Measurement of radon concentration in the air by PicoRad detectors

被引:0
|
作者
Chau, Nguyen Dinh [1 ]
机构
[1] AGH Univ Sci & Technol, Fac Phys & Appl Comp Sci, PL-30059 Krakow, Poland
关键词
radon; PicoRad detectors; charcoal; adsorption; desorption; liquid scintillation;
D O I
暂无
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The paper presents the theory and methodology of radon concentration measurement in the air by PicoRad detector, which consists of a porous canister held securely near the top of a plastic vial. The porous canister contains a bed of a controlled mass of charcoal (1.3 g) and silica desiccant (0.9 g). To measure the radon concentration, the vial detector is exposed in the radon laden air for a certain interval of time (24, 48 or 72 h or longer), then the liquid scintillation cocktail is added into the vial and measured by the help of a liquid scintillation counter. In this paper both radon adsorption by the charcoal detector, while it is being exposed in the radon laden air, and desorption of radon from the charcoal into the liquid scintillation cocktail were investigated. As a result, some conclusions concerning the methodology of radon measurement by PicoRad detectors have been done. The desorption factor of radon from the charcoal into the liquid scintillation cocktail was determined. The radon concentration values in the air measured by PicoRad detectors were in good agreement with those obtained by an AlphaGuard radon monitor.
引用
收藏
页码:S21 / S24
页数:4
相关论文
共 50 条
  • [1] RADON CONCENTRATION IN THE SOIL AIR MEASURED BY TRACK DETECTORS
    KVASNICKA, J
    NUCLEAR INSTRUMENTS & METHODS, 1980, 174 (03): : 599 - 604
  • [2] A SCINTILLATION COUNTER FOR MEASUREMENT OF RADON CONCENTRATION IN AIR
    COLLINSON, AJL
    HAQUE, AKM
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1963, 40 (11): : 521 - &
  • [3] Investigation of the influence of high humidity and exposure duration on the measurement results of radon concentration by means of PicoRad system in the CLOR calibration chamber
    Stawarz, Olga
    Karpinska, Maria
    Mamont-Ciesla, Kalina
    NUKLEONIKA, 2006, 51 : S53 - S57
  • [4] About ageing and fading of Cr-39 PADC track detectors used as air radon concentration measurement devices
    Caresana, M.
    Ferrarini, M.
    Garlati, L.
    Parravicini, A.
    RADIATION MEASUREMENTS, 2010, 45 (02) : 183 - 189
  • [5] MEASUREMENT OF RADON CONCENTRATION USING BARE CR-39 TRACK DETECTORS
    YONEHARA, H
    KIMURA, H
    AOYAMA, T
    MIFUNE, M
    SAKANOUE, M
    SUGAHARA, T
    JOURNAL OF RADIATION RESEARCH, 1987, 28 (01) : 73 - 73
  • [6] Further studies on ageing and fading of CR39 PADC track detectors used as air radon concentration measurement devices
    Caresana, M.
    Ferrarini, M.
    Garlati, L.
    Parravicini, A.
    RADIATION MEASUREMENTS, 2011, 46 (10) : 1160 - 1167
  • [7] EVALUATION OF A RADON AIR MONITOR IN THE MEASUREMENT OF RADON CONCENTRATION IN WATER IN COMPARISON WITH A LIQUID SCINTILLATION COUNTER
    Higuchi, Shoko
    Kamishiro, Yuuki
    Ishihara, Maki
    Yasuoka, Yumi
    Mori, Yasunori
    Hosoda, Masahiro
    Iwaoka, Kazuki
    Tokonami, Shinji
    Takahashi, Rikiya
    Janik, Miroslaw
    Muto, Jun
    Nagahama, Hiroyuki
    Mukai, Takahiro
    RADIATION PROTECTION DOSIMETRY, 2019, 184 (3-4) : 426 - 429
  • [8] Continuos measurement of radon concentration in the air with the Lucas cell by periodic sampling
    Machaj, B
    Urbanski, P
    NUKLEONIKA, 1999, 44 (04) : 579 - 594
  • [9] Radon concentration dependence of α radioactivity measurement using ionized air transportation
    Naito, Susumu
    Sano, Akira
    Izumi, Mikio
    Noda, Etsuo
    Hayashi, Kazuo
    Sato, Mitsuyoshi
    Suto, Osamu
    Miyamoto, Yasuaki
    Kondo, Shinichi
    Iinuma, Koichi
    Nanbu, Kenichi
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 495 - 499
  • [10] RADON CONCENTRATION IN WATER MEASURED BY TRACK DETECTORS
    KVASNICKA, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 206 (03): : 563 - 567