Accuracy of enhanced transcutaneous bilirubinometry according to various measurement sites

被引:9
|
作者
Lucanova, Lucia Casnocha [1 ,2 ]
Zibolenova, Jana [3 ]
Matasova, Katarina [1 ,2 ]
Docekalova, Lenka [1 ,2 ]
Zibolen, Mirko [1 ,2 ]
机构
[1] Univ Hosp Martin, Dept Neonatol, Martin, Slovakia
[2] Comenius Univ, Jessenius Fac Med Martin, Bratislava, Slovakia
[3] Comenius Univ, Jessenius Fac Med Martin, Dept Publ Hlth, Bratislava, Slovakia
来源
TURKISH ARCHIVES OF PEDIATRICS | 2021年 / 56卷 / 01期
关键词
Bilirubin; hyperbilirubinemia; jaundice; neonates; transcutaneous bilirubinometry; HOUR-SPECIFIC NOMOGRAM; SERUM BILIRUBIN; LATE PRETERM; SIGNIFICANT HYPERBILIRUBINEMIA; HEALTHY TERM; NEWBORNS; PREDICTION; BILICARE; INFANTS;
D O I
10.14744/TurkPediatriArs.2020.54514
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Objective: The goal of the study was to provide missing data on the accuracy of enhanced transcutaneous bilirubinometry in a monoracial population of term neonates, considering three different measurement sites. Material and Methods: Transcutaneous bilirubin was measured using the JM-105 device on the forehead, chest, and abdomen. Blood sampling for total serum bilirubin concentration has been performed within 10 minutes of transcutaneous measurements. Paired transcutaneous bilirubin and total serum bilirubin measurements were statistically analyzed. Results: The study group consisted of 102 healthy term Slovak infants. The correlation between total serum bilirubin and transcutaneous bilirubin was significant (coefficient of determination R2: 0.9045 forehead, 0.8808 sternum, 0.8467 abdomen). Transcutaneous measurements underestimated serum bilirubin levels significantly when total serum bilirubin values were higher than 15 mg/dL, irrespective of the site of transcutaneous measurements. The lowest mean difference between total serum bilirubin and transcutaneous bilirubin was identified on the sternum (median: -1.1 mg/dL). The area under the curve was >0.97 and >0.93 for detecting total serum bilirubin levels >10 mg/dL and >13 mg/dL, respectively, for all measurement sites. Transcutaneous measurements on the forehead and sternum provided very high sensitivity, with the best performance at the forehead. Conclusion: Transcutaneous bilirubinometry using an enhanced device is an accurate, sensitive, and convenient screening method in term Caucasian neonates. Transcutaneous bilirubin measurements on the forehead, sternum, and abdomen are reliable, with the best performance on the forehead. It is necessaryto confirm higher transcutaneous bilirubin values with a total serum bilirubin measurement.
引用
收藏
页码:15 / 21
页数:7
相关论文
共 50 条
  • [1] The accuracy of transcutaneous bilirubinometry in preterm infants
    Arman, Didem
    Topcuoglu, Sevilay
    Gursoy, Tugba
    Ovali, Fahri
    Karatekin, Guner
    JOURNAL OF PERINATOLOGY, 2020, 40 (02) : 212 - 218
  • [2] The accuracy of transcutaneous bilirubinometry in preterm infants
    Didem Arman
    Sevilay Topcuoğlu
    Tugba Gürsoy
    Fahri Ovalı
    Güner Karatekin
    Journal of Perinatology, 2020, 40 : 212 - 218
  • [3] Machine Learning to Improve Accuracy of Transcutaneous Bilirubinometry
    Morimoto, Daisaku
    Washio, Yosuke
    Fukuda, Kana
    Sato, Takeshi
    Okamura, Tomoka
    Watanabe, Hirokazu
    Yoshimoto, Junko
    Tanioka, Maki
    Tsukahara, Hirokazu
    NEONATOLOGY, 2024, 121 (06) : 772 - 779
  • [4] Diagnosis Accuracy of Transcutaneous Bilirubinometry in Very Preterm Newborns
    Rubio, Amandine
    Epiard, Chloe
    Gebus, Maya
    Deiber, Michel
    Samperiz, Sylvain
    Genty, Celine
    Ego, Anne
    Debillon, Thieny
    NEONATOLOGY, 2017, 111 (01) : 1 - 7
  • [5] Accuracy and precision of transcutaneous bilirubinometry in postdischarge Asian neonates
    Kitsommart, Ratchada
    Pornladnun, Pornpat
    Chomchai, Chulathida
    Urujchutchairut, Pornvilai
    Paes, Bosco
    EUROPEAN JOURNAL OF PEDIATRICS, 2013, 172 (06) : 781 - 786
  • [6] Accuracy and precision of transcutaneous bilirubinometry in postdischarge Asian neonates
    Ratchada Kitsommart
    Pornpat Pornladnun
    Chulathida Chomchai
    Pornvilai Urujchutchairut
    Bosco Paes
    European Journal of Pediatrics, 2013, 172 : 781 - 786
  • [8] Transcutaneous bilirubinometry versus total serum bilirubin measurement for newborns
    Okwundu, Charles I.
    Olowoyeye, Abiola
    Uthman, Olalekan A.
    Smith, Johan
    Wiysonge, Charles S.
    Bhutani, Vinod K.
    Fiander, Michelle
    Gautham, Kanekal S.
    COCHRANE DATABASE OF SYSTEMATIC REVIEWS, 2023, (05):
  • [9] Accuracy of transcutaneous bilirubinometry in the preterm infants: a comprehensive meta-analysis
    Shabuj, Mohammad Hassan
    Hossain, Jesmin
    Dey, Sanjoy
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2019, 32 (05): : 734 - 741
  • [10] Accuracy of transcutaneous bilirubinometry in term infants after phototherapy: a prospective observational study
    Joy, Charissa
    Collyer, Taya
    Mcmahon, Kathryn
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2024, 37 (01):