Impact of Stirring on the Luminescent Dissolved Oxygen Sensor

被引:1
|
作者
Colt, John [1 ]
机构
[1] NOAA, Natl Marine Fisheries Serv, NW Fisheries Sci Ctr, Seattle, WA 98112 USA
关键词
D O I
10.1080/15222055.2014.893474
中图分类号
S9 [水产、渔业];
学科分类号
0908 ;
摘要
The impact of stirring on the response of the field YSI ProODO luminescence dissolved oxygen sensor was measured in a 300-mL biological oxygen demand (BOD) test bottle with a magnetic stirrer either on or off. With a water sample from a fish transport experiment, the dissolved oxygen dropped by 2-3 mg/L when the magnetic stirrer was turned off for 11 min. The initial rate of dissolved oxygen decline ranged from -0.39 to -1.04 mg/L per min for this water. The dissolved oxygen returned to a value projected from the oxygen uptake rate when the stirrer was turned on. A much smaller impact was observed for 18.2 M Omega high-purity water (mean: -0.11 mg/L). Acid-washing the BOD bottle/stir bar and cleaning the sensor cap reduced the decrease to -0.08 mg/L. Based on t-tests between pairs of stirring-on and stirring-off results, the lack of stirring had a significant effect on measured dissolved oxygen. The source of this effect and its potential impact on the accuracy of the optical probe under typical operating conditions remain to be determined.
引用
收藏
页码:242 / 244
页数:3
相关论文
共 50 条
  • [1] A NIR luminescent copolymer based on platinum porphyrin as high permeable dissolved oxygen sensor for microbioreactors
    Jin, Pengwei
    Chu, Ju
    Miao, Yu
    Tan, Jun
    Zhang, Siliang
    Zhu, Weihong
    AICHE JOURNAL, 2013, 59 (08) : 2743 - 2752
  • [2] Unsteady 2D measurement of dissolved oxygen distribution using luminescent sensor film
    Matsuda, Yu
    Nagashima, Fumihiro
    Yamaguchi, Hiroki
    Egami, Yasuhiro
    Niimi, Tomohide
    SENSORS AND ACTUATORS B-CHEMICAL, 2011, 160 (01): : 1464 - 1467
  • [3] Luminescent nanobeads for optical sensing and imaging of dissolved oxygen
    Borisov, Sergey M.
    Klimant, Ingo
    MICROCHIMICA ACTA, 2009, 164 (1-2) : 7 - 15
  • [4] Luminescent nanobeads for optical sensing and imaging of dissolved oxygen
    Sergey M. Borisov
    Ingo Klimant
    Microchimica Acta, 2009, 164 : 7 - 15
  • [5] Wearable, Luminescent Oxygen Sensor for Transcutaneous Oxygen Monitoring
    Lim, Chang-Jin
    Lee, Soyeon
    Kim, Jin-Hoon
    Kil, Hye-Jun
    Kim, Yu-Chan
    Park, Jin-Woo
    ACS APPLIED MATERIALS & INTERFACES, 2018, 10 (48) : 41026 - 41034
  • [6] A Dissolved Oxygen Analyzer with an Optical Sensor
    N. K. Zaitsev
    V. I. Dvorkin
    P. V. Melnikov
    A. E. Kozhukhova
    Journal of Analytical Chemistry, 2018, 73 : 102 - 108
  • [7] Labview Based Dissolved Oxygen Sensor
    Chu, Fenghong
    Yang, Junjie
    2009 INTERNATIONAL CONFERENCE ON ARTIFICIAL INTELLIGENCE AND COMPUTATIONAL INTELLIGENCE, VOL II, PROCEEDINGS, 2009, : 101 - 103
  • [8] A ratiometric sensor for dissolved oxygen in seawater
    Poteau, X
    MacCraith, B
    OPTO-IRELAND 2002: OPTICS AND PHOTONICS TECHNOLOGIES AND APPLICATIONS, PTS 1 AND 2, 2003, 4876 : 886 - 893
  • [9] Phase fluorometric dissolved oxygen sensor
    McDonagh, C
    Kolle, C
    McEvoy, AK
    Dowling, DL
    Cafolla, AA
    Cullen, SJ
    MacCraith, BD
    SENSORS AND ACTUATORS B-CHEMICAL, 2001, 74 (1-3) : 124 - 130
  • [10] A Dissolved Oxygen Analyzer with an Optical Sensor
    Zaitsev, N. K.
    Dvorkin, V. I.
    Melnikov, P. V.
    Kozhukhova, A. E.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2018, 73 (01) : 102 - 108