A national field study of quality assessment of CoaguChek point-of-care testing prothrombin time monitors
被引:32
|作者:
Meijer, Piet
[1
]
Kluft, Cornelis
[1
]
Poller, Leon
[1
]
van der Meer, Felix J. M.
[1
]
Keown, Michelle
[1
]
Ibrahim, Saied
[1
]
van den Besselaar, Anton M. H. P.
[1
]
Tripodi, Armando
[1
]
Jespersen, Jorgen
[1
]
机构:
[1] Univ Manchester, Fac Life Sci, EAA Cent Facil, Manchester M13 9PT, Lancs, England
关键词:
quality assessment;
CoaguChek monitors;
prothrombin time;
significant deviation;
D O I:
10.1309/6Q8DY5J1THA8BQG3
中图分类号:
R36 [病理学];
学科分类号:
100104 ;
摘要:
A system for quality assessment (QA) of the CoaguChek (Roche Diagnostics, Mannheim, Germany) point-of-care testing prothrombin time monitor has been developed by the European Concerted Action on Anticoagulation. Hitherto there has not been an adequate rapid method for CoaguChek QA. Sets of 5 certified international normalized ratio (INR) plasma samples were tested on 539 CoaguChek monitors by experienced staff at 9 Netherlands Thrombosis Centers and results compared with certified INR. A 15% or more deviation has been classified as significant deviation. Overall mean and certified INR values were similar, but 20.3% of participants showed a 15% or more deviation from the certified INR on at least I of the 5 QA plasma samples. Statistically significant differences in results with different lots of CoaguChek test strips were found. There is need for large scale QA of CoaguChek monitors. The importance of the 5 CoaguChek certified INR QA plasma samples being tested on a single occasion is demonstrated.