Automated blood component collection with the MCS 3p cell separator: evaluation of three protocols for buffy coat-poor and white cell-reduced packed red cells and plasma

被引:21
|
作者
Zeiler, TA
Kretschmer, V
机构
[1] Dept. Transfus. Med. Hemostaseology, Philipps University Clinics Marburg, Marburg
[2] Dept. Transfus. Med. Hemostaseology, Philipps University Clinics, D-35043, Marburg, Conradistrasse
[3] Dept. Transfus. Med. Hemostaseology, University Clinics, Philipps University
关键词
D O I
10.1046/j.1537-2995.1997.37897424400.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND: Automated collection of blood components with a cell separator (MCS 3p, Haemonetics), was performed according to three protocols. STUDY DESIGN AND METHODS:The first protocol provided 2 units of fresh-frozen plasma (FFP); and one buy, coat-poor red cell (RBC) concentrate in additive solution. The second protocol included an additional inline filtration of the RBC in a closed system after storage at 4 degrees C for 24 hours. In the third protocol, an additional platelet concentrate (PC) was recovered from the buffy coat. Cell counts and biochemical characterization of the RBCs (n = 20 each) were determined on Days 0, 1, 14, 28, and 49. RESULTS: The RBC volume was 336 +/- 9 mt (first protocol), 337 +/- 7 mt (second protocol) and 293 +/- 12 mt (third protocol) with a hematocrit of 59 +/- 2, 53 +/- 3, and 61 +/- 5, percent respectively. On Day 49, hemolysis was 0.24 +/- 0.1 percent (first protocol), 0.33 +/- 0.32 percent (second protocol), and 0.38 +/- 0.1 percent(third protocol). The filtered RBC concentrate met the international standards for white cell-reduced RBCs. Filtration resulted in a clinically irrelevant increase of hemolysis. The in vitro RBC values (lactate dehydrogenase, 2-hydroxybutyrate dehydrogenase, hemolysis, potassium, 2,3 DPG, ATP) were at least equal to those in RBCs collected by conventional whole-blood donation. There is a trend toward extended preservation of 2,3 DPG in RBCs collected by apheresis. Two units of FFP could be collected with each donation (first protocol: 420 +/- 55 mt, 5.4 +/- 7 WBCs/mu L, 6.5 +/- 5 x 10(3) platelets/mu L; second protocol: 440 +/- 33 mL, 3 +/- 5.2 WBCs/mu L, 32 +/- 12 x 10(3) platelets/mu L; third protocol: 398 +/- 32 mt, 5 +/- 12 WBCs/mu L; 3.4 +/- 3.5 x 10(3) platelets/mu L). PCs prepared from the buffy coat collected by the third protocol contained 90 +/- 30 x 10(9) platelets in 88 +/- 14 mt of plasma. In vitro test results in these PCs were superior to those in PCs collected by conventional whole-blood donation. The procedure was well tolerated by all donors. No adverse reactions appeared. CONCLUSION: Erythroplasmapheresis with the MCS 3p cell separator is a useful alternative to conventional whole-blood donation and separation.
引用
收藏
页码:791 / 797
页数:7
相关论文
共 4 条
  • [1] Distribution of ''mobilized'' blood cells along the buffy coat produced by the hemonetics MCS-3P cell separator
    Ghielmini, M
    Pfister, U
    Bertoli, G
    VandenBosch, S
    Derivaz, JL
    Cavalli, F
    BLOOD, 1995, 86 (10) : 3920 - 3920
  • [2] A new separation protocol (DRBCP-F) for automatic blood component donation with the cell separator MCS 3p for collection of leukocyte-depleted red blood cell concentrates and plasma
    Zeiler, T
    Kretschmer, V
    TRANSFUSIONSMEDIZIN 1996/97, 1997, 34 : 118 - 122
  • [3] Distribution of mobilized progenitor cells in the buffy coat of the haemonetics MCS3p cell separator: A study to optimize the collection of progenitors by leukapheresis
    Ghielmini, M
    Pfister, U
    Zucca, E
    Van den Bosch, S
    Tamasy, P
    Marangoni, G
    Bertoli, G
    Derivaz, JL
    Cavalli, F
    JOURNAL OF HEMATOTHERAPY, 1998, 7 (03): : 251 - 256
  • [4] Immunologic changes after transfusion of autologous or allogeneic buffy coat-poor versus white cell-reduced blood to patients undergoing arthroplasty.: I.: Proliferative T-cell responses and the balance of helper and suppressor T cells
    Innerhofer, P
    Luz, G
    Spötl, L
    Hobisch-Hagen, P
    Schobersberger, W
    Fischer, M
    Nussbaumer, W
    Lochs, A
    Irschick, E
    TRANSFUSION, 1999, 39 (10) : 1089 - 1096