Clinical Results of the Wear Performance of Cross-Linked Polyethylene in Total Hip Arthroplasty Prospective Randomized Trial

被引:22
|
作者
Ise, Kentaro [1 ]
Kawanabe, Keiichi [1 ]
Tamura, Jiro [1 ]
Akiyama, Haruhiko [1 ]
Goto, Koji [1 ]
Nakamura, Takashi [1 ]
机构
[1] Kyoto Univ, Dept Orthoped Surg, Sakyo Ku, Kyoto 6068507, Japan
来源
JOURNAL OF ARTHROPLASTY | 2009年 / 24卷 / 08期
关键词
wear; cross-linked polyethylene; total hip arthroplasty; clinical result; prospective randomized trial; FOLLOW-UP; FEMORAL-HEAD; TOTAL JOINT; PENETRATION; OSTEOLYSIS; ALUMINA; SURFACE;
D O I
10.1016/j.arth.2009.05.020
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
To investigate the clinical results of cross-linked polyethylene (CLPE) and to compare the CLPE wear against zirconia and stainless steel heads, we studied the radiographic wear after a minimum 3-year follow-up in total hip arthroplasty (THA). Ninety-four hips were randomly implanted with a 22.225-mm head cemented THA-the group of non-CLPE against zirconia and CLPE against 2 different zirconias and stainless steel. The linear wear rate was significantly lower in the group of CLPE against zirconia (0.067, 0.059 mm/y) and against stainless steel (0.068 mm/y) compared with non-CLPE against zirconia (0.170 mm/y). in the short-term results, the wear performance of CLPE against zirconia was superior to that of non-CLPE; however, it did not show a better wear rate than CLPE against stainless steel. Furthermore, long-term investigations will be necessary for understanding CLPE wear in vivo.
引用
收藏
页码:1216 / 1220
页数:5
相关论文
共 50 条
  • [1] Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: A randomized prospective trial
    Martell, JM
    Verner, JJ
    Incavo, SJ
    JOURNAL OF ARTHROPLASTY, 2003, 18 (07): : 55 - 59
  • [2] Wear Rate of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty A Randomized Controlled Trial
    McCalden, Richard W.
    MacDonald, Steven J.
    Rorabeck, Cecil H.
    Bourne, Robert B.
    Chess, David G.
    Charron, Kory D.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (04): : 773 - 782
  • [3] Creep and Wear in Vitamin E-Infused Highly Cross-Linked Polyethylene Cups for Total Hip Arthroplasty A Prospective Randomized Controlled Trial
    Rochcongar, G.
    Buia, G.
    Bourroux, E.
    Dunet, J.
    Chapus, V.
    Hulet, C.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2018, 100 (02): : 107 - 114
  • [4] Unexpected Wear of a Uniquely Designed Moderately Cross-Linked Polyethylene in Total Hip Arthroplasty
    Yakkanti, Ramakanth R.
    Ocksrider, Justin L.
    Patel, Anand A.
    Kolevar, Matthew P.
    Moore, Rebecca D.
    Rimnac, Clare M.
    Kraay, Matthew J.
    Wright, Timothy M.
    Baral, Elexis C.
    Robinson, Raymond P.
    JOURNAL OF ARTHROPLASTY, 2022, 37 (06): : 1130 - 1135
  • [5] Does highly cross-linked polyethylene wear less than conventional polyethylene in total hip arthroplasty?
    Glyn-Jones, Sion
    Saac, Sherif
    Hauptfleisch, Jennifer
    McLardy-Smith, Peter
    Murray, David W.
    Singh, Harinderjit
    JOURNAL OF ARTHROPLASTY, 2008, 23 (03): : 337 - 343
  • [6] Conventional Versus Cross-Linked Polyethylene for Total Hip Arthroplasty
    Surace, Michele F.
    Monestier, Luca
    Vulcano, Ettore
    Harwin, Steven F.
    Cherubino, Paolo
    ORTHOPEDICS, 2015, 38 (09) : 556 - 561
  • [7] Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty
    Jacobs, Cale A.
    Christensen, Christian P.
    Greenwald, A. Seth
    McKellop, Harry
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (12): : 2779 - 2786
  • [8] Clinical performance of highly cross-linked Polyethylenes in total hip arthroplasty
    不详
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (03): : 628 - 628
  • [9] A Double-Blind, Prospective, Randomized Controlled Trial Comparing Highly Cross-Linked and Conventional Polyethylene in Primary Total Hip Arthroplasty
    Calvert, George T.
    Devane, Peter A.
    Fielden, Jan
    Adams, Kathryn
    Horne, J. Geoffrey
    JOURNAL OF ARTHROPLASTY, 2009, 24 (04): : 505 - 510
  • [10] Manufacturing, oxidation, mechanical properties and clinical performance of highly cross-linked polyethylene in total hip arthroplasty
    Singh, Gurpreet
    Klassen, Robert
    Howard, James
    Naudie, Douglas
    Teeter, Matthew
    Lanting, Brent
    HIP INTERNATIONAL, 2018, 28 (06) : 573 - 583