HAND-HELD LUMINOMETER WITH ECL-BASED BIOSENSOR FOR LACTATE DETERMINATION

被引:0
|
作者
Martinez-Olmos, A. [1 ]
Palma, A. J. [1 ]
Ballesta-Claver, J.
Valencia-Miron, M. C.
Capitan-Vallvey, L. F.
机构
[1] Univ Granada, Dept Elect & Tecnol Comp, E-18071 Granada, Spain
关键词
Electrochemiluminescence; Portable instrumentation; Potentiostat; Biosensor; Screen-printed device; Lactate determination; Saliva;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new hand-held luminometer for electrochemiluminescence (ECL)-based one-shot bioscnsor for lactate is described. The lactate recognition system is based on lactate oxidase and the transduction system consists of luminol, all reagents immobilized in a Methocel membrane. The measurement of ECL from a screen-printed electrode by a portable instrument designed and developed by the authors makes it possible to determine lactate concentration. The compositions of the membrane and reaction conditions have been optimized to obtain adequate sensitivity. The one-shot biosensor responds to lactate rapidly, with the typical ECL acquisition time being 3 min, with a linearized dependence whose dynamic range was from 9.10(-6) to 2.10(-3) M, a detection limit of 2.4.10(-6) M and a sensor-to-sensor reproducibility (relative standard deviation RSD) around 10% at the medium level of the range. The performance of the ECL one-shot biosensor and portable instrument was tested for the non invasive analysis of lactate in saliva, validating the results against a reference procedure.
引用
收藏
页码:343 / +
页数:2
相关论文
共 50 条
  • [1] A Portable Luminometer with a Disposable Electrochemiluminescent Biosensor for Lactate Determination
    Martinez-Olmos, Antonio
    Ballesta-Claver, Julio
    Palma, Alberto J.
    del Carmen Valencia-Miron, Maria
    Fermin Capitan-Vallvey, Luis
    [J]. SENSORS, 2009, 9 (10) : 7694 - 7710
  • [2] Hand-held biosensor developed for diagnostics
    不详
    [J]. MRS BULLETIN, 1998, 23 (01) : 9 - 10
  • [3] Validation of a hand-held lactate device in determination of blood lactate in critically injured patients
    Slomovitz, BM
    Lavery, RF
    Tortella, BJ
    Siegel, JH
    Bachl, BL
    Ciccone, A
    [J]. CRITICAL CARE MEDICINE, 1998, 26 (09) : 1523 - 1528
  • [4] Validation of a new hand-held lactate analyzer
    Lavery, RF
    Slomovitz, B
    Ciccone, A
    Tortella, BJ
    Siegel, JH
    [J]. CRITICAL CARE MEDICINE, 1997, 25 (07) : A156 - A156
  • [5] Evaluation of a hand-held lactate analyzer in dogs
    Thorneloe, Charlotte
    Bedard, Christian
    Boysen, Soren
    [J]. CANADIAN VETERINARY JOURNAL-REVUE VETERINAIRE CANADIENNE, 2007, 48 (03): : 283 - 288
  • [6] Validation of a Portable Hand-Held Lactate Analyzer for Determination of Blood Lactate in Patients on Antiretroviral Therapy in Uganda
    Kiragga, Agnes K.
    Ocama, Ponsiano
    Reynolds, Steven J.
    Kambugu, Andrew
    Ojiambo, Henry
    Castelnuovo, Barbara
    [J]. JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2008, 49 (05) : 564 - 566
  • [7] A hand-held electrochemiluminescence biosensor for detection of carcinoembryonic antigen
    Mohammadniaei, Mohsen
    Zhang, Ming
    Qin, Xianming
    Wang, Wentao
    Pia, Lorenza
    Gürbüz, Hakan
    Helalat, Seyed Hossein
    Naseri, Maryam
    Sun, Yi
    [J]. Talanta, 2024, 266
  • [8] A hand-held electrochemiluminescence biosensor for detection of carcinoembryonic antigen
    Mohammadniaei, Mohsen
    Zhang, Ming
    Qin, Xianming
    Wang, Wentao
    Pia, Lorenza
    Gurbuz, Hakan
    Helalat, Seyed Hossein
    Naseri, Maryam
    Sun, Yi
    [J]. TALANTA, 2024, 266
  • [9] Reliability and Accuracy of Six Hand-Held Blood Lactate Analysers
    Bonaventura, Jacinta M.
    Sharpe, Ken
    Knight, Emma
    Fuller, Kate L.
    Tanner, Rebecca K.
    Gore, Christopher J.
    [J]. JOURNAL OF SPORTS SCIENCE AND MEDICINE, 2015, 14 (01) : 203 - 214
  • [10] Lactate measurements in critically ill patients with a hand-held analyser
    Brinkert, W
    Rommes, JH
    Bakker, J
    [J]. INTENSIVE CARE MEDICINE, 1999, 25 (09) : 966 - 969