Anesthetic equipment fault leading to hypercapnia in a cat

被引:4
|
作者
Mattson, SF [1 ]
Kerr, CL [1 ]
Dyson, DH [1 ]
机构
[1] Univ Guelph, Ontario Vet Coll, Dept Clin Studies, Guelph, ON N1G 2W1, Canada
关键词
bain system; feline anesthesia; hypercapnia; nonrebreathing anesthetic system;
D O I
10.1111/j.1467-2987.2004.00136.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Spontaneous ventilation during positive pressure ventilation was observed in a 4-year-old DSH cat maintained under general anesthesia with isoflurane delivered with a nonrebreathing system. This was accompanied by an increase in heart rate and blood pressure. On investigation, neither an inadequate plane of anesthesia, nor hypoxemia, nor hyperthermia was present. The nonrebreathing system was replaced and the hypercapnia resolved. A defect in the inner fresh gas delivery tube of the Bain system was identified. A simple and quick test is described which can be performed to verify the integrity of the inner tube of the Bain breathing circuit.