Johnson noise thermometry

被引:46
|
作者
Qu, J. F. [1 ]
Benz, S. P. [2 ]
Rogalla, H. [2 ,3 ]
Tew, W. L. [4 ]
White, D. R. [5 ]
Zhou, K. L. [1 ]
机构
[1] Natl Inst Metrol, Beijing, Peoples R China
[2] Natl Inst Stand & Technol, Boulder, CO USA
[3] Univ Colorado, ECEE Dept, Boulder, CO 80309 USA
[4] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[5] Measurement Stand Lab New Zealand, Lower Hutt, New Zealand
基金
国家重点研发计划;
关键词
Johnson noise; temperature measurement; noise thermometer; LOW-TEMPERATURE SCALE; ZINC FREEZING-POINT; BOLTZMANN CONSTANT; THERMODYNAMIC TEMPERATURE; THERMAL AGITATION; SHOT-NOISE; ELECTRONIC MEASUREMENT; SYSTEMATIC-ERRORS; K; RESOLUTION;
D O I
10.1088/1361-6501/ab3526
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Johnson noise thermometers infer thermodynamic temperature from measurements of the thermally-induced current fluctuations that occur in all electrical conductors. This paper reviews the status of Johnson noise thermometry and its prospects for both metrological measurements and for practical applications in industry. The review begins with a brief description of the foundations and principles of Johnson noise thermometry before outlining the many different techniques and technological breakthroughs that have enabled the application of Johnson noise thermometry to high-accuracy, cryogenic, and industrial thermometry. Finally, the future of noise thermometry is considered. As the only purely electronic approach to thermodynamic temperature measurement, Johnson noise thermometry has appeal for metrological applications at temperatures ranging from below 1 mK up to 800 K. With the rapid advances in digital technologies, there are also expectations that noise thermometry will become a practical option for some industrial applications, perhaps reaching temperatures above 2000 K.
引用
收藏
页数:26
相关论文
共 50 条
  • [1] The status of Johnson noise thermometry
    White, DR
    Galleano, R
    Actis, A
    Brixy, H
    DeGroot, M
    Dubbeldam, J
    Reesink, AL
    Edler, F
    Sakurai, H
    Shepard, RL
    Gallop, JC
    METROLOGIA, 1996, 33 (04) : 325 - 335
  • [2] NEW METHOD OF JOHNSON NOISE THERMOMETRY
    BORKOWSKI, CJ
    BLALOCK, TV
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1974, 45 (02): : 151 - 162
  • [3] Johnson noise thermometry for harsh environments
    Kisner, R
    Britton, CL
    Jagadish, U
    Wilgen, JB
    Roberts, M
    Blalock, TV
    Holcomb, D
    Bobrek, M
    Ericson, MN
    2004 IEEE AEROSPACE CONFERENCE PROCEEDINGS, VOLS 1-6, 2004, : 2586 - 2596
  • [4] Non-linearity in Johnson noise thermometry
    White, D. R.
    METROLOGIA, 2012, 49 (06) : 651 - 665
  • [5] An AC Josephson source for Johnson noise thermometry
    Benz, SP
    Martinis, JM
    Dresselhaus, PD
    Nam, SW
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2003, 52 (02) : 545 - 549
  • [6] Practical realisation of the kelvin by Johnson noise thermometry
    Benz, Samuel P.
    Coakley, Kevin J.
    Flowers-Jacobs, Nathan E.
    Rogalla, Horst
    Tew, Weston L.
    Qu, Jifeng
    White, D. Rod
    Gaiser, Christof
    Pollarolo, Alessio
    Urano, Chiharu
    METROLOGIA, 2024, 61 (02)
  • [7] An ac Josephson source for Johnson noise thermometry'
    Benz, SP
    Martinis, JM
    Dresselhaus, PD
    Nam, SW
    2002 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CONFERENCE DIGEST, 2002, : 436 - 437
  • [8] Improved spectral aberration in Johnson Noise Thermometry
    Pollarolo, A.
    Rogalla, H.
    Fox, A.
    Coakley, K. J.
    Tew, W. L.
    Benz, S. P.
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [9] CORRELATION METHOD ERRORS IN JOHNSON NOISE THERMOMETRY
    Callegaro, L.
    Pisani, M.
    Ortolano, M.
    D'Elia, V.
    Manta, F.
    2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 504 - +
  • [10] Improvements in the NIST Johnson Noise Thermometry System
    Benz, Samuel P.
    Qu, Jifeng
    Rogalla, Horst
    White, D. Rod
    Dresselhaus, Paul D.
    Tew, Weston L.
    Nam, Sae Woo
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (04) : 884 - 890