Small-angle Compton Scatter Artifact in Tc-99m-IDA Hepatobiliary Scintigraphy Resulting in the Breast Overlying the Liver in Planar Dynamic Imaging

被引:0
|
作者
Qutbi, Mohsen [1 ]
机构
[1] Shahid Beheshti Univ Med Sci, Taleghani Educ Hosp, Dept Nucl Med, Tehran, Iran
关键词
Small-angle Compton scatter; artifact; liver; breast; hepatobiliary scan;
D O I
10.4274/mirt.galenos.2020.05658
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Compton scatter photons are generally considered problematic in nuclear medicine imaging. Therefore, efforts are being made to minimize the involvement of these photons by employing some special strategies in daily practice. Basically, photons scattering at a small angle and traveling in the proper direction stand a chance of getting recorded and thereby participate in the image formation. These photons may create artifactual hot spots in the vicinity of a region with high concentration of radioactivity. The present study focuses on the negative impact of such photons during routine imaging in clinical setting, through an artifact detected in technetium-99m-IDA hepatobiliary scintigraphy, with the purpose of highlighting this issue to the nuclear medicine practitioners.