Clinical Evaluation of Normalized Metal Artifact Reduction in kVCT Using MVCT Prior Images (MVCT-NMAR) for Radiation Therapy Treatment Planning

被引:19
|
作者
Paudel, Moti Raj [1 ]
Mackenzie, Marc [1 ]
Fallone, B. Gino [1 ,2 ,3 ]
Rathee, Satyapal [1 ,3 ]
机构
[1] Univ Alberta, Dept Oncol, Edmonton, AB T6G 1Z2, Canada
[2] Univ Alberta, Dept Phys, Edmonton, AB T6G 1Z2, Canada
[3] Cross Canc Inst, Dept Med Phys, Edmonton, AB T6G 1Z2, Canada
关键词
D O I
10.1016/j.ijrobp.2014.02.040
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the metal artifacts in diagnostic kilovoltage computed tomography (kVCT) images of patients that are corrected by use of a normalized metal artifact reduction (NMAR) method with megavoltage CT (MVCT) prior images: MVCT-NMAR. Methods and Materials: MVCT-NMAR was applied to images from 5 patients: 3 with dual hip prostheses, 1 with a single hip prosthesis, and 1 with dental fillings. The corrected images were evaluated for visualization of tissue structures and their interfaces and for radiation therapy dose calculations. They were compared against the corresponding images corrected by the commercial orthopedic metal artifact reduction algorithm in a Phillips CT scanner. Results: The use of MVCT images for correcting kVCT images in the MVCT-NMAR technique greatly reduces metal artifacts, avoids secondary artifacts, and makes patient images more useful for correct dose calculation in radiation therapy. These improvements are significant, provided the MVCT and kVCT images are correctly registered. The remaining and the secondary artifacts (soft tissue blurring, eroded bones, false bones or air pockets, CT number cupping within the metal) present in orthopedic metal artifact reduction corrected images are removed in the MVCT-NMAR corrected images. A large dose reduction was possible outside the planning target volume (eg, 59.2 Gy to 52.5 Gy in pubic bone) when these MVCT-NMAR corrected images were used in TomoTherapy treatment plans without directional blocks for a prostate cancer patient. Conclusions: The use of MVCT-NMAR corrected images in radiation therapy treatment planning could improve the treatment plan quality for patients with metallic implants. (C) 2014 Elsevier Inc.
引用
收藏
页码:682 / 689
页数:8
相关论文
共 19 条
  • [1] Clinical Evaluation of Normalized Metal Artifact Reduction in KVCT Using MVCT Prior Images (MVCT-NMAR) Technique in Radiotherapy
    Paudel, M.
    MacKenzie, M.
    Fallone, B.
    Rathee, S.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 527 - 528
  • [2] Evaluation of normalized metal artifact reduction (NMAR) in kVCT using MVCT prior images for radiotherapy treatment planning
    Paudel, M. R.
    Mackenzie, M.
    Fallone, B. G.
    Rathee, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (08)
  • [3] Deformable registration of the planning image (KVCT) and daily treatment images (MVCT) for Adaptive Radiation Therapy
    Lu, W
    Chen, Q
    Kupelian, P
    Langen, K
    Meeks, S
    Ruchala, K
    Olivera, G
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2132 - 2132
  • [4] Deformable registration of the planning image (kVCT) and the daily images (MVCT) for adaptive radiation therapy
    Lu, Weiguo
    Olivera, Gustavo H.
    Chen, Quan
    Ruchala, Kenneth J.
    Haimerl, Jason
    Meeks, Sanford L.
    Langen, Katja M.
    Kupelian, Patrick A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (17): : 4357 - 4374
  • [5] Evaluation of Metal Artifact Reduction Using MVCT and Model Based Image Reconstruction
    Paudel, M.
    Kirvan, P.
    Fallone, B. G.
    Rathee, S.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] Evaluation of CT Images Created Using a New Metal Artifact Reduction Reconstruction Algorithm for Radiation Therapy Treatment Planning
    Niemkiewicz, J.
    Palmiotti, A.
    Miner, M.
    Stunja, L.
    Bergene, J.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 207 - 207
  • [7] Phantom Study of Dose Verification for Prostate Cancer Radiation Therapy in Presence of Hip Prosthesis Using KVCT and MVCT for Treatment Planning on Helical Tomotherapy, EMRT and Rapid ARC
    Saini, J.
    Kumar, T.
    Starin, R.
    Rakowski, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3327 - +
  • [8] Evaluation of Metal Artifact Reduction Technique for the Radiation Therapy Planning
    Jeong, K.
    Kuo, H.
    Ritter, J.
    Shen, J.
    Basavatia, A.
    Yaparpalvi, R.
    Kalnicki, S.
    Tome, W.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3645 - 3645
  • [9] Clinical Evaluations of a Novel Metal Artifact Reduction Technique for Treatment Planning in Radiation Therapy
    Li, H.
    Noel, C.
    Thorstad, W.
    Li, H.
    Yu, L.
    Low, D.
    Moore, K.
    Mutic, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3693 - 3693
  • [10] Evaluation of Metal Artifact Reduction Algorithms in Computed Tomography and Their Application to Radiation Therapy Treatment Planning
    Norris, H.
    Rangaraj, D.
    Kim, S.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3448 - 3448