Conductivity tomography based on pulsed eddy current with squid magnetometer

被引:0
|
作者
Panaitov, GI [1 ]
Krause, HJ [1 ]
Zhang, Y [1 ]
机构
[1] Inst Thin Films & Interfaces, Res Ctr, D-52425 Julich, Germany
关键词
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Pulsed eddy current (EC) techniques have the advantage of potentially covering a broader depth range than standard single frequency EC testing. We developed a novel pulsed EC technique using a liquid-nitrogen cooled SQUID magnetometer. For two reasons, SQUID magnetometers are particularly well suited as sensors: first they constitute an extremely sensitive magnetic field sensor, second they measure the field directly which decays more slowly than its time derivative picked up by induction coils. A square waveform transmitter signal was used, with alternating slopes in order to eliminate drift effect, and stacking synchronous to the power line frequency in order to improve signal-to-noise. The early time (high frequency) data of the recorded transient correspond to the upper layers of the conducting medium, while late time data or low frequencies deliver information on deep layers. Measurements of crack's at different depths in a stacked aluminum sample are presented. From the measured data, the apparent conductivity of the sample was calculated for each position and depth by applying a technique known from geophysical data interpretation. Thus, the position and depth of the crack was determined from the tomographic conductivity image of the sample.
引用
收藏
页码:445 / 452
页数:8
相关论文
共 50 条
  • [1] Conductivity tomography for non-destructive evaluation using pulsed eddy current with HTS SQUID magnetometer
    Krause, HJ
    Panaitov, GI
    Zhang, Y
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2003, 13 (02) : 215 - 218
  • [2] Pulsed eddy current transient technique with HTS SQUID magnetometer for non-destructive evaluation
    Panaitov, G
    Krause, HJ
    Zhang, Y
    PHYSICA C-SUPERCONDUCTIVITY AND ITS APPLICATIONS, 2002, 372 : 278 - 281
  • [3] Aircraft wheel testing with remote eddy current technique using a HTS SQUID magnetometer
    Hohmann, R
    Lomparski, D
    Krause, HJ
    Von Kreurzbruck, M
    Becker, W
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2001, 11 (01) : 1279 - 1282
  • [4] Characterization of the damage process in GLARE® 2 using an eddy current technique based on HTS-SQUID magnetometer
    Bonavolonta, C.
    Valentino, M.
    Pepe, G. P.
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 2007, 20 (01): : 51 - 56
  • [5] Multiplexed HTS rf SQUID magnetometer array for eddy current testing of aircraft rivet joints
    Gärtner, S
    Krause, HJ
    Wolters, N
    Lomparski, D
    Wolf, W
    Schubert, J
    von Kreutzbruck, M
    Allweins, K
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 21A & B, 2002, 615 : 520 - 527
  • [6] Detection of low-conductivity objects using eddy current measurements with an optical magnetometer
    Jensen, Kasper
    Zugenmaier, Michael
    Arnbak, Jens
    Staerkind, Hans
    Balabas, Mikhail, V
    Polzik, Eugene S.
    PHYSICAL REVIEW RESEARCH, 2019, 1 (03):
  • [7] Eddy current tomography using rotating magnetic fields for deep SQUID NDE
    Haller, A
    Tavrin, Y
    Krause, HJ
    David, P
    Braginski, AI
    SUPERCONDUCTOR SCIENCE & TECHNOLOGY, 1997, 10 (12): : 901 - 903
  • [8] Thickness and conductivity of metallic layers from pulsed eddy current measurements
    Tai, CC
    Rose, JH
    Moulder, JC
    REVIEW OF PROGRESS IN QUANTITATIVE NONDESTRUCTIVE EVALUATION, VOLS 15A AND 15B, 1996, 15 : 409 - 416
  • [9] Decoupling the Influence of Permeability and Conductivity in Pulsed Eddy-Current Measurements
    Adewale, Ibukun Dapo
    Tian, Gui Yun
    IEEE TRANSACTIONS ON MAGNETICS, 2013, 49 (03) : 1119 - 1127
  • [10] Hard magnetic material measurements in a pulsed field magnetometer considering coating and eddy current effects
    Kern, Alexander
    Leuning, Nora
    Hameyer, Kay
    ARCHIVES OF ELECTRICAL ENGINEERING, 2023, 72 (02) : 407 - 414