Is automatic mode switching effective for atrial arrhythmias occurring at different rates? A study of the efficacy of automatic mode and rate switching to simulated atrial arrhythmias by chest wall stimulation

被引:7
|
作者
Leung, SK
Lau, CP [1 ]
Lam, CTF
Tse, HF
Lee, K
Chan, WK
Leung, Z
机构
[1] Univ Hong Kong, Dept Med, Div Cardiol, Queen Mary Hosp, Hong Kong, Hong Kong, Peoples R China
[2] Kwong Wah Hosp, Dept Med, Hong Kong, Hong Kong, Peoples R China
[3] United Christian Hosp, Hong Kong, Hong Kong, Peoples R China
来源
关键词
mode switching; atrial fibrillation; pacing;
D O I
10.1111/j.1540-8159.2000.tb00850.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Automatic mode switching (AMS) is a useful means to avoid rapid ventricular response during atrial fibrillation (AF), but AMS cannot occur if the detected atrial rate during AF is below the mode switching criteria. This may be the result of antiarrhythmic medications, or when the atrial events fall within the atrial blanking period, or if the atrial amplitudes during AF are too small to be sensed. We hypothesize that the addition of an automatic rate switching (ARS) algorithm may complement AMS response during AF with different detected atrial rates. We studied the Marathon DDDR pacemaker (Model 294-09, Intermedics Inc.) with the AMS and ARS algorithms that are independently programmable but can also operate in combination. AF sensed above the AMS rate (160 beats/min) will lead to VDIR pacing, whereas AF below AMS rate will be tracked at an interim rate as dictate by the ARS, at a ventricular response that is 20 beats/min above the sensor indicated rate. Atrial tachyarrhythmias were simulated by chest wall stimulation (CWS). CWS was applied to 33 patients (16 men, 17 women, mean age 69 +/- 11 years) with a Marathon DDDR pacemaker using an external pacer to simulate AF occurring at two rate levels: above the AMS rate (programmed at 160 beats/min) at 180 beats/min and below the AMS rate at 120 beats/min. The maximum, minimum, and mean ventricular rates during CWS in DDDR mode with AIMS alone, ARS alone, and their combination were compared. During CWS at 120 beats/min, the AIMS plus ARS setting showed a mean ventricular rate of 79 +/- 3 beats/min and 124 +/- 14 beats/min in the AIMS setting alone (P < 0.01). With CWS at 180 beats/min, the mean ventricular rate in the AMS plus ARS setting compared to the AMS setting alone was not significantly different. However, the variation in ventricular pacing rate was 7 +/- 14 beats/min in the AMS plus ARS setting and 40 +/- 42 beats/min in the AIMS setting (P < 0.05). In conclusion, AMS is effective for simulated atrial tachyarrhythmias sensed above the AIMS rate. Combined AIMS with ARS is useful to handle simulated atrial tachyarrhythmia at a slower rate and to avoid rate fluctuation during AMS. There is also a possibility that this can be applied to the naturally occurring atrial tachyarrhythmias.
引用
收藏
页码:824 / 831
页数:8
相关论文
共 19 条
  • [1] Improved efficacy of mode switching during atrial fibrillation using automatic atrial sensitivity adjustment
    Lam, CTF
    Lau, CP
    Leung, SK
    Tse, HF
    Ayers, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (01): : 17 - 25
  • [2] Atrial tachyarrhythmias detected by automatic mode switching: Quo vadis?
    Estes, NAM
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (07) : 778 - 779
  • [3] An Unusual Cause of Automatic Mode Switching in the Absence of an Atrial Tachyarrhythmia
    Almehairi, Mohammed
    Baranchuk, Adrian
    Johri, Amer
    Caldwell, Jane
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (06): : 777 - 780
  • [4] How should atrial sensitivity be programmed for optimal automatic mode switching?
    Leung, SK
    Lau, CP
    Lam, C
    Tse, HF
    Chung, F
    Tang, MO
    Ayres, G
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (02) : 57175 - 57175
  • [5] Verification of pacemaker automatic mode switching for the detection of atrial fibrillation and atrial tachycardia with Holter recording
    de Voogt, Willem G.
    van Hemel, Norbert M.
    van de Bos, Arjan A.
    Koistinen, Juhani
    Fast, Jules H.
    EUROPACE, 2006, 8 (11): : 950 - 961
  • [6] Programmed atrial sensitivity: A critical determinant in atrial fibrillation detection and optimal automatic mode switching
    Leung, SK
    Lau, CP
    Lam, CTF
    Tse, HF
    Tang, MO
    Chung, F
    Ayers, G
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11): : 2214 - 2219
  • [7] Automatic mode switching variants: Dual demand pacing, retriggerable atrial refractory periods, automatic mode adaptation, and pseudomode switching. Enlightenment or obfuscation?
    Barold, SS
    Byrd, CL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (07): : 1065 - 1067
  • [8] Efficacy of automatic mode switching in DDDR mode pacemakers: The most 2 study
    Santomauro, Maurizio
    Ottaviano, Luca
    Borrelli, Alessio
    Galasso, Gennaro
    Duilio, Carlo
    Monteforte, Nicola
    Padeletti, Luigi
    Montenero, Annibale Sandro
    Andrew, Peter
    Chiariello, Massimo
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2008, 21 (01) : 13 - 17
  • [9] Efficacy of automatic mode switching in DDDR mode pacemakers: The most 2 study
    Maurizio Santomauro
    Luca Ottaviano
    Alessio Borrelli
    Gennaro Galasso
    Carlo Duilio
    Nicola Monteforte
    Luigi Padeletti
    Annibale Sandro Montenero
    Peter Andrew
    Massimo Chiariello
    Journal of Interventional Cardiac Electrophysiology, 2008, 21 : 13 - 17
  • [10] Atrial Undersensing and Cycle Prolongation Related to Automatic Mode Switching: What is the Mechanism?
    Barold, S. S.
    Stroobandt, Roland X.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (12): : 1507 - 1508