Parallel-detection microwave spectroscopy system for breast imaging

被引:55
|
作者
Li, D [1 ]
Meaney, PM [1 ]
Raynolds, T [1 ]
Pendergrass, SA [1 ]
Fanning, MW [1 ]
Paulsen, KD [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2004年 / 75卷 / 07期
关键词
D O I
10.1063/1.1764609
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A liquid-coupled, noncontacting, broadband microwave imaging system has been designed and fabricated. Extension of the operating bandwidth allows us to exploit the potential of new clinical information for breast cancer diagnosis at frequencies higher than previously achieved. The new system design implements a parallel-detection scheme that allows signals to be simultaneously sampled at multiple receiving antenna sites (in 8 s for a single tomographic slice at a single frequency). It also has important features such as high cross-channel isolation (>120 dB), smooth broad bandwidth receiver response, and adjustable intermediate frequency signal amplification factors of 1 to 2000 to ensure successful realization of a large linear dynamic range which is especially important to counteract the increased signal loss at the higher operating frequencies. The new system is capable of recovering dielectric properties of breastlike phantoms with tumor inclusions over the frequency range from 0.5 to 2.1 GHz when embedded in an 87%/13% glycerin/Water background. Errors in the measurement data are less than 0.5% in signal amplitude and 1degrees in phase, on average. (C) 2004 American Institute of Physics.
引用
收藏
页码:2305 / 2313
页数:9
相关论文
共 50 条
  • [1] 3D parallel-detection microwave tomography for clinical breast imaging
    Epstein, N. R.
    Meaney, P. M.
    Paulsen, K. D.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (12):
  • [2] A parallel-detection frequency-domain near-infrared tomography system for hemoglobin imaging of the breast in vivo
    McBride, TO
    Pogue, BW
    Jiang, S
    Osterberg, UL
    Paulsen, KD
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (03): : 1817 - 1824
  • [3] A compact, parallel-detection diffuse optical mammography system
    Iftimia, N
    Gu, XJ
    Xu, Y
    Jiang, HB
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (05): : 2836 - 2842
  • [4] PROGRESS IN PARALLEL-DETECTION EELS
    KRIVANEK, OL
    KUNDMANN, MK
    [J]. INSTITUTE OF PHYSICS CONFERENCE SERIES, 1990, (98): : 33 - 40
  • [5] VIDICON-CAMERA PARALLEL-DETECTION SYSTEM FOR ANGLE-RESOLVED ELECTRON-SPECTROSCOPY
    WEEKS, SP
    ROWE, JE
    CHRISTMAN, SB
    CHABAN, EE
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1979, 50 (10): : 1249 - 1255
  • [6] DATA ACQUISITION AND PROCESSING FOR PARALLEL-DETECTION ELECTRON ENERGY-LOSS SPECTROSCOPY
    KUNDMANN, MK
    CHABERT, X
    MEYER, C
    KRIVANEK, OL
    [J]. JOURNAL OF ELECTRON MICROSCOPY, 1989, 38 (04): : 299 - 299
  • [7] AUTOMATED PROCESSING OF PARALLEL-DETECTION EELS DATA
    KUNDMANN, MK
    KRIVANEK, OL
    [J]. MICROSCOPY MICROANALYSIS MICROSTRUCTURES, 1991, 2 (2-3): : 245 - 256
  • [8] Microwave Imaging System for Early Detection of Breast Cancer
    Eltigani, Faihaa Mohammed
    Yahya, Maged Ali Ali
    Osman, Mazin Elfadil
    [J]. 2017 INTERNATIONAL CONFERENCE ON COMMUNICATION, CONTROL, COMPUTING AND ELECTRONICS ENGINEERING (ICCCCEE), 2017,
  • [9] An ultra wideband microwave imaging system for breast cancer detection
    Khort, Wee Chang
    Bialkowski, Marek E.
    Abbosh, Amin
    Seman, Norhudah
    Crozier, Stuart
    [J]. IEICE TRANSACTIONS ON COMMUNICATIONS, 2007, E90B (09) : 2376 - 2381
  • [10] Design and Modeling of a Microwave Imaging System for Breast Cancer Detection
    Tobon, Jorge A., V
    Attardo, Elia A.
    Dassano, Gianluca
    Vipiana, Francesca
    Casu, Mario R.
    Vacca, Marco
    Pulimeno, Azzurra
    Vecchi, Giuseppe
    [J]. 2015 9TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2015,