Power MOSFET Linearizer of a High-Voltage Power Amplifier for High-Frequency Pulse-Echo Instrumentation

被引:29
|
作者
Choi, Hojong [1 ]
Woo, Park Chul [1 ]
Yeom, Jung-Yeol [2 ]
Yoon, Changhan [3 ]
机构
[1] Kumoh Natl Inst Technol, Dept Med IT Convergence Engn, Gumi 39253, South Korea
[2] Korea Univ, Sch Biomed Engn, Seoul 02841, South Korea
[3] Inje Univ, Dept Biomed Engn, Gimhae 50834, South Korea
基金
新加坡国家研究基金会;
关键词
high-voltage power amplifier; ultrasonic transducer; power MOSFET linearizer; ULTRASOUND; DISTORTION;
D O I
10.3390/s17040764
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A power MOSFET linearizer is proposed for a high-voltage power amplifier (HVPA) used in high-frequency pulse-echo instrumentation. The power MOSFET linearizer is composed of a DC bias-controlled series power MOSFET shunt with parallel inductors and capacitors. The proposed scheme is designed to improve the gain deviation characteristics of the HVPA at higher input powers. By controlling the MOSFET bias voltage in the linearizer, the gain reduction into the HVPA was compensated, thereby reducing the echo harmonic distortion components generated by the ultrasonic transducers. In order to verify the performance improvement of the HVPA implementing the power MOSFET linearizer, we measured and found that the gain deviation of the power MOSFET linearizer integrated with HVPA under 10 V DC bias voltage was reduced (-1.8 and -0.96 dB, respectively) compared to that of the HVPA without the power MOSFET linearizer (-2.95 and -3.0 dB, respectively) when 70 and 80 MHz, three-cycle, and 26 dB(m) input pulse waveforms are applied, respectively. The input 1-dB compression point (an index of linearity) of the HVPA with power MOSFET linearizer (24.17 and 26.19 dB(m) at 70 and 80 MHz, respectively) at 10 V DC bias voltage was increased compared to that of HVPA without the power MOSFET linearizer (22.03 and 22.13 dB(m) at 70 and 80 MHz, respectively). To further verify the reduction of the echo harmonic distortion components generated by the ultrasonic transducers, the pulse-echo responses in the pulse-echo instrumentation were compared when using HVPA with and without the power MOSFET linearizer. When three-cycle 26 dB(m) input power was applied, the second, third, fourth, and fifth harmonic distortion components of a 75 MHz transducer driven by the HVPA with power MOSFET linearizer (-48.34, -44.21, -48.34, and -46.56 dB, respectively) were lower than that of the HVPA without the power MOSFET linearizer (-45.61, -41.57, -45.01, and -45.51 dB, respectively). When five-cycle 20 dB(m) input power was applied, the second, third, fourth, and fifth harmonic distortions of the HVPA with the power MOSFET linearizer (-41.54, -41.80, -48.86, and -46.27 dB, respectively) were also lower than that of the HVPA without the power MOSFET linearizer (-25.85, -43.56, -49.04, and -49.24 dB, respectively). Therefore, we conclude that the power MOSFET linearizer could reduce gain deviation of the HVPA, thus reducing the echo signal harmonic distortions generated by the high-frequency ultrasonic transducers in pulse-echo instrumentation.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Digital high-frequency high-voltage pulse power supply for electrostatic precipitators
    Geng, Tao
    Li, Ge
    Cao, Liang
    Zhou, Yin-Gui
    Wang, Hai-Tian
    Li, Zhe
    [J]. Gaodianya Jishu/High Voltage Engineering, 2009, 35 (07): : 1695 - 1699
  • [2] A cascaded high-voltage high-frequency linear amplifier for power transmission line measurement
    Wang, Shouming
    Liu, Ji
    Huang, Jun
    Zhang, Haiyue
    Li, Zhen
    Cao, Jinming
    [J]. MEASUREMENT, 2024, 227
  • [3] High-frequency characteristics of high-voltage power line
    Suljanovic, N
    Mujcic, A
    Zajc, M
    Tasic, JF
    [J]. IEEE REGION 8 EUROCON 2003, VOL B, PROCEEDINGS: COMPUTER AS A TOOL, 2003, : 310 - 314
  • [4] Design of a high-frequency high-voltage pulse power supply for cable fault detection
    [J]. Liu, Z. (vystdy@163.com), 2013, Institution of Engineers (Australia) (10):
  • [5] A HIGH-VOLTAGE ELECTROOPTIC DRIVE AMPLIFIER USING A POWER MOSFET PAIR
    THEOPHANOUS, N
    TSITOMENEAS, S
    ALEXAKIS, G
    ARAPOYIANNI, A
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1988, 37 (01) : 49 - 52
  • [6] A Wideband High-Voltage Power Amplifier Post-Linearizer for Medical Ultrasound Transducers
    Choi, Hojong
    Yoon, Changhan
    Yeom, Jung-Yeol
    [J]. APPLIED SCIENCES-BASEL, 2017, 7 (04):
  • [7] Design and implementation of a high-voltage high-frequency pulse power generation system for plasma applications
    Tsai, M. T.
    Ke, C. W.
    [J]. 2009 INTERNATIONAL CONFERENCE ON POWER ELECTRONICS AND DRIVE SYSTEMS, VOLS 1 AND 2, 2009, : 35 - 39
  • [8] Low-power high-voltage high-frequency power supply for ozone generation
    Alonso, JM
    Cardesín, J
    Corominas, EL
    Rico-Secades, M
    García, J
    [J]. IEEE TRANSACTIONS ON INDUSTRY APPLICATIONS, 2004, 40 (02) : 414 - 421
  • [9] Low-power high-voltage high-frequency power supply for ozone generation
    Alonso, JM
    Rico-Secades, M
    Corominas, E
    Cardesín, J
    García, J
    [J]. CONFERENCE RECORD OF THE 2002 IEEE INDUSTRY APPLICATIONS CONFERENCE, VOLS 1-4, 2002, : 257 - 264
  • [10] High-voltage, high-frequency SiC power MOSFETs model validation
    Ortiz-Rodriguez, J. M.
    Duong, T.
    Rivera-Lopez, A.
    Hefner, A. R., Jr.
    [J]. 2007 IEEE POWER ELECTRONICS SPECIALISTS CONFERENCE, VOLS 1-6, 2007, : 1018 - 1022