Signal-to-Noise Ratio Analysis of a Phase-Sensitive Voltmeter for Electrical Impedance Tomography

被引:26
|
作者
Murphy, Ethan K. [1 ]
Takhti, Mohammad [1 ]
Skinner, Joseph [1 ]
Halter, Ryan J. [1 ]
Odame, Kofi [1 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
Analog-to-digital converter (ADC); digital voltmeter; electrical impedance imaging; matched filter; random noise; SYSTEM;
D O I
10.1109/TBCAS.2016.2601692
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper, thorough analysis along with mathematical derivations of the matched filter for a voltmeter used in electrical impedance tomography systems are presented. The effect of the random noise in the system prior to the matched filter, generated by other components, are considered. Employing the presented equations allow system/circuit designers to find the maximum tolerable noise prior to the matched filter that leads to the target signal-to-noise ratio (SNR) of the voltmeter, without having to over-design internal components. A practical model was developed that should fall within 2 dB and 5 dB of the median SNR measurements of signal amplitude and phase, respectively. In order to validate our claims, simulation and experimental measurements have been performed with an analog-to-digital converter (ADC) followed by a digital matched filter, while the noise of the whole system was modeled as the input referred at the ADC input. The input signal was contaminated by a known value of additive white Gaussian noise (AWGN) noise, and the noise level was swept from 3% to 75% of the least significant bit (LSB) of the ADC. Differences between experimental and both simulated and analytical SNR values were less than 0.59 and 0.35 dB for RMS values >= 20% of an LSB and less than 1.45 and 2.58 dB for RMS values <20% of an LSB for the amplitude and phase, respectively. Overall, this study provides a practical model for circuit designers in EIT, and a more accurate error analysis that was previously missing in EIT literature.
引用
收藏
页码:360 / 369
页数:10
相关论文
共 50 条
  • [1] Digital phase-sensitive demodulator for Electrical Impedance Tomography
    Oh, TI
    Lee, JW
    Kim, KS
    Lee, JS
    Woo, EJ
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 1070 - 1072
  • [2] A Digital Phase-sensitive Detector for Electrical Impedance Tomography
    He, Chuanhong
    Zhang, Li
    Liu, Bin
    Xu, Zheng
    Zhang, Zhanlong
    2008 WORLD AUTOMATION CONGRESS PROCEEDINGS, VOLS 1-3, 2008, : 1769 - 1772
  • [3] DESIGN OF A PHASE-SENSITIVE DETECTOR TO MAXIMIZE SIGNAL-TO-NOISE RATIO IN THE PRESENCE OF GAUSSIAN WIDE-BAND NOISE
    SMITH, RWM
    FREESTON, IL
    BROWN, BH
    SINTON, AM
    MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (11) : 1054 - 1062
  • [4] Signal-to-noise ratio enhancement of phase-sensitive optical time-domain reflectometry based on power spectrum analysis
    Li, Qin
    Zhang, Chunxi
    Li, Lijing
    Zhong, Xiang
    OPTICAL ENGINEERING, 2014, 53 (02)
  • [5] Enhancement of Signal-to-Noise Ratio Based on Variational Mode Decomposition for Phase-Sensitive Optical Time Domain Reflectometry
    Gao Haotian
    Kang Jiehu
    Zhang, Zhen
    Wu, Bin
    ACTA OPTICA SINICA, 2023, 43 (21)
  • [6] Signal-to-noise ratio analysis and improvement for fluorescence tomography imaging
    Wang, Huiquan
    Feng, Xing
    Shi, Boya
    Liang, Wenxuan
    Chen, Yongping
    Wang, Jinhai
    Li, Xingde
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (09):
  • [7] Signal-to-noise ratio analysis of the computed tomography imaging spectrometer
    Volin, CE
    Descour, MR
    Dereniak, EL
    IMAGING SPECTROMETRY IV, 1998, 3438 : 107 - 113
  • [8] Analysis of in-plane signal-to-noise ratio in computed tomography
    Hara, Takanori
    Ichikawa, Katsuhiro
    Sanada, Shigeru
    Ida, Yoshihiro
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [9] Signal-to-Noise Ratio Enhancement Based on Empirical Mode Decomposition in Phase-Sensitive Optical Time Domain Reflectometry Systems
    Qin, Zengguang
    Chen, Hui
    Chang, Jun
    SENSORS, 2017, 17 (08):
  • [10] Improved Signal-to-Noise Ratio in Laboratory-Based Phase Contrast Tomography
    Boone, Matthieu N.
    De Witte, Yoni
    Dierick, Manuel
    Almeida, Ana
    Van Hoorebeke, Luc
    MICROSCOPY AND MICROANALYSIS, 2012, 18 (02) : 399 - 405