Characterization of Standard CMOS Compatible Photodiodes and Pixels for Lab-on-Chip Devices

被引:0
|
作者
Koeklue, Goezen [1 ,2 ]
Etienne-Cummings, Ralph [3 ]
Leblebici, Yusuf [2 ]
De Micheli, Giovanni [1 ]
Carrara, Sandro [1 ]
机构
[1] Swiss Fed Inst Technol EPFL, LSI, Lausanne, Switzerland
[2] Swiss Fed Inst Technol EPFL, LSM, Lausanne, Switzerland
[3] Johns Hopkins Univ, Computat Sensory Motor Syst, Baltimore, MD USA
来源
2013 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS) | 2013年
基金
瑞士国家科学基金会;
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
High quality CMOS image sensors are of great importance for LoC -Lab-on-Chip devices based on optical measurements. The main target in these devices is to minimize the cost and area while achieving a good resolution. The performance parameters of image sensor pixels and CMOS compatible photodiodes depend on the size, type and the geometry of the photodiode layout and varies for each technology. In this study, we present a comparative analysis of CMOS compatible photodiode types at different areas. The results have shown n -well/ p -sub type photodiode with 5 x 5 mu m(2) diffusion area achieves the highest sensitivity (6 9 : 8 1 x 1 0(1 2) V.s(-1).cm(-2)/ W.cm(2)) and with 4 0 x 4 0 mu m(2) diffusion area, highest SNR -Signal-to-Noise Ratio (7 2 : 2 6 d B) at 630 nm, while the p(+)/n -well/rho-sub type photodiode with 4 0 x 4 0 mu m(2) diffusion area results in highest responsivity (0 : 4 6 6 A.cm(2) = W.cm(2) ) at the same wavelength.
引用
收藏
页码:1075 / 1078
页数:4
相关论文
共 50 条
  • [1] Microfluidic degasser for lab-on-chip devices
    Al-Adhami, Mustafa
    Gutierrez, Erick
    Andar, Abhay
    Rao, Govind
    Kostov, Iordan
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2019, 257
  • [2] Automated substance testing for lab-on-chip devices
    Lutz Kloke
    Katharina Schimek
    Sven Brincker
    Alexandra Lorenz
    Annika Jänicke
    Christopher Drewell
    Silke Hoffmann
    Mathias Busek
    Frank Sonntag
    Norbert Danz
    Christoph Polk
    Florian Schmieder
    Alexey Borchanikov
    Viacheslav Artyushenko
    Frank Baudisch
    Mario Bürger
    Reyk Horland
    Roland Lauster
    Uwe Marx
    BMC Proceedings, 7 (Suppl 6)
  • [3] The fabrication of lab-on-chip devices from fluoropolymers
    Taberham, A.
    Kraft, M.
    Mowlem, M.
    Morgan, H.
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2008, 18 (06)
  • [4] Lab-on-chip for Mechanical Characterization of Oocyte
    Pokrzywnicka, Aleksandra
    Lizanets, Danylo
    Walczak, Rafal
    Sniadek, Patrycja
    PROCEEDINGS OF THE 25TH INTERNATIONAL CONFERENCE MIXED DESIGN OF INTEGRATED CIRCUITS AND SYSTEM (MIXDES 2018), 2018, : 441 - 444
  • [5] A 0.18 μm CMOS capacitive detection lab-on-chip
    Ghafar-Zadeh, Ebrahim
    Sawan, Mohamad
    PROCEEDINGS OF THE IEEE 2007 CUSTOM INTEGRATED CIRCUITS CONFERENCE, 2007, : 165 - 172
  • [6] Integratable capacitive bubble counter for lab-on-chip devices
    Barbosa, Catia
    Silva, Carlos A.
    Dong, Tao
    2014 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2014, : 253 - 257
  • [7] Integration of Biosensors and Associated Electronics on Lab-on-Chip Devices
    Giannitsis, A. T.
    Parve, T.
    Min, M.
    ELEKTRONIKA IR ELEKTROTECHNIKA, 2011, (04) : 61 - 66
  • [8] Lab-on-Chip Devices: Gaining Ground Losing Size
    Romao, Veronica C.
    Martins, Sofia A. M.
    Germano, Jose
    Cardoso, Filipe A.
    Cardoso, Susana
    Freitas, Paulo P.
    ACS NANO, 2017, 11 (11) : 10659 - 10664
  • [9] Lab-on-Chip for Exosomes and Microvesicles Detection and Characterization
    Chiriaco, Maria Serena
    Bianco, Monica
    Nigro, Annamaria
    Primiceri, Elisabetta
    Ferrara, Francesco
    Romano, Alessandro
    Quattrini, Angelo
    Furlan, Roberto
    Arima, Valentina
    Maruccio, Giuseppe
    SENSORS, 2018, 18 (10)
  • [10] CMOS Dielectrophoretic Lab-on-Chip Platform for Manipulation and Monitoring of cells
    Park, Kyoungchul
    Kabiri, Shideh
    Sonkusale, Sameer
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 7530 - 7533