Whole brain 31P MRSI at 7T with a dual-tuned receive array

被引:15
|
作者
Rowland, Benjamin C. [1 ,2 ]
Driver, Ian D. [2 ]
Tachrount, Mohamed [2 ]
Klomp, Dennis W. J. [3 ,4 ]
Rivera, Debra [4 ]
Forner, Ria [4 ]
Pham, Anh [4 ]
Italiaander, Michel [4 ]
Wise, Richard G. [2 ]
机构
[1] Univ Manchester, Div Canc Sci, Manchester M13 9PL, Lancs, England
[2] Cardiff Univ, Brain Res Imaging Ctr, Sch Psychol, Cardiff, S Glam, Wales
[3] UMC Utrecht, Dept Radiol, Utrecht, Netherlands
[4] MR Coils, Zaltbommel, Netherlands
基金
英国医学研究理事会;
关键词
dual-tuned; MRSI; P-31; phosphorus; receive array; spectroscopy; MAGNETIC-RESONANCE-SPECTROSCOPY; PHASED-ARRAY; RF COIL; H-1; RECONSTRUCTION; COMBINATION; FREQUENCY;
D O I
10.1002/mrm.27953
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The design and performance of a novel head coil setup for P-31 spectroscopy at ultra-high field strengths (7T) is presented. The described system supports measurements at both the H-1 and P-31 resonance frequencies. Methods The novel coil consists of 2, actively detunable, coaxial birdcage coils to give homogeneous transmit, combined with a double resonant 30 channel receive array. This allows for anatomical imaging combined with P-31 acquisitions over the whole head, without changing coils or disturbing the subject. A phosphate buffer phantom and 3 healthy volunteers were scanned with a pulse acquire CSI sequence using both the novel array coil and a conventional transceiver birdcage. Four different methods of combining the array channels were compared at 3 different levels of SNR. Results The novel coil setup delivers significantly increased P-31 SNR in the peripheral regions of the brain, reaching up to factor 8, while maintaining comparable performance relative to the birdcage in the center. Conclusions The new system offers the potential to acquire whole brain P-31 MRSI with superior signal relative to the standard options.
引用
收藏
页码:765 / 775
页数:11
相关论文
共 50 条
  • [1] Whole-body radiofrequency coil for 31P MRSI at 7T
    Loring, J.
    van der Kemp, W. J. M.
    Almujayyaz, S.
    van Oorschot, J. W. M.
    Luijten, P. R.
    Klomp, D. W. J.
    NMR IN BIOMEDICINE, 2016, 29 (06) : 709 - 720
  • [2] Repeatability of 31P MRSI in the human brain at 7T with and without the nuclear Overhauser effect
    Lagemaat, Miriam W.
    van de Bank, Bart L.
    Sati, Pascal
    Li, Shizhe
    Maas, Marnix C.
    Scheenen, Tom W. J.
    NMR IN BIOMEDICINE, 2016, 29 (03) : 256 - 263
  • [3] Volumetric mapping of intra- and extracellular pH in the human brain using 31P MRSI at 7T
    Korzowski, Andreas
    Weinfurtner, Nina
    Mueller, Sebastian
    Breitling, Johannes
    Goerke, Steffen
    Schlemmer, Heinz-Peter
    Ladd, Mark E.
    Paech, Daniel
    Bachert, Peter
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (04) : 1707 - 1723
  • [4] Phosphorus liver MRSI at 3 T using a novel dual-tuned eight-channel 31P/1H coil
    Panda, Anshuman
    Jones, Scott
    Stark, Helmut
    Raghavan, Rahul S.
    Sandrasegaran, Kumar
    Bansal, Navin
    Dydak, Ulrike
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (05) : 1346 - 1356
  • [5] Evaluation of a new 1H/31P dual-tuned birdcage coil for 31P spectroscopy
    Potter, W. M.
    Wang, L.
    McCully, K. K.
    Zhao, Q.
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2013, 43 (03) : 90 - 99
  • [6] An 8-channel receive array for improved P-31 MRSI of the whole brain at 3T
    van Uden, Mark J.
    Peeters, Tom H.
    Rijpma, Anne
    Rodgers, Christopher T.
    Heerschap, Arend
    Scheenen, Tom W. J.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 82 (02) : 825 - 832
  • [7] A novel multifrequency-tuned transceiver array for human-brain 31P-MRSI at 7 T
    Li, Xin
    Zhu, Xiao-Hong
    Zhang, Xiao-Liang
    Waks, Matt
    Chen, Wei
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (06) : 2640 - 2654
  • [8] Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T
    Valkovic, Ladislav
    Dragonu, Iulius
    Almujayyaz, Salam
    Batzakis, Alex
    Young, Liam A. J.
    Purvis, Lucian A. B.
    Clarke, William T.
    Wichmann, Tobias
    Lanz, Titus
    Neubauer, Stefan
    Robson, Matthew D.
    Klomp, Dennis W. J.
    Rodgers, Christopher T.
    PLOS ONE, 2017, 12 (10):
  • [9] Mapping an Extended Metabolic Profile of Gliomas Using High-Resolution 31P MRSI at 7T
    Korzowski, Andreas
    Weckesser, Nina
    Franke, Vanessa L.
    Breitling, Johannes
    Goerke, Steffen
    Schlemmer, Heinz-Peter
    Ladd, Mark E.
    Bachert, Peter
    Paech, Daniel
    FRONTIERS IN NEUROLOGY, 2021, 12
  • [10] 31P MRSI and 1H MRS at 7T: initial results in human breast cancer
    Klomp, Dennis W. J.
    van de Bank, Bart L.
    Raaijmakers, Alexander
    Korteweg, Mies A.
    Possanzini, Cecilia
    Boer, Vincent O.
    de Berg, Cornelius A. T. van
    van de Bosch, Maurice A. A. J.
    Luijten, Peter R.
    NMR IN BIOMEDICINE, 2011, 24 (10) : 1337 - 1342