Depth-of-interaction study of a dual-readout detector based on TOFPET2 application-specific integrated circuit

被引:24
|
作者
Li, Mohan [1 ]
Abbaszadeh, Shiva [1 ]
机构
[1] Univ Illinois, Dept Nucl Plasma & Radiol Engn, Urbana, IL 61801 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2019年 / 64卷 / 17期
基金
美国国家卫生研究院;
关键词
dual-ended readout; PET; depth-of-interaction; TOFPET2; ASIC; PET DETECTOR; ENDED READOUT; PERFORMANCE; SCANNER; DESIGN; SYSTEM; PHOTODIODE; MM;
D O I
10.1088/1361-6560/ab3866
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Depth-of-interaction (DOI) capability is important for achieving high spatial resolution and sensitivity in dedicated organ and small animal positron emission tomography (PET) scanners. The dual-ended readout is one of the common methods that can achieve good DOI resolution. The aim of this study is to evaluate a dual-ended readout detector based on silicon photomultiplier (SiPM) and TOFPET2 application-specific integrated circuit (ASIC). The detector is based on 4 x 4 lutetium-yttrium oxyorthosilicate (LYSO) units, each unit contained 6 x 6 LYSO crystals, and the crystal size was 1 x 1 x 20 min(3). The four lateral surfaces of LYSO crystals were mechanically ground to W14 (surface roughness 10-14 mu m), and the two ended surfaces were polished (surface roughness <0.5 mu m). The reflector was Toray Lumirror E60, and the packing fraction of the LYSO block was 86.5%. Each LYSO unit was read out from both ends with two Hamamatsu S13361-3050AE-08 SiPM arrays. The analog output signals of SiPM were digitized by PETsys TOFPET2 ASIC and acquired by PETsys SiPM Readout System. The ASIC and SiPM were cooled by a fan and a Peltier element. To investigate the crystal resolvability, different light guide thicknesses including 0.8,1,1.2 and 2 min were tested. The light guide was made of optical glass (H-K9L-Foctek Photoincs), and the size and refractive index were 6.45 x 6.45 mm(2) and 1.53 (at 420 nm), respectively. To characterize the detector performance at different depths, another 1 x 25.8 x 20 mm 3 single LYSO slab was used. Data were acquired at 10 depths (1, 3, ... ,19 mm), and each depth had a 10 min acquisition time and about 40 thousand coincidence events. During the experiment, the SiPM temperature was controlled as 27.6 +/- 0.4 degrees C. The results showed that the 1.2 mm light guide offered the best crystal resolvability. The energy, coincidence time, and DOI resolution full-width at half-maximum of the detector were characterized as 15.66% +/- 0.66%, 602.98 +/- 10.58 ps, and 2.33 +/- 0.07 mm, respectively. The good DOI resolution indicates the potential of utilizing the detector for high-resolution PET applications.
引用
收藏
页数:8
相关论文
共 16 条
  • [1] Evaluation of high-resolution and depth-encoding PET detector modules based on single-ended readout with TOFPET2 ASIC
    Wang, Mingchen
    Wang, Yonggang
    Wang, Liwei
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2021, 5 (03) : 451 - 458
  • [2] Evaluation of high-resolution and depth-encoding PET detector modules based on single-ended readout with TOFPET2 ASIC
    Mingchen Wang
    Yonggang Wang
    Liwei Wang
    Radiation Detection Technology and Methods, 2021, 5 : 451 - 458
  • [3] Calibration Methodology for a Dual-ended Readout Silicon Photomultiplier Based Depth-of-interaction PET Detector Module
    Xi, W.
    McKisson, J. E.
    Weisenberger, A. G.
    Lee, S.
    Taylor, M.
    Stepanyan, A.
    Zorn, C.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3493 - 3495
  • [4] Choice of crystal surface finishing for a dual-ended readout depth-of-interaction (DOI) detector
    Fan, Peng
    Ma, Tianyu
    Wei, Qingyang
    Yao, Rutao
    Liu, Yaqiang
    Wang, Shi
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (03): : 1041 - 1056
  • [5] Depth-of-interaction positron emission tomography detector with 45° tilted silicon photomultipliers using dual-ended signal readout
    Seo, Minjee
    Park, Haewook
    Lee, Seungeun
    Ko, Guen Bae
    Lee, Jae Sung
    MEDICAL PHYSICS, 2023, 50 (07) : 4112 - 4121
  • [6] A depth-of-interaction encoding PET detector module with dual-ended readout using large-area silicon photomultiplier arrays
    Du, Junwei
    Bai, Xiaowei
    Cherry, Simon R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (24):
  • [7] Preliminary Results of a Depth-of-Interaction based PET System with Dual Ended Readout for Combined Breast PET/CT Imaging
    Judenhofer, Martin
    Sumanasena, M. G. Buddika
    Gong, Kuang
    Peng, Qiyu
    Burkett, George
    Spencer, Benjamin
    Qi, Jinyi
    Badawi, Ramsey
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [8] Test system of the front-end readout for an application-specific integrated circuit for the water Cherenkov detector array at the large high-altitude air shower observatory
    Er-Lei Chen
    Lei Zhao
    Li Yu
    Jia-Jun Qin
    Yu Liang
    Shu-Bin Liu
    Qi An
    Nuclear Science and Techniques, 2017, 28
  • [9] Test system of the front-end readout for an application-specific integrated circuit for the water Cherenkov detector array at the large high-altitude air shower observatory
    Chen, Er-Lei
    Zhao, Lei
    Yu, Li
    Qin, Jia-Jun
    Liang, Yu
    Liu, Shu-Bin
    An, Qi
    NUCLEAR SCIENCE AND TECHNIQUES, 2017, 28 (06)
  • [10] Test system of the front-end readout for an application-specific integrated circuit for the water Cherenkov detector array at the large high-altitude air shower observatory
    Er-Lei Chen
    Lei Zhao
    Li Yu
    Jia-Jun Qin
    Yu Liang
    Shu-Bin Liu
    Qi An
    NuclearScienceandTechniques, 2017, 28 (06) : 53 - 62