Evaluation of intraocular pressure change after laser in situ keratomileusis using the pressure phosphene tonometer

被引:15
|
作者
Naruse, S
Mori, K
Kojo, M
Hieda, O
Kinoshita, S
机构
[1] Kyoto Prefectural Univ Med, Dept Ophthalmol, Kamigyo Ku, Kyoto 6020841, Japan
[2] Baptist Eye Clin, Kyoto, Japan
来源
JOURNAL OF CATARACT AND REFRACTIVE SURGERY | 2004年 / 30卷 / 02期
关键词
D O I
10.1016/S0886-3350(03)00614-X
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the applicability of pressure phosphene tonometry in measuring intraocular pressure (IOP) after laser in situ keratomileusis (LASIK). Setting: Baptist Eye Clinic, Kyoto, Japan. Methods: Thirty-six eyes of 22 consecutive patients who had had LASIK for myopia were enrolled in this prospective comparative study. The mean age of the patients was 32.6 years +/-9.0 (SD). The IOP was measured using the pressure phosphene tonometer (FPT), Goldmann applanation tonometry (GAT), and noncontact tonometry (NCT). The patients themselves took measurements with the FPT before and 1 week after surgery. One-way analysis of variance followed by a Bonferroni/Dunn multiple comparison post hoc test was administered to compare the preoperative and postoperative measurements with each tonometer. These data were represented graphically using a method described by Bland and Altman. The relationship between each tonometric change and the central corneal thickness (CCT) or keratometry (K) changes after LASIK were evaluated with by Pearson correlation coefficient. Results: Statistically significant differences were found between preoperative and postoperative measurements with GAT and NCT but not with FPT. Preoperative and postoperative FPT readings conformed to the British Standard for reproducibility of a standard test method for IOP estimation. No significant correlation between FPT changes and CCT or K changes was found. Conclusions: Pressure phosphene tonometry measures IOP through the upper eyelid, and changes in corneal shape do not appear to have an effect on the IOP. The IOP measurements with FPT were constant before and after LASIK.
引用
收藏
页码:390 / 397
页数:8
相关论文
共 50 条
  • [1] Evaluation of intraocular pressure change after laser in situ keratomileusis using pressure phosphene tonometer
    Naruse, S
    Mori, K
    Kojo, M
    Hieda, O
    Kinoshita, S
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2002, 43 : U484 - U484
  • [2] Changes in intraocular pressure after laser in situ keratomileusis
    Rashad, KM
    Bahnassy, AA
    JOURNAL OF REFRACTIVE SURGERY, 2001, 17 (04) : 420 - 427
  • [3] Reduction in intraocular pressure after laser in situ keratomileusis
    Park, HJ
    Uhm, KB
    Hong, C
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (02): : 303 - 309
  • [4] Intraocular pressure behavior after laser in situ keratomileusis
    Rojas-Diaz, J
    Baca, O
    Velasco, R
    Viggiano, D
    Betech-Hanono, M
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U12 - U12
  • [5] Intraocular pressure change measured by Goldmann tonometry after laser in situ keratomileusis
    Fournier, AV
    Podtetenev, M
    Lemire, J
    Thompson, P
    Duchesne, R
    Perreault, C
    Chehade, N
    Blondeau, P
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (07): : 905 - 910
  • [6] Intraocular pressure assessment after laser in situ keratomileusis: a review
    Tsai, Andrew S. H.
    Loon, Seng Chee
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (03): : 295 - 304
  • [7] Underestimation of intraocular pressure in eyes after laser in situ keratomileusis
    Arimoto, A
    Shimizu, K
    Shoji, N
    Enomoto, K
    Kohara, M
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2002, 46 (06) : 645 - 649
  • [8] The pressure phosphene tonometer - a clinical evaluation
    Chew, GSM
    Sanderson, GF
    Molteno, ACB
    EYE, 2005, 19 (06) : 683 - 685
  • [9] The pressure phosphene tonometer—a clinical evaluation
    G S M Chew
    G F Sanderson
    A C B Molteno
    Eye, 2005, 19 : 683 - 685
  • [10] Evaluation of the pressure phosphene tonometer as a self-tonometer
    Naruse, S
    Mori, K
    Kinoshita, S
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) : 421 - 428