A regenerative biosensing surface in microfluidics using electrochemical desorption of short-chain self-assembled monolayer

被引:28
|
作者
Choi, Seokheun [1 ]
Chae, Junseok [1 ]
机构
[1] Arizona State Univ, Dept Elect Engn, Tempe, AZ 85287 USA
基金
美国国家科学基金会;
关键词
Self-assembled monolayer; Electrochemical reductive desorption; Biosensor Regenerative; Surface plasmon resonance; N-ALKANETHIOL MONOLAYERS; IN-SITU; REDUCTIVE DESORPTION; THIOCTIC ACID; THIOL FILMS; AU(111); GOLD; ADSORPTION; KINETICS; RECOGNITION;
D O I
10.1007/s10404-009-0440-7
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
We report a novel method to regenerate a biosensor surface in microfluidics. By applying a low DC voltage (0.9 V) between two electrodes submerged in phosphate buffered saline, the sensing surface resets to be reusable and reconfigurable; streptavidin-bound COOH-SAM completely desorbs and CH3-terminated self-assembled monolayer (SAM) forms on the sensing surface to capture the subsequent target molecule, fibrinogen, in a microfluidic device. The biomolecular interactions are monitored by surface plasmon resonance in real time, and ellipsometry and linear sweep voltammetry are used to evaluate the results. Despite much study on the theoretical mechanism of electrochemical SAM desorption, relatively little research has been carried out on its full integration into a microfluidic system. This is because of electrode peeling-off and electrolysis occurring at a similar potential to the potential of the SAM desorption. In this paper, we report that the potential for the reductive desorption of thiol SAMs depends on the length of the alkyl chin, the type of terminal groups and the binding of proteins and that our approach using short-chain SAMs (n < 3) can be a good candidate to minimize these limitations. While the surface modified by proteins on the long-chain SAMs (n > 10) needs more than 1.0 V between two electrodes to be completely regenerated, the protein-bound surface on the short-chain SAMs (n < 3) does around 0.9-DC voltage. Linear sweep voltammetry demonstrates that hydrogen evolution (approx. -1.2 V) does not overlap the short-chain SAM desorption and the electrode does not peel off during the desorption process as well. It is shown that the modified proteins in the same microfluidic device are still stable even after 10 cycles showing a relative standard deviation lower than 1.86%.
引用
收藏
页码:819 / 827
页数:9
相关论文
共 50 条
  • [1] A regenerative biosensing surface in microfluidics using electrochemical desorption of short-chain self-assembled monolayer
    Seokheun Choi
    Junseok Chae
    Microfluidics and Nanofluidics, 2009, 7
  • [2] Modified fabrication process of protein chips using a short-chain self-assembled monolayer
    Jang, Ling-Sheng
    Keng, Hao-Kai
    BIOMEDICAL MICRODEVICES, 2008, 10 (02) : 203 - 211
  • [3] Modified fabrication process of protein chips using a short-chain self-assembled monolayer
    Ling-Sheng Jang
    Hao-Kai Keng
    Biomedical Microdevices, 2008, 10 : 203 - 211
  • [4] Electrochemical desorption of a self-assembled monolayer of alkanethiol in ionic liquids
    Oyamatsu, Daisuke
    Fujita, Takeshi
    Arimoto, Satoshi
    Munakata, Hirokazu
    Matsumoto, Hajime
    Kuwabata, Susumu
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2008, 615 (02) : 110 - 116
  • [5] Investigation of self-assembled monolayers of two short-chain molecules by the electrochemical titration method
    Zhu Yang-Song
    Hou Dong-Yan
    Shao Hui-Bo
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (02) : 201 - 205
  • [6] Effects of chain-chain interaction on the configuration of short-chain alkanethiol self-assembled monolayers on a metal surface
    Liao, Chia-Li
    Faizanuddin, Syed Mohammed
    Haruyama, Jun
    Liao, Wei-Ssu
    Wen, Yu-Chieh
    JOURNAL OF CHEMICAL PHYSICS, 2024, 160 (21):
  • [7] Impact of Self-Assembled Monolayer Design and Electrochemical Factors on Impedance-Based Biosensing
    Brothers, Michael C.
    Moore, David
    St Lawrence, Michael
    Harris, Jonathan
    Joseph, Ronald M.
    Ratcliff, Erin
    Ruiz, Oscar N.
    Glavin, Nicholas
    Kim, Steve S.
    SENSORS, 2020, 20 (08)
  • [8] Electrochemical Determination of Uranyl Ions Using a Self-Assembled Monolayer
    Becker, Amit
    Tobias, Haim
    Mandler, Daniel
    ANALYTICAL CHEMISTRY, 2009, 81 (20) : 8627 - 8631
  • [9] Self-Spreading of Lipid Bilayer on a Hydrophobic Surface Made by Self-Assembled Monolayer with Short Alkyl Chain
    Omori, Yuya
    Sakaue, Hiroyuki
    Takahagi, Takayuki
    Suzuki, Hitoshi
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2016, 16 (04) : 3426 - 3430
  • [10] Novel electrochemical biosensing platform using self-assembled peptide nanotubes
    Yemini, M
    Reches, M
    Rishpon, J
    Gazit, E
    NANO LETTERS, 2005, 5 (01) : 183 - 186