Large volume sample stacking for rapid and sensitive determination of antidiabetic drug metformin in human urine and serum by capillary electrophoresis with contactless conductivity detection

被引:36
|
作者
Tuma, Petr [1 ]
机构
[1] Charles Univ Prague, Fac Med 3, Inst Biochem Cell & Mol Biol, Ruska 87, Prague 10000 10, Czech Republic
关键词
Antidiabetic drug; Capillary electrophoresis; Contact less conductivity detection; Large volume sample stacking; Rapid separation; PERIAQUEDUCTAL GRAY-MATTER; TYPE-2; DIABETES-MELLITUS; TANDEM MASS-SPECTROMETRY; ZONE-ELECTROPHORESIS; LIQUID-CHROMATOGRAPHY; PHARMACEUTICAL-PREPARATIONS; DOSAGE FORMULATIONS; HUMAN PLASMA; AMINO-ACIDS; VALIDATION;
D O I
10.1016/j.chroma.2014.04.016
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Two CE methods with contactless conductivity detection have been developed for determining the oral antidiabetic drug metformin in human urine and blood. The determination of metformin is performed on a separation capillary with an effective length of 14 cm, using a maximum voltage of 30 kV and with a small injection of 50-fold diluted urine into the capillary. Under these conditions, the migration time of metformin is 35 s and the LOD is 0.3 mu M Large-volume sample stacking was used to determine low metformin levels in serum. The injection of a sample of serum deproteinized with acetonitrile was 10 times greater compared to the injected amount of urine. This enabled reduction of the LOD to 0.03 mu M and the metformin migration time equalled 86 s. The undesirable solvent from sample zone was forced out of the capillary to ensure rapidity and good repeatability of the determination. The RSD values for the migration time are 0.1% for urine and 0.7% for serum; RSD for the peak areas equalled 1.4% for urine and 2.6% for serum. The developed CE technique was tested on performance of routine analyses of metformin in the urine and serum of patients suffering from type II diabetes mellitus. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:207 / 211
页数:5
相关论文
共 50 条
  • [1] Rapid and Sensitive Determination of Metformin in Human Urine and Serum by Capillary Electrophoresis with Contactless Conductivity Detection
    Tuma, Petr
    Matejckova, Jana Fauknerova
    Jurka, Vlastimil
    Samcova, Eva
    XXXIV. MODERNI ELEKTROCHEMICKE METODY, 2014, : 212 - 215
  • [2] The Use of Large Volume Sample Stacking with Acetonitrile for Determination of Neurotransmitters by Capillary Electrophoresis with Contactless Conductivity Detection
    Tuma, Petr
    Pavlicek, Vaclav
    Malkova, Klara
    Samcova, Eva
    XXXIII MODERNI ELEKTROCHEMICKE METODY, 2013, : 227 - 230
  • [3] Determination of tobramycin in human serum by capillary electrophoresis with contactless conductivity detection
    Law, Wai Siang
    Kuban, Pavel
    Yuan, Ling Ling
    Zhao, Jian Hong
    Li, Sam Fong Yau
    Hauser, Peter C.
    ELECTROPHORESIS, 2006, 27 (10) : 1932 - 1938
  • [4] Highly sensitive rapid determination of orotic acid in urine samples using a field-amplified sample stacking approach in capillary electrophoresis coupled with contactless conductivity detection
    Adimcilar, Veselina
    Saygili, Merve Taser
    Cansever, Mehmet Serif
    Oztekin, Nevin
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2024, 238
  • [5] Rapid and sensitive determination of strychnine and brucine in human urine by capillary electrophoresis with field-amplified sample stacking
    Li, Junmei
    Jiang, Ye
    BIOMEDICAL CHROMATOGRAPHY, 2010, 24 (02) : 186 - 194
  • [6] Rapid determination of meldonium in urine samples by capillary electrophoresis with capacitively coupled contactless conductivity detection
    Sampova, Andrea
    Kuban, Pavel
    JOURNAL OF CHROMATOGRAPHY A, 2016, 1468 : 236 - 240
  • [7] Determination of major inorganic ions in blood serum and urine by capillary electrophoresis with contactless conductivity detection
    Wan, QJ
    Kubán, P
    Tanyanyiwa, J
    Rainelli, A
    Hauser, PC
    ANALYTICA CHIMICA ACTA, 2004, 525 (01) : 11 - 16
  • [8] Sensitive Determination of Barbiturates in Biological Matrix by Capillary Electrophoresis Using Online Large-Volume Sample Stacking
    Fan, Liu-Yin
    He, Ting
    Tang, Yun-Yun
    Zhang, Wei
    Song, Chao-Jin
    Zhao, Xia
    Zhao, Xiao-Yu
    Cao, Cheng-Xi
    JOURNAL OF FORENSIC SCIENCES, 2012, 57 (03) : 813 - 819
  • [9] CAPILLARY ELECTROPHORESIS WITH CONTACTLESS CONDUCTIVITY DETECTION FOR AMINO ACID PROFILING OF HUMAN AND URINE SAMPLES
    Madr, Ales
    Cela, Andrea
    Musilova, Jindra
    Zeisbergerova, Marta
    Glatz, Zdenek
    CHEMICKE LISTY, 2013, 107 : S318 - S319
  • [10] Large-volume sample stacking-capillary electrophoresis used for the determination of 3-nitrotyrosine in rat urine
    Maeso, N
    Cifuentes, A
    Barbas, C
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 809 (01): : 147 - 152