Bending stiffness and pull-out strength of tapered, variable pitch screws, and 6.5-mm cancellous screws in acute Jones fractures

被引:32
|
作者
Sides, Steven D.
Fetter, Nicole L.
Glisson, Richard
Nunley, James A.
机构
[1] Duke Univ, Med Ctr, Durham, NC 27710 USA
[2] Madigan Army Med Ctr, Ft Lewis, WA USA
关键词
fracture; Jones; intramedullary; metatarsal; screw;
D O I
10.1177/107110070602701012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Stabilization of fifth metatarsal Jones fractures with intramedullary screws is popular, particularly in athletes, because nonoperative treatment involves prolonged casting and a distinct risk of nonunion or delayed union. Conventional lag screws of various diameters are routinely used for Jones fracture fixation. More recently, tapered, headless, variable pitch screws have become available as an option. These screws have the advantage of not having a protruding screw head, but information regarding their performance in Jones fracture fixation is limited. To determine whether differences exist in the mechanical integrity of fifth metatarsals fixed with each type of screw, this study was designed to compare Jones fracture fixation with 6.5-mm partially-threaded lag screws and headless, tapered, variable pitch compression screws with a 4-mm leading-thread diameter and 5-mm trailing-thread diameter. Methods: Simulated Jones fractures were created in 20 matched pairs of fresh-frozen fifth metatarsals. One bone from each pair was stabilized with a tapered, variable pitch screw, and the contralateral with a 6.5-mm partially-threaded cancellous lag screw. The stiffness in lateral-to-medial bending of the resulting constructs and the resistance of the screws to pulling out of the distal fragment were quantified. Results: There was no demonstrable difference in bending stiffness between metatarsals fixed with the two types of screws (p = 0.688). The 6.5-mm screw provided significantly higher resistance to pull-out (p = 0.001). Conclusions: Headless, tapered, variable pitch compression screws of the size tested are not entirely comparable to 6.5-mm lag screws in this application. They are effective in resisting bending but do not offer equivalent resistance to thread pull-out.
引用
收藏
页码:821 / 825
页数:5
相关论文
共 7 条
  • [1] Interfragmentary compression and pull-out properties of 6.5-mm AO cancellous lag screws in a uniform synthetic material during tightening procedures
    Sun, Peidong
    Xu, Daqiang
    Zhao, Weidong
    Jiao, Peifeng
    Li, Zeyu
    Liu, Chang
    Ouyang, Jun
    [J]. MEDICAL ENGINEERING & PHYSICS, 2017, 42 : 116 - 120
  • [2] Effect of thread pitch on pull-out strength of laparoscopic myoma screws
    Tintara, Hatern
    Aiyarak, Pattara
    Mitarnun, Winyou
    Geater, Alan
    [J]. JOURNAL OF OBSTETRICS AND GYNAECOLOGY RESEARCH, 2006, 32 (04) : 428 - 433
  • [3] Proximal half angle of the screw thread is a critical design variable affecting the pull-out strength of cancellous bone screws
    Wang, Yingxu
    Mori, Ryuji
    Ozoe, Nobuaki
    Nakai, Takahisa
    Uchio, Yuji
    [J]. CLINICAL BIOMECHANICS, 2009, 24 (09) : 781 - 785
  • [4] Evaluation of uniaxial pull-out strength of 1.5mm monocortical screws.
    Rabelo, LRS
    Passeri, LA
    [J]. JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1139 - 1139
  • [5] INTERNAL-FIXATION OF FEMORAL-NECK FRACTURES - A COMPARATIVE BIOMECHANICAL STUDY OF KNOWLES PINS AND 6.5-MM CANCELLOUS SCREWS
    SPRINGER, ER
    LACHIEWICZ, PF
    GILBERT, JA
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1991, (267) : 85 - 92
  • [6] Axial pull-out strength of 3.5 cortical and 4.0 cancellous bone screws placed in canine proximal tibias using manual and power tapping
    Demko, J. L.
    Soniat, M. E.
    Elder, S.
    McLaughlin, R.
    [J]. VETERINARY AND COMPARATIVE ORTHOPAEDICS AND TRAUMATOLOGY, 2008, 21 (04) : 323 - 328
  • [7] Estimation of Shear Strength and Pull-out Strength of Timber Joints Using Deformed Steel Bars for Civil Engineering Construction with Large Cross Sections. Performance comparison of specimens utilizing lag screws and deformed steel bars of φ 16 mm and long length
    Noda, Ryu
    Sasaki, Takanobu
    Chida, Tomohiro
    [J]. MOKUZAI GAKKAISHI, 2014, 60 (05): : 249 - 260