Design and Performance Investigation of a Multi-pinhole Collimator for a Small Field of View Gamma Imaging System

被引:3
|
作者
Bae, Jaekeon [1 ,2 ]
Bae, Seungbin [1 ,2 ]
Lee, Kisung [1 ,2 ]
Choi, Yong [3 ]
Kim, Yongkwon [4 ]
Joung, Jinhun [4 ]
机构
[1] Korea Univ, Dept Bioconvergence Engn, Seoul 136701, South Korea
[2] Korea Univ, Dept Med Devices, Guro Hosp, Seoul 152703, South Korea
[3] Sogang Univ, Dept Elect Engn, Seoul 121742, South Korea
[4] NuCare Med Syst Inc, Inchon 406840, South Korea
基金
新加坡国家研究基金会;
关键词
Limited angle system; Nuclear medicine; Multi-pinhole collimator;
D O I
10.3938/jkps.64.970
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The aim of this study is to design a collimator for a gamma imaging system that has a smallfootprint, a fast-scan-time, and a organ-specific applicability. To achieve such features, the collimator must have a high resolution, a high sensitivity, and a gantry with a simple geometry. For ensuring high resolution and high sensitivity, we designed a multi-pinhole collimator. For realizing a simplified gantry, we carried out studies with limited angle reconstruction. On the designed multi-pinhole collimator has eight-pinholes, whose diameters are 2 mm. Limited-angle reconstruction has been conducted with angle intervals of 3, 6, 9, and 12 degrees and with an angle coverage of 60, 90, 120, 150, and 180 degrees. The reconstruction of an image was separately developed based on the ray-driven and voxel-driven methods in order to overcome the sampling problem and to reduce the amount of computation required. To evaluate the performance of the designed system, performed studies on three spherical phantoms and a heart phantom by using the Geant4 application for tomographic emission (GATE) simulation tool. The results showed that the full width at half maximum (FWHM) of the center source were 6.25 mm and 7.18 mm for single-pinhole and multi-pinhole collimators, respectively. Moreover, limited angle reconstruction resulted in a higher efficiency of the imaging system because it overcame the limitation of the gantry geometry. Limited-angle reconstruction was optimized at an angle coverage of 120. with an angle interval of 6 degrees, then reconstructed image was shown a 12.18 mm FWHM. This suggests that the designed system needs only one-third the number of projections to acquire a reconstructed image with a slight degradation in image quality. This also suggests that our proposed multi-pinhole collimator is suitable for applications requiring a small-footprint, a fast-scan-time, and organ-specificity.
引用
收藏
页码:970 / 975
页数:6
相关论文
共 50 条
  • [1] Design and performance investigation of a multi-pinhole collimator for a small field of view gamma imaging system
    Jaekeon Bae
    Seungbin Bae
    Kisung Lee
    Yong Choi
    Yongkwon Kim
    Jinhun Joung
    Journal of the Korean Physical Society, 2014, 64 : 970 - 975
  • [2] An Investigation to Design High Performance Multi-Pinhole Collimator
    Min, Byung Jun
    Choi, Yong
    Lee, Nam-Yong
    Song, Tae Yong
    Jung, Jin Ho
    Hong, Key Jo
    Ahn, Young Bok
    Joung, Jinhun
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 3004 - 3007
  • [3] System Design and Performance Evaluation for Cardiac SPECT Imaging with Multi-pinhole Collimator
    Sun L.
    Lyu Z.
    Hou Y.
    Liu W.
    Jiang N.
    Liu H.
    Ma T.
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2021, 55 : 407 - 413
  • [4] Two-Level Multi-Pinhole Collimator for SPECT Imaging Using a Small-Field-of-View Gamma Camera
    Bae, Jaekeon
    Bae, Seungbin
    Lee, Soo-young
    Lee, Kisung
    Kim, Yongkwon
    Joung, Jinhun
    Kim, MinHo
    Kim, Kyeong Min
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 70 (02) : 192 - 200
  • [5] Two-level multi-pinhole collimator for SPECT imaging using a small-field-of-view gamma camera
    Jaekeon Bae
    Seungbin Bae
    Soo-young Lee
    Kisung Lee
    Yongkwon Kim
    Jinhun Joung
    MinHo Kim
    Kyeong Min Kim
    Journal of the Korean Physical Society, 2017, 70 : 192 - 200
  • [6] Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera
    Jaekeon Bae
    Seungbin Bae
    Young-Jun Jung
    Kisung Lee
    Yongkwon Kim
    Jinhun Joung
    Kyeong Min Kim
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2017, 70 : 416 - 423
  • [7] Experimental evaluation of a multi-pinhole collimator for a small organ by using a small-field-of-view gamma camera
    Bae, Jaekeon
    Bae, Seungbin
    Jung, Young-Jun
    Lee, Kisung
    Kim, Yongkwon
    Joung, Jinhun
    Kim, Kyeong Min
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 70 (04) : 416 - 423
  • [8] Design of a multi-pinhole collimator for high performance clinical and preclinical molecular imaging
    Mok, Greta
    Tsui, Benjamin
    Wang, Yi-Xiang
    Choi, Frankie
    Yip, Gabriel
    Ahuja, Anil
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [9] Evaluation of a Multi-pinhole Collimator for Imaging Small Animals with Different Sizes
    Mok, Greta S. P.
    Yu, Jianhua
    Du, Yong
    Wang, Yuchuan
    Tsui, Benjamin M. W.
    MOLECULAR IMAGING AND BIOLOGY, 2012, 14 (01) : 60 - 69
  • [10] Evaluation of a Multi-pinhole Collimator for Imaging Small Animals with Different Sizes
    Greta S. P. Mok
    Jianhua Yu
    Yong Du
    Yuchuan Wang
    Benjamin M. W. Tsui
    Molecular Imaging and Biology, 2012, 14 : 60 - 69