A new central venous catheter cap: decreased microbial growth and risk for catheter-related bloodstream infection

被引:7
|
作者
Buchman, A. L. [1 ]
Spapperi, J. [2 ]
Leopold, P. [2 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Div Gastroenterol, Chicago, IL 60611 USA
[2] Med Murray Inc, N Barrington, IL USA
来源
JOURNAL OF VASCULAR ACCESS | 2009年 / 10卷 / 01期
关键词
Central venous catheter; Catheter cap; Catheter-related blood stream infection; Infection; Sepsis; HOME PARENTERAL-NUTRITION; CRITICALLY-ILL PATIENTS; HUB COLONIZATION; POVIDONE-IODINE; PREVENTION; SEPSIS; CARE; SYSTEM; TRIAL; BACTEREMIA;
D O I
10.1177/112972980901000103
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: Catheter-related blood stream infection (CRBI) is a major cause of morbidity and mortality, and is a source of significant healthcare expenditures in patients that require central venous catheters for intravenous nutrition, chemotherapy, and other products. The source of many catheter-related infections is contamination of the catheter hub. Herein an antimicrobial catheter cap, the AB Cap is described. Methods: The AB Cap device is a catheter cleaning device designed to keep needleless; luer valves clean by encapsulating them in a cleaning solution. This device was evaluated using an in vitro model of hub contamination with Staphylococcus aureus, Staphylococcus epidermidis (S. epidermidis), Klebsiella pneumonia (K pneumonia), Pseudomonas aeruginosa, Escherichia coli and Candida. albicans (C. albicans). Following hub contamination on days 1, 3, 5 and 7, saline was infused through the AB Cap and effluent collected from the efferent end. The effluent fluid was cultured for the index organisms, and allowed to incubate in culture for up to 7 days. Negative control caps were not contaminated and positive controls lacked cleaning solution and were contaminated. Results: Microbial growth developed for all index organisms, and generally within 1 day of culture growth following the first day of contamination (day 1) in effluent from all positive controls, while no growth occurred in effluent from negative controls. No growth of any organism occurred in any of the test items after the first day of contamination. Growth of three organisms was detected in two of the three test AB Caps following contamination day 3, after 1-4 days of incubation. All organisms could be cultured in the effluent from two of the three test items at contamination day 5, generally by the second day of incubation. One test item remained free of growth for the entire test period except for one organism. By day 7, this particular test item grew an additional organism and the testing was concluded. All positive growth test items remained positive on subsequent inoculations during culture of newly obtained effluent with the exception of test item A, from which effluent following inoculation on day 3 showed growth of S. epidermidis and K pneumonia, but no growth for these organisms from effluent obtained on inoculation day 5. In addition, effluent from test item C showed growth of C. albicans from inoculation day 5, but no growth from effluent obtained on inoculation day 7. The growth of S. epidermidis from effluent of test item A from the day 3 inoculation, and C albicans from effluent of test items B and C did not occur until day 4 of incubation, suggesting a very small amount of contamination. Conclusion: An antimicrobial catheter cap is not a complete substitute for a proper catheter cleaning technique and other anti-infection precautions. However, we describe a unique catheter cap that significantly decreased the likelihood of a catheter-related infection from a non-cleaned cap in an in vitro model. (J Vasc Access 2009; 10: 11-21)
引用
收藏
页码:11 / 21
页数:11
相关论文
共 50 条
  • [1] Risk factors for central venous catheter-related bloodstream infection in neonates
    Ferreira, Janita
    Moreira Camargos, Paulo Augusto
    Rosado, Viviane
    Orlandi Mourao, Paulo Henrique
    de Castro Romanelli, Roberta Maia
    [J]. AMERICAN JOURNAL OF INFECTION CONTROL, 2020, 48 (09) : 1102 - 1103
  • [2] Catheter-related bloodstream infection according to central venous accesses
    L Lorente Ramos
    C García
    J Iribarren
    J Castedo
    J Jiménez
    M Brouard
    C Henry
    R Santacreu
    M Martín
    M Mora
    [J]. Critical Care, 12 (Suppl 2):
  • [3] Central Venous Catheter-Related Bloodstream Infection and Cryptococcus neoformans
    Tuon, Felipe F.
    Morales, Hugo M. P.
    Penteado-Filho, Sergio R.
    da-Silva, Margarete M.
    de Quadros, Isabel
    El Hamoui, Amina
    [J]. BRAZILIAN JOURNAL OF INFECTIOUS DISEASES, 2009, 13 (04): : 317 - 318
  • [4] Impact of oligon central venous catheters on catheter colonization and catheter-related bloodstream infection
    Ranucci, M
    Isgrò, G
    Giomarelli, PP
    Pavesi, M
    Luzzani, A
    Cattabriga, I
    Carli, M
    Giomi, P
    Compostella, A
    Digito, A
    Mangani, V
    Silvestri, V
    Mondelli, E
    [J]. CRITICAL CARE MEDICINE, 2003, 31 (01) : 52 - 59
  • [5] Femoral Venous Catheterization Is a Major Risk Factor for Central Venous Catheter-Related Bloodstream Infection
    Ishizuka, Mitsuru
    Nagata, Hitoshi
    Takagi, Kazutoshi
    Kubota, Keiichi
    [J]. JOURNAL OF INVESTIGATIVE SURGERY, 2009, 22 (01) : 16 - 21
  • [6] Dressing Change Reduces the Central Venous Catheter-Related Bloodstream Infection
    Ishizuka, Mitsuru
    Nagata, Hitoshi
    Takagi, Kazutoshi
    Kubota, Keiichi
    [J]. HEPATO-GASTROENTEROLOGY, 2011, 58 (112) : 1882 - 1886
  • [7] Central venous catheter-related bloodstream infection - prevention, records and auditing
    AV Varela
    P Vala
    M Dykyy
    R Almeida
    J Arduán
    A Gonçalves
    MA Pereira
    I Hubert
    C Granja
    [J]. Intensive Care Medicine Experimental, 3 (Suppl 1)
  • [8] Central venous catheter-related infection
    M Mer
    AG Duse
    J Galpin
    R Taylor
    GA Richards
    [J]. Critical Care, 6 (Suppl 1):
  • [9] Catheter-related bloodstream infection rate according to the change of central venous catheter dressing protocols
    Park, HR
    Ki, YH
    Kim, SH
    Song, BE
    Cho, HS
    Ban, JY
    Youn, JH
    Bae, YY
    Kim, KS
    Kim, KM
    Hong, EY
    [J]. BONE MARROW TRANSPLANTATION, 2005, 35 : S320 - S321
  • [10] Prevention of central venous catheter-related bloodstream infections
    Rupp, ME
    Craig, R
    [J]. INFECTIONS IN MEDICINE, 2004, 21 (03) : 123 - 127