A precise navigation device for fixation of patella fractures with modified K-wire tension band:a comparative retrospective study

被引:2
|
作者
Wang, Fuming [1 ,2 ]
Xiong, Haolan [3 ]
Long, Xiaotao [2 ]
Li, Yang [2 ]
Chen, Xiaohua [2 ]
Wang, Gang [1 ]
机构
[1] Southern Med Univ, Nanfang Hosp, Dept Orthopaed & Traumatol, Guangzhou, Guangdong, Peoples R China
[2] Univ Chinese Acad Sci, Chongqing Gen Hosp, Dept Orthopaed & Traumatol, Chongqing, Peoples R China
[3] Univ Chinese Acad Sci, Chongqing Gen Hosp, Dept Anesthesiol, Chongqing, Peoples R China
关键词
Navigation device; Patella fractures; Modified K-wire tension band;
D O I
10.1186/s13018-021-02235-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Traditionally, the technique of modified tension band wires (MTBW) has been the most commonly used surgical procedure. The purpose of this study is to design a precise navigation device that can obtain a standard position of K-wires for (MTBW) and to compare the precise MTBW (P-MTBW) by a navigation device with the conventional MTBW (C-MTBW) by hands in a retrospective study. Methods: The device was designed by solidworks2012 software (USA), which could provide a precise guidance for obtaining parallel K-wires. Besides, it could set the distance between two K-wires and the level of K-wires below patellar anterior surface. From June 2014 to August 2018, a total of 112 patients were employed in this retrospective study. The patients were divided into P-MTBW group and C-MTBW group according to the surgical technique with or without the precise navigation device. We needed to record and analyze the operation time and the number of fluoroscopy, postoperative internal fixation imaging, knee function and complications. Results: There were 54 patients in P-MTBW group and 58 patients in C-MTBW group. There were statistically significant differences (P < 0.001) in the operation time between P-MTBW group (39.5 +/- 4.7; range, 32-49 min) and C-MTBW group (53.7 +/- 6.8; range, 42-71 min). The number of intraoperative fluoroscopy was significantly less (P < 0.001) in P-MTBW group (4.2 +/- 1.4) versus that of C-MTBW group (8.3 +/- 2.7). According to Iowa knee score, there was no significant difference (P = 0.268 at 1 year) in function between the two groups. According to our own evaluation criteria for MTBW, anyone in the P-MTBW group was excellent and 26 patients were excellent, 20 patients were good, and 2 patients were fair in the C-MTBW group. Conclusion: The navigation device can reduce operation time and intraoperative fluoroscopy frequency. P-MTBW fixation is an accurate and effective surgical procedure for patella fractures.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] A precise navigation device for fixation of patella fractures with modified K-wire tension band:a comparative retrospective study
    Fuming Wang
    Haolan Xiong
    Xiaotao Long
    Yang Li
    Xiaohua Chen
    Gang Wang
    Journal of Orthopaedic Surgery and Research, 16
  • [2] Efficacy of K-wire tension band fixation compared with other alternatives for patella fractures: a meta-analysis
    Zhang, Yinwang
    Xu, Zhen
    Zhong, Wuxue
    Liu, Fuhai
    Tang, Jie
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2018, 13
  • [3] Efficacy of K-wire tension band fixation compared with other alternatives for patella fractures: a meta-analysis
    Yinwang Zhang
    Zhen Xu
    Wuxue Zhong
    Fuhai Liu
    Jie Tang
    Journal of Orthopaedic Surgery and Research, 13
  • [4] Clinical results of treatment using a modified K-wire tension band versus a cannulated screw tension band in transverse patella fractures: A strobe-compliant retrospective observational study
    Tan, Honglue
    Dai, Pengyi
    Yuan, Yanhao
    MEDICINE, 2016, 95 (40)
  • [5] Cable pin system versus K-wire tension band fixation for patella fractures in Chinese Han population: A meta-analysis
    Zha, Kun
    Liu, Guo-hui
    Yang, Shu-hua
    Zhou, Wu
    Liu, Yi
    Wu, Qi-peng
    JOURNAL OF HUAZHONG UNIVERSITY OF SCIENCE AND TECHNOLOGY-MEDICAL SCIENCES, 2017, 37 (05) : 667 - 674
  • [6] Cable Pin System versus K-wire Tension Band Fixation for Patella Fractures in Chinese Han Population: A Meta-analysis
    查琨
    刘国辉
    杨述华
    周武
    刘毅
    吴其鹏
    Current Medical Science, 2017, (05) : 667 - 674
  • [7] Cable pin system versus K-wire tension band fixation for patella fractures in Chinese Han population: A meta-analysis
    Kun Zha
    Guo-hui Liu
    Shu-hua Yang
    Wu Zhou
    Yi Liu
    Qi-peng Wu
    Current Medical Science, 2017, 37 : 667 - 674
  • [9] K-wire and tension band wire fixation in treating sternoclavicular joint dislocation
    Chen Qing-yu
    Cheng Shao-wen
    Wang Wei
    Lin Zhong-qin
    Zhang Wei
    Kou Dong-quan
    Shen Yue
    Ying Xiao-zhou
    Cheng Xiao-jie
    Lu Chuan-zhu
    Peng Lei
    CHINESE JOURNAL OF TRAUMATOLOGY, 2011, 14 (01) : 53 - 57
  • [10] Treatment of displaced fractures of the patella: Tension band wiring technique with the one-end or both-ends K-wire bending fixation method
    Huang, Po-Hua
    Hsu, Chi-Hsiang
    Hsu, Shan-Ling
    Liu, Hao-Chen
    JOURNAL OF ORTHOPAEDIC SURGERY, 2021, 29 (01)