A Wearable Thermometry for Core Body Temperature Measurement and Its Experimental Verification

被引:35
|
作者
Huang, Ming [1 ]
Tamura, Toshiyo [2 ]
Tang, Zunyi [2 ]
Chen, Wenxi [3 ]
Kanaya, Shigehiko [1 ]
机构
[1] Nara Inst Sci & Technol, Nara 6300192, Japan
[2] Osaka Electrocommun Univ, Neyagawa, Osaka 5750063, Japan
[3] Univ Aizu, Aizu Wakamatsu, Fukushima 9658580, Japan
关键词
Circadian rhythm; clinical use; core body temperature; healthcare; wearable thermometry;
D O I
10.1109/JBHI.2016.2532933
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A wearable thermometry for core body temperature (CBT) measurement has both healthcare and clinical applications. On the basis of the mechanism of bioheat transfer, we earlier designed and improved a wearable thermometry using the dual-heat-flux method for CBT measurement. In this study, this thermometry is examined experimentally. We studied a fast-changing CBT measurement (FCCM, 55 min, 12 subjects) inside a thermostatic chamber and performed long-term monitoring of CBT (LTM, 24 h, six subjects). When compared with a reference, the Core Temp CM-210 by Terumo, FCCM shows 0.07 degrees C average difference and a 95% CI of [-0.27, 0.12] degrees C. LTM shows no significant difference in parameters for the inference of circadian rhythm. The FCCM and LTM both simulated scenarios in which this thermometry could be used for intensive monitoring and daily healthcare, respectively. The results suggest that because of its convenient design, this thermometry may be an ideal choice for conventional CBT measurements.
引用
下载
收藏
页码:708 / 714
页数:7
相关论文
共 50 条
  • [1] Core temperature measurement by wearable device
    Nakagawa S.
    Nakagawa, Shinya, 1600, Japan Institute of Electronics Packaging (23): : 342 - 346
  • [2] An Ergonomic Wearable Core Body Temperature Sensor
    Atallah, Louis
    Ciuhu, Calina
    Wang, Chao
    Bongers, Edwin
    Blom, Toon
    Paulussen, Igor
    Noordergraaf, Gerrit-Jan
    2018 IEEE 15TH INTERNATIONAL CONFERENCE ON BIOMEDICAL AND HEALTH INFORMATICS (BHI) AND THE WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS (BSN), 2018, : 70 - 73
  • [3] Zero-Heat-Flux Thermometry for Non-Invasive Measurement of Core Body Temperature in Pigs
    Guschlbauer, Maria
    Maul, Alexandra C.
    Yan, Xiaowei
    Herff, Holger
    Annecke, Thorsten
    Sterner-Kock, Anja
    Boettiger, Bernd W.
    Schroeder, Daniel C.
    PLOS ONE, 2016, 11 (03):
  • [4] Measurement of horse core body temperature
    Green, AR
    Gates, SG
    Lawrence, LM
    JOURNAL OF THERMAL BIOLOGY, 2005, 30 (05) : 370 - 377
  • [5] Correlation of tympanic thermometry with standard measures of core body temperature.
    Potti, A
    Pandita, D
    Rabadi, KM
    Carson, PJ
    JOURNAL OF INVESTIGATIVE MEDICINE, 1998, 46 (07) : 275A - 275A
  • [6] Wearable Personal Core Body Temperature Measurement Considering Individual Differences and Dynamic Tissue Blood Perfusion
    Shan, Chengcheng
    Hu, Jiawen
    Zou, Jincheng
    Zhang, Aili
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2022, 26 (05) : 2158 - 2168
  • [7] CORE BODY-TEMPERATURE IN THE ELDERLY AND FACTORS WHICH INFLUENCE ITS MEASUREMENT
    MARION, GS
    MCGANN, KP
    CAMP, DL
    GERONTOLOGY, 1991, 37 (04) : 225 - 232
  • [8] Measurement of the core temperature of the body - A methodic challenge?
    Esch, JSA
    Horn, EP
    ANASTHESIOLOGIE INTENSIVMEDIZIN NOTFALLMEDIZIN SCHMERZTHERAPIE, 1997, 32 (07): : 399 - 400
  • [9] Temporal thermometry fails to track body core temperature during heat stress
    Low, David A.
    Vu, Albert
    Brown, Marilee
    Davis, Scott L.
    Keller, David M.
    Levine, Benjamin D.
    Crandall, Craig G.
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2007, 39 (07): : 1029 - 1035
  • [10] Microchip transponder thermometry for monitoring core body temperature of antelope during capture
    Rey, Benjamin
    Fuller, Andrea
    Hetem, Robyn S.
    Lease, Hilary M.
    Mitchell, Duncan
    Meyer, Leith C. R.
    JOURNAL OF THERMAL BIOLOGY, 2016, 55 : 47 - 53