Combined effect of the body mass index and implant size on the wear of retrieved total knee prostheses

被引:7
|
作者
Battaglia, Santina [1 ]
Taddei, Paola [2 ]
Castiello, Emanuela [1 ,3 ]
Tozzi, Silvia [2 ]
Sudanese, Alessandra [1 ,3 ]
Affatato, Saverio [1 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, Bologna, Italy
[2] Univ Bologna, Dipartimento Sci Biomed & Neuromotorie, I-40126 Bologna, Italy
[3] Ist Ortoped Rizzoli, Bologna, Italy
关键词
TOTAL HIP; IN-VITRO; OBESITY; ARTHROPLASTY; POLYETHYLENE; REPLACEMENT; OSTEOARTHRITIS; STRESS; WEIGHT; RISK;
D O I
10.1016/j.jmbbm.2014.06.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Total Knee Arthroplasty is a well-established surgical procedure performed to relieve pain and to restore function in knee osteoarthritis. A proper choice of the implant size is mandatory in order to guarantee the success of the implant and to respect the bone stock of tibial plateau and femoral condyles. The aim of this study was to investigate the combined effect of Body Mass Index and implant size on the wear damage of retrieved knee implants, a still debated subject. To this purpose, twelve total knee prostheses of the same design (NexGen, Zimmer, Warsaw, Indiana) but with different sizes were investigated. These prostheses were all cemented fixed bearing inserts and posterior stabilised. Roughness analyses were performed on femoral components to assess surface modifications. Micro-Raman spectroscopic characterization was performed on polyethylene inserts to evaluate crystallinity changes. A worsening of the surface and a nonhomogeneous aspect was observed in the prostheses characterized by a high ratio between preoperative Body Mass Index and implant size, which we called Phi factor. Delamination, a good indicator for polyethylene osteolysis, was more pronounced in the inserts having a high Phi than in those with a low Phi. Micro-Raman analyses showed more significant crystallinity changes in the prostheses of lower size, in agreement with the results previously obtained in vitro. The findings of the present study suggested that the Phi factor can be considered a predictor factor of the Total Knee Arthroplasty overall performance. Potentially, it could be used as a discriminating factor to redact Total Knee Arthroplasty outcome and prevent complications. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:69 / 77
页数:9
相关论文
共 50 条
  • [1] Toward the interpretation of the combined effect of size and body weight on the tribological performance of total knee prostheses
    Battaglia, Santina
    Taddei, Paola
    Tozzi, Silvia
    Sudanese, Alessandra
    Affatato, Saverio
    INTERNATIONAL ORTHOPAEDICS, 2014, 38 (06) : 1183 - 1190
  • [2] Toward the interpretation of the combined effect of size and body weight on the tribological performance of total knee prostheses
    Santina Battaglia
    Paola Taddei
    Silvia Tozzi
    Alessandra Sudanese
    Saverio Affatato
    International Orthopaedics, 2014, 38 : 1183 - 1190
  • [3] Implant migration and failure in total knee arthroplasty is related to body mass index and tibial component size
    Berend, Michael E.
    Ritter, Merrill A.
    Hyldahl, Hans Christian
    Meding, John B.
    Redelman, Ryan
    JOURNAL OF ARTHROPLASTY, 2008, 23 (06): : 104 - 109
  • [4] Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses
    Blunn, GW
    Joshi, AB
    Minns, RJ
    Lidgren, L
    Lilley, P
    Ryd, L
    Engelbrecht, E
    Walker, PS
    JOURNAL OF ARTHROPLASTY, 1997, 12 (03): : 281 - 290
  • [5] FAILURE ANALYSIS OF RETRIEVED CEMENTED TOTAL KNEE PROSTHESES
    HALVORSON, TL
    COOK, SD
    SANDBORN, PM
    BRUNET, ME
    SOUTHERN MEDICAL JOURNAL, 1988, 81 (09) : S54 - S54
  • [6] Polyethylene wear characteristics in retrieved total knee replacements
    Hall, Emily A.
    Bland, Tracie M.
    Lemons, Jack E.
    Cuckler, John M.
    Eberhardt, Alan W.
    American Society of Mechanical Engineers, Bioengineering Division (Publication) BED, 1999, 43 : 7 - 8
  • [7] Effect of Body Mass Index on Blood Transfusion in Total Hip and Knee Arthroplasty
    Frisch, Nicholas
    Wessell, Nolan M.
    Charters, Michael
    Peterson, Ed
    Cann, Brett
    Greenstein, Alex
    Silverton, Craig D.
    ORTHOPEDICS, 2016, 39 (05) : E844 - E849
  • [8] Effect of Body Mass Index on Limb Alignment After Total Knee Arthroplasty
    Estes, Chris S.
    Schmidt, Kenneth J.
    McLemore, Ryan
    Spangehl, Mark J.
    Clarke, Henry D.
    JOURNAL OF ARTHROPLASTY, 2013, 28 (08): : 101 - 105
  • [9] Effect of Body Mass Index on Reoperation and Complications After Total Knee Arthroplasty
    Wagner, Eric R.
    Kamath, Atul F.
    Fruth, Kristin
    Harmsen, William S.
    Berry, Daniel J.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2016, 98 (24): : 2052 - 2060
  • [10] Effect of Body Mass Index on Range of Motion and Manipulation After Total Knee Arthroplasty
    Gadinsky, Naomi E.
    Ehrhardt, Jessica K.
    Urband, Christopher
    Westrich, Geoffrey H.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (08): : 1194 - 1197