Reproducibility of disc and macula optical coherence tomography using the Canon OCT-HS100 as compared with the Zeiss Cirrus HD-OCT

被引:12
|
作者
Brautaset, Rune [1 ]
Birkeldh, Ulrika [1 ]
Rosen, Rebecka [1 ]
Ramsay, Marika Wahlberg [1 ]
Nilsson, Maria [1 ]
机构
[1] Karolinska Inst, St Eriks Eye Hosp, Dept Clin Neurosci, Unit Optometry, S-10420 Stockholm, Sweden
关键词
Canon OCT-HS100; Optical coherence tomography; Reproducibility; Spectral domain; NERVE-FIBER LAYER; THICKNESS MEASUREMENTS; ULTRAHIGH-RESOLUTION; HIGH-SPEED; SD-OCT; DEGENERATION; INSTRUMENTS; AGREEMENT; EYES;
D O I
10.5301/ejo.5000437
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: In a clinical setting, the usefulness of optical coherence tomography (007) is strongly dependent on reproducibility of the measurement. The aim of the present study was to evaluate macula and optic disc measurement reproducibility with the new spectral-domain OCT (SD-OCT) from Canon (Canon OCT-HS100) and to compare reproducibility and obtained measurements with the Zeiss Cirrus HD-OCT. Methods: Macula and optic disc parameters from the right eyes of 31 subjects were obtained twice with both instruments. Interoperator reproducibility was evaluated by use of the coefficient of repeatability (CR), and the obtained measurements were compared between the instruments. Results: No difference in interoperator reproducibility could be found when comparing the 2 instruments and reproducibility ranged from 3.94% to 12.77% for optic disc parameters and from 1.19% to 3.54% for macula parameters. The lowest reproducibility was found for cup volume and vertical cup/disc ratio with both instruments. For all macula and retinal nerve fiber layer (RNFL) thickness measurements, there was a statistical difference when comparing the 2 instruments, except for RFNL measurements of the superior quadrant, with the Canon OCT-HS100 always evaluating the thickness to be thicker; however, the 2 instruments correlated well. Conclusions: The Canon OCT-HS100 is a reproducible instrument that matches the Zeiss Cirrus HD-OCT well. It remains to be evaluated how sensitive the Canon OCT-HS100 is to detect small subtle changes in optic disc parameters and macular nerve fiber layer thickness. Furthermore, due to the differences in thickness estimation, it is important to emphasize that SD-OCTs are not interchangeable.
引用
收藏
页码:722 / 727
页数:6
相关论文
共 17 条
  • [1] Repeatability Using Automatic Tracing with Canon OCT-HS100 and Zeiss Cirrus HD-OCT 5000
    Brautaset, Rune
    Birkeldh, Ulrika
    Alstig, Petra Frehr
    Wiken, Petra
    Nilsson, Maria
    [J]. PLOS ONE, 2016, 11 (02):
  • [2] Reproducibility of Cirrus Optical Coherence Tomography (OCT) in Multiple Sclerosis
    Syc, Stephanie B.
    Warner, Christina V.
    Hiremath, Girish S.
    Farrell, Sheena K.
    Balcer, Laura J.
    Frohman, Elliot M.
    Calabresi, Peter A.
    [J]. NEUROLOGY, 2009, 72 (11) : A245 - A245
  • [3] Reproducibility of Central Corneal Thickness Measurements in Normal Eyes Using the Zeiss Cirrus 5000 HD-OCT and Pentacam HR
    Baghdasaryan, Elmira
    Huang, Xiwen
    Marion, Kenneth M.
    Tepelus, Tudor C.
    Bagherinia, Homayoun
    Sadda, SriniVas R.
    Hsu, Hugo Y.
    [J]. OPEN OPHTHALMOLOGY JOURNAL, 2018, 12 : 72 - 83
  • [4] Measurement of Optic Disc Cup Surface Depth Using Cirrus HD-OCT
    Kim, Young Kook
    Ha, Ahnul
    Lee, Won June
    Jeoung, Jin Wook
    Park, Ki Ho
    [J]. JOURNAL OF GLAUCOMA, 2017, 26 (12) : 1072 - 1080
  • [5] Comparison of macular thickness in diabetic patients acquired from optical coherence tomography mode and optical coherence tomography angiography mode in Cirrus HD-OCT 5000
    Wong, Ho-yin
    Ahmat, Ricky
    Zee, Benny Chung-ying
    Luk, Simon Chun-wa
    Cheing, Gladys Lai-ying
    Lam, Andrew Kwok-cheung
    [J]. JOURNAL OF OPTOMETRY, 2024, 17 (04)
  • [6] Comparison of Automated Analysis of Cirrus HD OCT Spectral-Domain Optical Coherence Tomography with Stereo Photographs of the Optic Disc
    Sharma, Ashish
    Oakley, Jonathan D.
    Schiffman, Joyce C.
    Budenz, Donald L.
    Anderson, Douglas R.
    [J]. OPHTHALMOLOGY, 2011, 118 (07) : 1348 - 1357
  • [7] Reproducibility of retinal thickness measurements using optical coherence tomography (OCT)
    Liebmann, JM
    Baumann, M
    Gentile, RC
    DeLuca, R
    Ozols, A
    Chew, SJ
    Ritch, R
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1996, 37 (03) : 4464 - 4464
  • [8] Development and evaluation of an R library to align macula and optic disc centered optical coherence tomography (OCT) scans
    Wang, Hui
    Rauscher, Franziska
    Martinez-Perez, M.
    Peschel, Thomas
    Eslami, Mohammad
    Kazeminasab, Saber
    Luo, Yan
    Shi, Min
    Tian, Yu
    Loeffler, Markus
    Kirsten, Toralf
    Wang, Mengyu
    Elze, Tobias
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [9] IN-VIVO IMAGING OF OPTIC DISC PITS USING OPTICAL COHERENCE TOMOGRAPHY (OCT)
    KRIVOY, D
    LIEBMANN, JM
    GENTILE, R
    ROSEN, R
    STEGMAN, Z
    WALSH, JB
    RITCH, R
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S208 - S208
  • [10] Reproducibility of foveal thickness measurements obtained by retinal thickness analyzer (RTA) compared to optical coherence tomography (OCT).
    Neubauer, AS
    Thiel, M
    Niederdellmann, C
    Bechmann, M
    Ulbig, MW
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S173 - S173