Non-enhanced ultrasound is not a satisfactory modality for measuring necrotic ablated volume after radiofrequency ablation of benign thyroid nodules: a comparison with contrast-enhanced ultrasound

被引:27
|
作者
Yan, Lin [1 ,2 ]
Luo, Yukun [1 ]
Xiao, Jing [1 ]
Lin, Lin [1 ]
机构
[1] Chinese Peoples Liberat Army Gen Hosp, Dept Ultrasound, Med Ctr 1, 28 Fuxing Rd, Beijing, Peoples R China
[2] Dalian Med Univ, Hosp 2, Hlth Management Ctr, 467 Zhongshan Rd, Dalian, Peoples R China
基金
中国国家自然科学基金;
关键词
Radiofrequency ablation; Thyroid nodule; Ultrasonography; Reproducibility of results; INTEROBSERVER VARIATION; REPORTING STANDARDS; CLINICAL-PRACTICE; THERMAL ABLATION; COLOR DOPPLER; GUIDELINES; CEUS; RECURRENCE; AGREEMENT; UPDATE;
D O I
10.1007/s00330-020-07398-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective To investigate the intra- and inter-observer reliability and agreement between gray-scale and Doppler ultrasound (US) and contrast-enhanced ultrasound (CEUS) in measuring ablated volume (Va) after radiofrequency ablation (RFA) for benign thyroid nodules. Methods A total of 173 patients with 190 benign thyroid nodules who underwent RFA were included in this study. After RFA, the total volume of ablated nodule was divided into Va and the incompletely treated vital volume. Patients were followed up at 1, 3, 6, 12 months, and every 12 months thereafter. Two independent observers measured Va using US and CEUS during the same follow-up visit. The intra- and inter-observer reliability of the two measurement modalities was assessed using intraclass correlation coefficient (ICC) with 95% confidence interval. The Bland-Altman analysis was used to evaluate agreement, which was expressed as a mean difference with 95% limits of agreement (LOA). Results The mean follow-up time was 23.17 +/- 12.70 months. Va measured by US was significantly larger than by CEUS (p < 0.001). The intra- and inter-observer reliability decreased over the follow-up period and became moderate in both subgroups at 12 months (all ICC < 0.75). The mean difference and LOA became larger and wider during the follow-up. The best agreement was found in nodules < 10 ml at 1 month with a mean difference of 1.166 and LOA between 0.413 and 3.294. Conclusions The intra- and inter-observer reliability and agreement of US and CEUS in measuring Va were unsatisfactory. CEUS should be considered when Va was needed for further evaluation or in the case of nodules with suspected regrowth.
引用
收藏
页码:3226 / 3236
页数:11
相关论文
共 50 条
  • [1] Non-enhanced ultrasound is not a satisfactory modality for measuring necrotic ablated volume after radiofrequency ablation of benign thyroid nodules: a comparison with contrast-enhanced ultrasound
    Lin Yan
    Yukun Luo
    Jing Xiao
    Lin Lin
    European Radiology, 2021, 31 : 3226 - 3236
  • [2] Contrast-enhanced ultrasound is a reliable and reproducible assessment of necrotic ablated volume after radiofrequency ablation for benign thyroid nodules: a retrospective study
    Yan, Lin
    Li, XinYang
    Xiao, Jing
    Li, YingYing
    Zhu, Yaqiong
    He, Hongying
    Luo, Yukun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2022, 39 (01) : 40 - 47
  • [3] Comparison of the effects of contrast-enhanced ultrasound and conventional ultrasound-guided radiofrequency ablation on benign thyroid nodules
    Xu, Shouxing
    He, Huiling
    Jiang, Meijuan
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2023, 39 (06) : 1661 - 1665
  • [4] Application of contrast-enhanced ultrasound in minimally invasive ablation of benign thyroid nodules
    Jiaying Cao
    Peili Fan
    Feihang Wang
    Shuainan Shi
    Lingxiao Liu
    Zhiping Yan
    Yi Dong
    Wenping Wang
    Journal of Interventional Medicine, 2022, (01) : 32 - 36
  • [5] Reproducibility of Ablated Volume Measurement Is Higher with Contrast-Enhanced Ultrasound than with B-Mode Ultrasound after Benign Thyroid Nodule Radiofrequency Ablation-A Preliminary Study
    Schiaffino, Simone
    Serpi, Francesca
    Rossi, Duccio
    Ferrara, Valerio
    Buonomenna, Ciriaco
    Ali, Marco
    Monfardini, Lorenzo
    Sconfienza, Luca Maria
    Mauri, Giovanni
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (05)
  • [6] Comparison of Micro- flow Imaging and Contrast-Enhanced Ultrasound in Ultrasound-Guided Microwave Ablation of Benign Thyroid Nodules
    Zhuang, Min
    Lu, Man
    Jiang, Zirui
    Liang, Yin
    Wang, Shishi
    Wang, Lu
    Li, Jiami
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2024, 50 (05): : 729 - 734
  • [7] Evaluation of the clinical efficacy of microwave ablation for benign thyroid nodules based on contrast-enhanced ultrasound
    Zhang, Yun
    Feng, Jun
    Fu, Gang
    MEDICINE, 2024, 103 (49)
  • [8] Contrast-enhanced ultrasound in the diagnosis of solitary thyroid nodules
    Zhang Yuan
    Jiang Quan
    Zhang Yunxiao
    Chen Jian
    He Zhu
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2015, 11 (01) : 41 - 45
  • [9] Value of Contrast-Enhanced Ultrasound in Mummified Thyroid Nodules
    Chen, Sijie
    Tang, Kui
    Gong, Yi
    Ye, Fei
    Liao, Liyan
    Li, Xiaodu
    Zhang, Qi
    Xu, Yan
    Zhang, Rongsen
    Niu, Chengcheng
    FRONTIERS IN ENDOCRINOLOGY, 2022, 13
  • [10] Utility of Contrast-Enhanced Ultrasound for Evaluation of Thyroid Nodules
    Zhang, Bo
    Jiang, Yu-Xin
    Liu, Ji-Bin
    Yang, Meng
    Dai, Qing
    Zhu, Qing-Li
    Gao, Pin
    THYROID, 2010, 20 (01) : 51 - 57