Modelling the response function of energy dispersive X-ray spectrometers with silicon detectors

被引:107
|
作者
Scholze, F. [1 ]
Procop, M. [2 ]
机构
[1] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
[2] Bundesanstalt Mat Forsch & Prufung, D-12200 Berlin, Germany
关键词
ELECTRON-HOLE PAIR; SI(LI) DETECTOR; RANGE; CALIBRATION; SPECTRA; PHOTOIONIZATION; TRANSMISSION; PARAMETERS; SCATTERING; EFFICIENCY;
D O I
10.1002/xrs.1165
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
A new, analytical description of the physical processes determining the spectral response of an energy dispersive X-ray spectrometer with a silicon detector (Si(Li) or silicon drift detector (SDD)) is presented. The model considers the detector statistical noise, the electronic noise, the incomplete charge collection (ICC) that gives rise to the peak tailing, the escape effect, the fluorescence of the front contact or the dead layer and hot photoelectrons that cause the shelf. Only five free parameters are necessary to model the response function: the electronic noise, three parameters describing the shape of the charge collection efficiency beneath the front contact and the thickness of the detector front layer. Once the five parameters are adjusted to have agreement between a measured and a calculated response function, the response function can be calculated for any other photon energy in the range from 0.1 keV to 30 keV. The algorithm is implemented in IDL and MATLAB and is available also as MATLAB stand-alone program. It enables the determination of the optimum parameter set by fitting a calculated response function to a measured one for monochromatic radiation. A (m,n)-type matrix can be calculated whereby m represents the number of channels for the response function and n the number of photon energies in the selected range. The matrix can be used to convolute a calculated spectrum for comparison with a measured one. The calculated response functions are in agreement with the pulse height distributions measured with monochromatic synchrotron radiation in the energy range from 0.1 keV to 10 keV for three spectrometers with detector crystals different in construction. It is shown that the improved description of the detector response enables the detection of minor components of characteristic lines in fluorescence spectra, which have been attributed earlier to the detector. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:312 / 321
页数:10
相关论文
共 50 条
  • [1] THERMAL DETECTORS AS X-RAY SPECTROMETERS
    MOSELEY, SH
    MATHER, JC
    MCCAMMON, D
    [J]. JOURNAL OF APPLIED PHYSICS, 1984, 56 (05) : 1257 - 1262
  • [2] The spectral response of silicon X-ray detectors
    Eggert, T.
    Boslau, O.
    Kemmer, J.
    Pahke, A.
    Wiest, F.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 568 (01): : 1 - 11
  • [3] Gaseous detectors for energy dispersive X-ray fluorescence analysis
    Veloso, J. F. C. A.
    Silva, A. L. M.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 878 : 24 - 39
  • [4] THERMAL DETECTORS AS X-RAY SPECTROMETERS.
    Moseley, S.H.
    Mather, J.C.
    McCammon, D.
    [J]. Journal of Applied Physics, 1984, 56 (05): : 1257 - 1262
  • [5] The use of 'Joule-Thomson' type cooled detectors for field transportable Energy Dispersive X-ray Fluorescence spectrometers
    Russell, A
    Feldmann, T
    [J]. FIELD SCREENING EUROPE, 1997, : 339 - 342
  • [6] The determination of the efficiency of energy dispersive X-ray spectrometers by a new reference material
    Alvisi, Marco
    Blome, Markus
    Griepentrog, Michael
    Hodoroaba, Vasile-Dan
    Karduck, Peter
    Mostert, Marco
    Nacucchi, Michele
    Procop, Mathias
    Rohde, Martin
    Scholze, Frank
    Statham, Peter
    Terborg, Ralf
    Thiot, Jean-Francois
    [J]. MICROSCOPY AND MICROANALYSIS, 2006, 12 (05) : 406 - 415
  • [7] Application of energy dispersive X-ray fluorescence spectrometers to iron and steel industry
    Ji, Ang
    Zhuo, Shangjun
    Tao, Guangyi
    [J]. Kang T'ieh/Iron and Steel (Peking), 2001, 36 (10): : 64 - 68
  • [8] Monte Carlo methods for X-ray dispersive spectrometers
    Peterson, JR
    Jernigan, JG
    Kahn, SM
    [J]. ASTRONOMICAL DATA ANALYSIS II, 2002, 4847 : 195 - 206
  • [9] THERMAL DETECTORS AS SINGLE PHOTON X-RAY SPECTROMETERS
    MOSELEY, SH
    KELLEY, RL
    MATHER, JC
    MUSHOTZKY, RF
    SZYMKOWIAK, AE
    MCCAMMON, D
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) : 134 - 138
  • [10] Superconductive tunnel junction detectors as X-ray spectrometers
    deKorte, PAJ
    vandenBerg, ML
    Bruijn, MP
    Gomez, J
    Kiewiet, F
    vanLieshout, HL
    Luiten, OJ
    Brons, CGS
    Flokstra, J
    Hamster, AW
    [J]. EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY VII, 1996, 2808 : 506 - 515