postures in both seated and standing positions. The dynamics of the system was formulated to resolve validly the moment generated by trunk muscles in the three anatomic planes. The optimal gain and offset of the system were obtained using deadweights based on the least-squares linear regression analysis. The R-2 results of calibration for all loading courses of all joints were higher than 0.99, which indicated an excellent linear correlation. The results of the validation analysis of the regression model suggested that the mean absolute error and the r.m.s. error were less than 2 per cent of the applied load. The maximum value of the minimum detectable change was found to be 1.63 N m for the sagittal plane torque measurement, 0.8 per cent of the full-scale load. The trial-to-trial variability analysis of the device using deadweights provided intra-class correlation coefficients of higher than 0.99, suggesting excellent reliability. The cross-talk analysis of the device indicated maximum cross-talks of 1.7 per cent and 3.4 per cent when the system was subjected to flexion-extension and lateral bending torques respectively. The trial-to-trial variability of the system during in-vivo strength measurement tests resulted in good to excellent reliability, with intra-class correlation coefficients ranging from 0.69 to 0.91. The results of the maximum voluntary isometric torques exertion measurements for 30 subjects indicated good agreement with the previously published data in the literature. The extensive capabilities and high reliability of the system are promising for more comprehensive investigations on the trunk biomechanics in future, e.g. isometric strength measurement at symmetric and asymmetric postures, muscle endurance, and recruitment pattern analysis.
引用
收藏
页码:755 / 766
页数:12
相关论文
共 50 条
  • [1] Reliability of isometric muscle strength tests for the trunk, hands and shoulders
    Essendrop, M
    Schibye, B
    Hansen, K
    INTERNATIONAL JOURNAL OF INDUSTRIAL ERGONOMICS, 2001, 28 (06) : 379 - 387
  • [2] CORRELATIONS BETWEEN ISOTONIC AND ISOMETRIC MEASUREMENTS OF TRUNK MUSCLE STRENGTH
    LEWIS, VM
    MERRITT, JL
    PIPER, SM
    SINAKI, M
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1987, 68 (09): : 639 - 640
  • [3] Maximum isometric trunk muscle strength and activity at trunk axial rotation during sitting
    Torén, A
    Öberg, K
    APPLIED ERGONOMICS, 1999, 30 (06) : 515 - 525
  • [4] A SYSTEM FOR MEASURING ISOMETRIC MUSCLE STRENGTH
    AMUNDSEN, L
    PATTERSON, R
    BAXTER, T
    SCUDDER, G
    PHYSICAL THERAPY, 1987, 67 (05): : 757 - 757
  • [5] Reliability of a Test for Assessment of Isometric Trunk Muscle Strength in Elderly Women
    Mesquita, Marceli M. A.
    Santos, Marta S.
    Vasconcelos, Alan B. S.
    de Sa, Clodoaldo A.
    Pereira, Luana C. D.
    da Silva-Santos, Inea B. M.
    da Silva Junior, Walderi M.
    de Matos, Dihogo G.
    Fontes, Alan dos S.
    Oliveira, Paulo M. P.
    Aidar, Felipe J.
    DeSantana, Josimari M.
    Fernandes, Iohanna G. S.
    Da Silva-Grigoletto, Marzo E.
    JOURNAL OF AGING RESEARCH, 2019, 2019
  • [6] Reliability of isokinetic trunk muscle strength measurement
    Karatas, GK
    Gögüs, F
    Meray, J
    AMERICAN JOURNAL OF PHYSICAL MEDICINE & REHABILITATION, 2002, 81 (02) : 79 - 85
  • [7] Screen Time Viewing Behaviors and Isometric Trunk Muscle Strength in Youth
    Grontved, Anders
    Ried-Larsen, Mathias
    Froberg, Karsten
    Wedderkopp, Niels
    Brage, Soren
    Kristensen, Peter Lund
    Andersen, Lars Bo
    Moller, Niels Christian
    MEDICINE & SCIENCE IN SPORTS & EXERCISE, 2013, 45 (10) : 1975 - 1980
  • [8] A novel adaptable isometric strength analysis and exercise development system design
    Fidan, Murat
    Bayrak, Alper
    Karli, Umid
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2021, 235 (08) : 913 - 926
  • [9] Isometric muscle strength measurement and its clinical use
    Stoll, T
    Proceedings of the 3rd World Congress of the International Society of Physical and Rehabilitation Medicine ISPRM, 2005, : 5 - 7
  • [10] Portable, One-Dimensional, Trunk-Flexor Muscle Strength Measurement System
    Loss, Jefferson Fagundes
    Wagner Neto, Edgar Santiago
    de Siqueira, Tatiane Borsoi
    Winck, Aline Dill
    de Moura, Laura Silveira
    Gertz, Luiz Carlos
    JOURNAL OF SPORT REHABILITATION, 2020, 29 (06) : 851 - 854