Integrated refractive index optical ring resonator detector for capillary electrophoresis

被引:92
|
作者
Zhu, Hongying
White, Ian M.
Suter, Jonathan D.
Zourob, Mohammed
Fan, Xudong
机构
[1] Univ Missouri, Dept Biol Engn, Columbia, MO 65211 USA
[2] Univ Cambridge, Inst Biotechnol, Cambridge CB2 1QT, England
关键词
D O I
10.1021/ac061279q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We developed a novel miniaturized and multiplexed, on-capillary, refractive index (RI) detector using liquid core optical ring resonators (LCORRs) for future development of capillary electrophoresis (CE) devices. The LCORR employs a glass capillary with a diameter of similar to 100 mu m and a wall thickness of a few micrometers. The circular cross section of the capillary forms a ring resonator along which the light circulates in the form of the whispering gallery modes (WGMs). The WGM has an evanescent field extending into the capillary core and responds to the RI change due to the analyte conducted in the capillary, thus permitting label-free measurement. The resonating nature of the WGM enables repetitive light-analyte interaction, significantly enhancing the LCORR sensitivity. This LCORR architecture achieves dual use of the capillary as a sensor head and a CE fluidic channel, allowing for integrated, multiplexed, and noninvasive on-capillary detection at any location along the capillary. In this work, we used electro-osmotic flow and glycerol as a model system to demonstrate the fluid transport capability of the LCORRs. In addition, we performed flow speed measurement on the LCORR to demonstrate its flow analysis capability. Finally, using the LCORR's label-free sensing mechanism, we accurately deduced the analyte concentration in real time at a given point on the capillary. A sensitivity of 20 nm/RIU (refractive index units) was observed, leading to an RI detection limit of 10(-6) RIU. The LCORR marries photonic technology with microfluidics and enables rapid on-capillary sample analysis and flow profile monitoring. The investigation in this regard will open a door to novel high-throughput CE devices and lab-on-a-chip sensors in the future.
引用
收藏
页码:930 / 937
页数:8
相关论文
共 50 条
  • [1] ON-COLUMN LASER-BASED REFRACTIVE-INDEX DETECTOR FOR CAPILLARY ELECTROPHORESIS
    BRUNO, AE
    KRATTIGER, B
    MAYSTRE, F
    WIDMER, HM
    ANALYTICAL CHEMISTRY, 1991, 63 (23) : 2689 - 2697
  • [2] Integrated on-capillary electrochemical detector for capillary electrophoresis
    Zhong, Min
    Lunte, Susan M.
    1600, Kluwer Academic Publishers, Dordrecht, Netherlands (18):
  • [3] Integrated on-capillary electrochemical detector for capillary electrophoresis
    Zhong, M
    Lunte, SM
    ANALYTICAL CHEMISTRY, 1996, 68 (15) : 2488 - 2493
  • [4] GeAsSe chalcogenide slot optical waveguide ring resonator for refractive index sensing
    Ashok, N.
    Lee, Yeung Lak
    Shin, WooJin
    2017 25TH INTERNATIONAL CONFERENCE ON OPTICAL FIBER SENSORS (OFS), 2017, 10323
  • [5] Design and optimization of liquid core optical ring resonator for refractive index sensing
    Lin, Nai
    Jiang, Lan
    Wang, Sumei
    Xiao, Hai
    Lu, Yongfeng
    Tsai, Hai-Lung
    APPLIED OPTICS, 2011, 50 (20) : 3615 - 3621
  • [6] Dependence of output power in modulated optical fiber ring resonator on nonlinear refractive index
    Takehito Suzuki
    Shi Wei
    Yoh Imai
    Optical Review, 2010, 17 : 327 - 330
  • [7] Sensitive Refractive Index Detection Using a Broad-Band Optical Ring Resonator
    Oates, Thomas C.
    Burgess, Lloyd W.
    ANALYTICAL CHEMISTRY, 2012, 84 (18) : 7713 - 7720
  • [8] Dependence of output power in modulated optical fiber ring resonator on nonlinear refractive index
    Suzuki, Takehito
    Wei, Shi
    Imai, Yoh
    OPTICAL REVIEW, 2010, 17 (03) : 327 - 330
  • [9] Refractive index biosensor based on topological ring resonator
    Liu, Na
    Wang, Shutao
    Lv, Jiangtao
    Lu, Yuhong
    Zhang, Jinqing
    OPTICS COMMUNICATIONS, 2023, 541
  • [10] HOLOGRAM-BASED REFRACTIVE-INDEX DETECTOR FOR CAPILLARY ELECTROPHORESIS - SEPARATION OF METAL-IONS
    KRATTIGER, B
    BRUIN, GJM
    BRUNO, AE
    ANALYTICAL CHEMISTRY, 1994, 66 (01) : 1 - 8