Results from a European clinical investigation of the Nucleus® multichannel auditory brainstem implant

被引:96
|
作者
Nevison, B
Laszig, R
Sollmann, WP
Lenarz, T
Sterkers, O
Ramsden, R
Fraysse, B
Manrique, M
Rask-Andersen, H
Garcia-Ibanez, E
Colletti, V
von Wallenberg, E
机构
[1] Cochlear Europe Ltd, London SW19 4DF, England
[2] Univ Freiburg, Freiburg, Germany
[3] Stadt Klinikum, Braunschweig, Germany
[4] Hannover Med Sch, Hannover, Germany
[5] Hop Beaujon, Paris, France
[6] Manchester Royal Infirm, Manchester M13 9WL, Lancs, England
[7] CHU Purpan, Toulouse, France
[8] Univ Navarra Clin, Pamplona, Spain
[9] Uppsala Univ, Uppsala, Sweden
[10] Inst Otol, Barcelona, Spain
[11] Univ Verona, I-37100 Verona, Italy
[12] Cochlear AG, Basel, Switzerland
来源
EAR AND HEARING | 2002年 / 23卷 / 03期
关键词
D O I
10.1097/00003446-200206000-00002
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: This study was designed to investigate the perceptual benefits and potential risks of implanting the Nucleus(R) multichannel auditory brainstem implant. Design: Between September 1992 and October 1997 a total of 27 subjects received a Nucleus 20- or 21-channel Auditory Brainstem Implant (ABI). All subjects involved in the trial had bilateral acoustic tumour as a result of neurofibromatosis type 2 (NF2) resulting in complete dysfunction of the VIIIth nerve. The study used each subject as their own control without a preoperative baseline because residual hearing, if existing, was destroyed at surgery by tumour removal. A battery of speech tests was conducted to evaluate each patient's performance and communication abilities. Tests were conducted, where possible, in the auditory-only, visual-only, and auditory-visual conditions at 3 days postoperatively (baseline), at 3-mo intervals for the first year and every 12 mo thereafter. A subjective performance questionnaire was administered together with an extensive neurological examination at each test interval. Results: 27 subjects involved in this trial were successfully implanted with a Nucleus ABI. One subject died 2 days postoperatively due to a lung embolism unrelated to the device. Twenty-six subjects underwent device activation and all but one patient received auditory sensation at initial stimulation (96.2%). On average 8.6 (+/-4.2) of the available 21 electrodes were used in the patients' MAPs. Performance evaluation measures showed that the majority of users had access to auditory information such as environmental sound awareness together with stress and rhythm cues in speech that assist with lipreading. Although most subjects did not achieve any functional auditory-alone, open-set speech understanding, two subjects from this series (7.4%) did receive sufficient benefit to be able to use the ABI in conversation without lipreading. Conclusions: Although the medical risks and surgical complexity associated with ABI device implantation are far greater than those for a cochlear implant, the clinical results from this trial show that the Nucleus multichannel ABI is capable of providing a significant patient benefit over risk ratio for subjects suffering loss of hearing due to bilateral retrocochlear lesions.
引用
下载
收藏
页码:170 / 183
页数:14
相关论文
共 50 条
  • [1] Multichannel auditory brainstem implant: US clinical trial results
    Ebinger, K
    Otto, S
    Arcaroli, J
    Staller, S
    Arndt, P
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2000, 114 : 50 - 53
  • [2] Electrical stimulation of the cochlear nucleus. First results of the European auditory brainstem implant clinical trial
    Sollman, WP
    Laszig, R
    Marangos, N
    Charachon, R
    Fraysse, B
    Sterkers, O
    Rey, A
    Ramsden, R
    Samii, M
    Matthies, C
    Lenarz, T
    RaskAndersen, H
    Nyberg, G
    Manrique, M
    11TH INTERNATIONAL CONGRESS OF NEUROLOGICAL SURGERY, VOLS 1 AND 2, 1997, : 1569 - 1573
  • [3] The multichannel auditory brainstem implant: 6-month coinvestigator results
    Otto, SR
    Brackmann, DE
    Staller, S
    Menapace, CM
    COCHLEAR IMPLANT AND RELATED SCIENCES UPDATE, 1997, 52 : 1 - 7
  • [4] 1 year results from the clinical trial of the Nucleus(R) 21 channel auditory brainstem implant (ABI)
    Laszig, R
    Marangos, N
    Sollmann, WP
    Nevison, B
    Ramsden, R
    Lenarz, T
    Fraysse, B
    Sterkers, O
    COCHLEAR IMPLANTS, 1996, : 145 - 150
  • [5] Surgical experiences in 58 cases using the Nucleus 22 multichannel auditory brainstem implant
    Sollmann, WP
    Laszig, R
    Marangos, N
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2000, 114 : 23 - 26
  • [6] Multichannel auditory brainstem implant: update on performance in 61 patients
    Otto, SR
    Brackmann, DE
    Hitselberger, WE
    Shannon, RV
    Kuchta, J
    JOURNAL OF NEUROSURGERY, 2002, 96 (06) : 1063 - 1071
  • [7] Use of a multichannel auditory brainstem implant for neurofibromatosis type 2
    Schwartz, MS
    Otto, SR
    Brackmann, DE
    Hitselberger, WE
    Shannon, RV
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2003, 81 (1-4) : 110 - 114
  • [8] Results of the MXM digisonic auditory brainstem implant clinical trials in Europe
    Vincent, C
    Zini, C
    Gandolfi, A
    Triglia, JM
    Pellet, W
    Truy, E
    Fischer, G
    Maurizi, M
    Meglio, M
    Lejeune, JP
    Vaneecloo, FM
    OTOLOGY & NEUROTOLOGY, 2002, 23 (01) : 56 - 60
  • [9] The multichannel auditory brainstem implant: how many electrodes make sense?
    Kuchta, J
    Otto, SR
    Shannon, RV
    Hitselberger, WE
    Brackmann, DE
    JOURNAL OF NEUROSURGERY, 2004, 100 (01) : 16 - 23
  • [10] Stimulation of the cochlear nucleus with multichannel auditory brainstem implants and long-term results: Freiburg patients
    Marangos, N
    Stecker, M
    Sollmann, WP
    Laszig, R
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2000, 114 : 27 - 31