Characterisation of wear areas on UHMWPE total knee replacement prostheses through study of their areal surface topographical parameters

被引:2
|
作者
Holland, Matthew [1 ]
Fleming, Leigh [1 ]
Walton, Karl [1 ]
Cerquiglini, Arianna [2 ]
Hothi, Harry [2 ]
Hart, Alister [2 ]
Skinner, John [2 ]
Bills, Paul [1 ]
机构
[1] Univ Huddersfield, EPSRC Future Metrol Hub, Huddersfield, W Yorkshire, England
[2] UCL, London Implant Retrieval Ctr, London, England
来源
基金
英国工程与自然科学研究理事会;
关键词
biometrology; total knee replacement; focus variation microscopy; surface measurement; FOCUS-VARIATION MICROSCOPY; COMPARING MOBILE BEARING; FIXED-BEARING; ARTHROPLASTY; METAANALYSIS; OUTCOMES; TKA; BEHAVIOR; PS; CR;
D O I
10.1088/2051-672X/aac071
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Total knee replacement is one of the most common elective surgeries in the world, and presents a number of challenges related to the wear of ultra-high molecular weight polyethylene (UHMWPE). This paper presents an analysis of the surface topographical properties of the worn and unworn condylar surfaces on a small cohort of both wear simulated and retrieved prostheses of varying designs. A number of measurement points were taken on each prostheses in a mixture of worn and unworn areas through the use of focus-variation microscopy (FVM), a non-contact method of surface measurement. Surface areal parameters were extracted from this data to analyse and search for patterns within the data. It was found that in general, worn implant surfaces appear to show smoother, less peak dominated surfaces than unworn area. It was also found that wear simulated and retrieved implants display similar characteristics of surface topography. In addition, variation was noted between different designs of TKR device, with posterior stabilised designs found to be peak dominated and cruciate retaining type implants being valley dominated.
引用
收藏
页数:13
相关论文
共 24 条
  • [1] THE INFLUENCE OF CONTACT STRESS ON THE WEAR OF UHMWPE FOR TOTAL REPLACEMENT HIP PROSTHESES
    BARBOUR, PSM
    BARTON, DC
    FISHER, J
    WEAR, 1995, 181 : 250 - 257
  • [2] COLD-EXTRUSION DEFORMATION OF UHMWPE IN TOTAL KNEE REPLACEMENT PROSTHESES
    CHANG, DCM
    GOH, JCH
    TEOH, SH
    BOSE, K
    BIOMATERIALS, 1995, 16 (03) : 219 - 223
  • [3] A study of the wear resistance of three types of clinically applied UHMWPE for total replacement hip prostheses
    Barbour, PSM
    Stone, MH
    Fisher, J
    BIOMATERIALS, 1999, 20 (22) : 2101 - 2106
  • [4] In vitro wear assessments of fixed and mobile UHMWPE total knee replacement
    Affatato, Saverio
    Bracco, Pierangiola
    Sudanese, Alessandra
    MATERIALS & DESIGN, 2013, 48 : 44 - 51
  • [5] Study on implantation of total knee replacement prostheses
    Cortes-Aburto, Obed
    Hernandez-Perez, Jose-Alfredo
    Rojas-Rodriguez, Rafael
    Arroyo-Diaz, Salvador-Antonio
    Aceves-Perez, Rita-Marina
    Canchola-Chavez, Marco-Antonio
    DYNA, 2017, 92 (05): : 522 - 524
  • [6] A New Method of In Vitro Wear Assessment of the UHMWPE Tibial Insert in Total Knee Replacement
    Affatato, Saverio
    Cristofolini, Luca
    Leardini, Walter
    Erani, Paolo
    Zavalloni, Mara
    Tigani, Domenico
    Viceconti, Marco
    ARTIFICIAL ORGANS, 2008, 32 (12) : 942 - 948
  • [7] A comparative surface topographical analysis of explanted total knee replacement prostheses: Oxidised zirconium vs cobalt chromium femoral components
    Kennard, Emma
    Scholes, Susan C.
    Sidaginamale, Raghavendra
    Gangadharan, Rajkumar
    Weir, David J.
    Holland, James
    Deehan, David
    Joyce, Thomas J.
    MEDICAL ENGINEERING & PHYSICS, 2017, 50 : 59 - 64
  • [8] Wear of ultra-high molecular weight polyethylene (UHMWPE) in total knee prostheses: a review of key influences
    McGloughlin, TM
    Kavanagh, AG
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H4) : 349 - 359
  • [9] Ultra high molecular-weight polyethylene (UHMWPE) in total knee replacement: Fabrication, sterilisation and wear
    Blunn, G
    del Preva, EMB
    Costa, L
    Fisher, J
    Freeman, MAR
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (07): : 946 - 949
  • [10] Influence of conformity on the wear of total knee replacement: An experimental study
    Brockett, Claire L.
    Carbone, Silvia
    Fisher, John
    Jennings, Louise M.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2018, 232 (02) : 127 - 134