Intraoperative microvascular Doppler sonography in aneurysm surgery

被引:51
|
作者
Akdemir, H.
Oktem, I. S.
Tucer, B.
Menku, A.
Basaslan, K.
Gunaldi, O.
机构
[1] 2nd Neurosurg Clin, Bakirkoy Psychiat & Neurol Dis Hosp, Istanbul, Turkey
[2] Erciyes Univ, Sch Med, Dept Neurosurg, Kayseri, Turkey
关键词
aneurysm; intraoperative; microvascular; Doppler;
D O I
10.1055/s-2006-954577
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: The goal of this study was to evaluate the efficacy and reliability of intraoperative microvascular Doppler sonography (MDS) for the assessment of cerebral haemodynamics in aneurysm surgery. Methods: For 40 patients (21 men, 19 women, mean age 54.0 years, range 23-73 years) who underwent surgery for the treatment of 48 intracranial aneurysms, microvascular Doppler sonography with a 20-MHz microprobe was used before and after clip application, to confirm the complete obliteration of the aneurysm. Postoperative angiography was performed to assess the complete occlusion of the aneurysm and the patency of adjacent vessels. The findings of MDS were analysed and compared with the postoperative angiography. Results: A 1mm diameter microprobe was able to insonate all vessels of the circle of Willis and their major branches and perforating arteries were reliably insonated. The aneurysm clip was repositioned on the basis of the MDS findings in 12 out of 48 patients (25%). For 9 aneurysms (18.7%) MDS exposed a relevant stenosis of an adjacent vessel induced by clip positioning that had escaped detection by visual inspection. Clip repositioning resulted in complete occlusion of the aneurysms in 7 of 9 cases (14.5%). In two cases, additional wrapping became necessary as it was not possible to achieve complete clipping. The mean duration of MDS investigations was 5.3 minutes. There were no complications of intraoperative MDS probe use. Conclusion: Intraoperative MDS should. be used routinely in cerebral aneurysm surgery, especially for large, complicated and giant aneurysms. Intraoperative MDS is a feasible, safe, and very reliable technique in aneurysm surgery. This technique is a valuable tool, in many instances, in place of intraoperative angiography for the surgical treatment of aneurysms.
引用
下载
收藏
页码:312 / 316
页数:5
相关论文
共 50 条
  • [1] Intraoperative microvascular Doppler sonography in aneurysm surgery
    Bailes, JE
    Tantuwaya, LS
    Fukushima, T
    Schurman, GW
    Davis, D
    NEUROSURGERY, 1997, 40 (05) : 965 - 970
  • [2] Practicability of intraoperative microvascular Doppler sonography in aneurysm surgery
    Firsching, R
    Synowitz, HJ
    Hanebeck, J
    MINIMALLY INVASIVE NEUROSURGERY, 2000, 43 (03) : 144 - 148
  • [3] Intraoperative microvascular Doppler sonography in aneurysm surgery - Comments
    Batjer, HH
    Ogilvy, CS
    Barrow, DL
    NEUROSURGERY, 1997, 40 (05) : 970 - 972
  • [4] Intraoperative microvascular Doppler in intracranial aneurysm surgery
    Bailes, JE
    SURGICAL NEUROLOGY, 2005, 63 (04): : 342 - 342
  • [5] Intraoperative Doppler and Duplex sonography in cerebral aneurysm surgery
    Heiroth, Hi-Jae
    Etminan, Nima
    Steiger, Hans-Jakob
    Haenggi, Daniel
    BRITISH JOURNAL OF NEUROSURGERY, 2011, 25 (05) : 586 - 590
  • [6] MICROVASCULAR INTRAOPERATIVE DOPPLER SONOGRAPHY
    GILSBACH, JM
    ULTRASCHALL IN DER MEDIZIN, 1984, 5 (05): : 246 - 254
  • [7] Intraoperative microvascular Doppler ultrasonography in cerebral aneurysm surgery
    Stendel, R
    Pietilä, T
    Al Hassan, AA
    Schilling, A
    Brock, M
    JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2000, 68 (01): : 29 - 35
  • [8] Intraoperative microvascular Doppler monitoring in intracranial aneurysm surgery
    Hui Pin-jing
    Yan Yan-hong
    Zhang Shi-ming
    Wang Zhong
    Yu Zheng-quan
    Zhou You-xin
    Li Xiang-dong
    Cui Gang
    Zhou Dai
    Hui Guo-zhen
    Lan Qing
    CHINESE MEDICAL JOURNAL, 2013, 126 (13) : 2424 - 2429
  • [9] Intraoperative microvascular Doppler monitoring in intracranial aneurysm surgery
    HUI Pin-jing
    YAN Yan-hong
    ZHANG Shi-ming
    WANG Zhong
    YU Zheng-quan
    ZHOU You-xin
    LI Xiang-dong
    CUI Gang
    ZHOU Dai
    HUI Guo-zhen
    LAN Qing
    中华医学杂志(英文版), 2013, 126 (13) : 2424 - 2429
  • [10] Use of microvascular Doppler sonography in aneurysm surgery on the anterior choroidal artery
    Shibata, Y
    Fujita, S
    Kawaguchi, T
    Hosoda, K
    Komatsu, H
    Tamaki, N
    NEUROLOGIA MEDICO-CHIRURGICA, 2000, 40 (01): : 30 - 35