A new physiological signal acquisition patch designed with advanced respiration monitoring algorithm based on 3-axis accelerator and gyroscope

被引:0
|
作者
Wang, Sikai [1 ,2 ]
Liu, Ming [1 ,2 ]
Pang, Bo [1 ,2 ]
Li, Peng [1 ,2 ]
Yao, Zhaolin [1 ,2 ]
Zhang, Xu [1 ,2 ]
Chen, Hongda [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst Semicond, State Key Lab Integrated Optoelect, Beijing 100083, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
SYSTEM;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In a gradually aging society, families and hospitals have a growing demand for reliable and unobtrusive physiological signal monitoring for elderly people. However, the existing respiration rate monitoring methods and algorithms are still unsatisfactory. In this work, we introduce a physiological signal acquisition patch which integrates 3-axis accelerator and 3-axis gyroscope to estimate respiration rate, as well as ECG(electrocardiogram) sensor and surface temperature sensor. A complete set of respiration rate estimation algorithms is embedded in our patch, which can be used to identify whether the patch is worn or not, and to recognize, segment, de-noise and reconstruct the respiration signal. In-situ experiments have been conducted to prove the validity of the algorithms described in this paper and the possibilities of estimating respiration rate using a physiological signal acquisition patch. The mean absolute error (MAE) is 0.11(about +/- 0.7 times in a minute), which is the least among similar studies that acquire respiratory rate from 3-axis accelerators or electrocardiogram.