Multi-channel plastic-scintillator-based detection system for monitoring tritium in air

被引:2
|
作者
Bae, Jun Woo [1 ]
Kang, Ki Joon [1 ]
Kim, Hee Reyoung [1 ]
Jeon, SangJune [2 ]
机构
[1] Ulsan Natl Inst Sci & Technol, Dept Nucl Engn, 50 Unist Gil, Ulsan 44919, South Korea
[2] Sans Frontier Technol, 506 Samsung Leaders Tower, Seoul 08511, South Korea
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2019年 / 90卷 / 09期
基金
新加坡国家研究基金会;
关键词
IONIZATION CHAMBERS; HYDROGEN STORAGE; TRITIATED-WATER; EFFICIENCY; COUNTER; VAPOR;
D O I
10.1063/1.5092543
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To overcome the limitations of the ionization chamber-based tritium monitor, a design for a multichannel plastic scintillator-based detection chamber for monitoring tritium in air is proposed. The performance of the chamber was characterized by Monte Carlo-based calculations with various design parameters such as thickness of the plastic scintillator (t) and number of channels (n). We considered the volume and detection efficiency of the chamber to evaluate the performance of the detector. The expected counting rate curve was in good agreement with the expected simplified physical model. The minimum detectable activity (MDA) was shown at t = 0.5 mm and n = 48 and estimated to be 29.9 kBq/m(3). Compared with the experimentally estimated MDA, the relative difference was approximately 24%. However, this difference is understandable considering the different lower level of discrimination and light loss during transportation assumed in the model. The proposed tritium detection chamber is expected to be useful for environmental monitoring at some level as well as for monitoring of tritium leaks from fusion or CANada Deuterium Uranium (CANDU)-type reactors.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Detection efficiency of plastic scintillator for gaseous tritium sampling and measurement system
    Uda, Tatsuhiko
    Kawano, Takao
    Tanaka, Masahiro
    Tomatsuri, Satoshi
    Ito, Takeshi
    Tatenuma, Katsuyoshi
    FUSION ENGINEERING AND DESIGN, 2010, 85 (7-9) : 1474 - 1478
  • [2] Multi-channel electrochemical detection system based on LabVIEW
    Li, HQ
    Luo, XB
    Liu, CX
    Jiang, LY
    Cui, DF
    Cai, XX
    Yang, QD
    ICIA 2004: Proceedings of 2004 International Conference on Information Acquisition, 2004, : 224 - 227
  • [3] UNIQUE CHANNEL DETECTION IN A MULTI-CHANNEL SYSTEM
    MARINO, PF
    SIAM REVIEW, 1963, 5 (01) : 93 - &
  • [4] Design of Multi-Channel Weighting Monitoring System
    Shi Lilian
    Xu Guoxing
    Shen Hongwei
    Ye Jun
    PROCEEDINGS OF THE THIRD INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1 - 4, 2010, : 1275 - 1278
  • [5] Design of Multi-channel analyzer's monitoring system based on embedded system
    Yang, Tao
    Wei, Yi-Xiang
    Hedianzixue Yu Tance Jishu/Nuclear Electronics and Detection Technology, 2007, 27 (04): : 745 - 747
  • [6] Design of an Internet of Things - based multi-channel temperature monitoring system
    Clerbaux, B.
    de Gracia, D. Gomez
    Petitjean, P.
    Yang, Y.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (04)
  • [7] Remote based multi-channel leak monitoring system using correlation
    Liang Wei
    Zhang Laibin
    Wang Zhaohui
    ENGINEERING STRUCTURAL INTEGRITY: RESEARCH, DEVELOPMENT AND APPLICATION, VOLS 1 AND 2, 2007, : 209 - 212
  • [8] Collaborative biofluid analysis based multi-channel integrated wearable detection system for the monitoring of wound infection
    Chen Z.
    Dong Y.
    Fu J.
    Bai Y.
    Gao Q.
    Qin Z.
    Wang J.
    Li S.
    Biosensors and Bioelectronics: X, 2024, 17
  • [9] Novel multi-layer plastic-scintillator-based solid active proton target for inverse-kinematics experiments
    Tran, D. T.
    Terashima, S.
    Ong, H. J.
    Hirakawa, K.
    Matsuda, Y.
    Aoi, N.
    Harakeh, M. N.
    Itoh, M.
    Kawabata, T.
    Kohda, A.
    Matsumoto, S. Y.
    Nishi, T.
    Okamoto, J.
    Tanihata, I
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2020, 959
  • [10] Evaluation of multi-channel PMTs for readout of scintillator arrays
    Shao, Yiping
    Cherry, Simon R.
    Siegel, Stefan
    Silverman, Robert W.
    Majewski, Stan
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997, 390 (1-2): : 209 - 218