Evaluation of the SAS Legionella Test, a new immunochromatographic assay for the detection of Legionella pneumophila serogroup 1 antigen in urine

被引:15
|
作者
Diederen, B. M. W. [1 ]
Peeters, M. F. [1 ]
机构
[1] St Elizabeth Hosp, Lab Med Microbiol & Immunol, Tilburg, Netherlands
关键词
diagnosis; evaluation; Legionella pneumophila; SAS Legionella Test; urinary antigen test;
D O I
10.1111/j.1469-0691.2006.01587.x
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
This study evaluated a new immunochromatographic assay (SAS Legionella Test) for its ability to detect Legionella pneumophila serogroup 1 antigen in urine. Results were compared with those obtained using the Binax Now urinary antigen test. Sensitivity and specificity were estimated as 82.9% and 99.0%, respectively, for the SAS Legionella Test, and 91.4% and 100%, respectively, for the Binax Now urinary antigen test. The sensitivity of both tests increased to 97.1% (p 0.009) and 94.2% (p 0.7), respectively, if the tests were examined after 1 h.
引用
收藏
页码:86 / 88
页数:3
相关论文
共 50 条