Radiofrequency ablation:: In vivo comparison of four commercially available devices in pig livers

被引:159
|
作者
Pereira, PL
Trübenbach, J
Schenk, M
Subke, J
Kroeber, S
Schaefer, I
Remy, CT
Schmidt, D
Brieger, J
Claussen, CD
机构
[1] Univ Tubingen, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Univ Tubingen, Sect Expt Radiol, D-72076 Tubingen, Germany
[3] Univ Tubingen, Dept Gen Surg, D-72076 Tubingen, Germany
[4] Univ Tubingen, Dept Forens Surg, D-72076 Tubingen, Germany
[5] Univ Tubingen, Dept Pathol, D-72076 Tubingen, Germany
关键词
animals; experimental study; liver; interventional procedures; radiofrequency (RF) ablation;
D O I
10.1148/radiol.2322030184
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare in vivo coagulation necrosis obtained with four radiofrequency (RF) ablation devices, to determine shape and reproducibility of induced coagulation by means of three-dimensional measurements of the ablation zone, and to achieve representations of the coagulated areas in three-dimensional spaces. MATERIALS AND METHODS: Four commercially available RF devices (perfusion, internally cooled cluster, and nine- and 12-tine expandable electrodes) that represent the most widely used systems on the market were tested. Sixteen in vivo ablation procedures were performed in porcine livers (four ablations for each RF system). After macroscopic and histopathologic analyses of 3-mm-thick liver sections, morphometric and volumetric findings in the central zone of white coagulation necrosis were assessed. Coagulation volume, diameter, length, and shape were determined digitally. After analysis of variance, measurements with each system were tested with the Tukey post hoc test. RESULTS: Mean coagulation volumes were 31.5 cm(3) +/- 15.8 (SD) for the perfusion electrode, 20.5 cm(3) +/- 2.6 for the cluster electrode, 16.2 cm(3) +/- 7.3 for the 12-tine electrode, and 9.8 cm(3) +/- 3.2 for the nine-tine electrode (P < .05, perfusion vs nine-tine electrode). No significant differences were observed regarding the mean short axis perpendicular to the needle shaft: 2.30 cm +/- 0.94, 3.04 cm +/- 0.26, 3.44 cm +/- 0.21, and 2.70 cm +/- 0.76, respectively. Variation coefficients were 0.50, 0.13, 0.45, and 0.33, respectively. CONCLUSION: Larger coagulation volumes were obtained with the perfusion and internally cooled cluster devices. More spherical volumes of ablation were achieved with the 12-tine and cluster electrodes. The former proved superior with regard to the short axis perpendicular to the needle shaft. The cluster and nine-tine electrode produced better reproducibility, which is suggestive of improved predictability of the extent of coagulation with these systems.
引用
收藏
页码:482 / 490
页数:9
相关论文
共 50 条
  • [1] In vivo comparison of the size of thermal injuries produced by three commercially available radiofrequency (RF) ablation devices
    Dodd, GD
    Napier, DH
    Hubbard, LG
    Chintapalli, KN
    Chopra, S
    Halff, GA
    [J]. RADIOLOGY, 2000, 214 (02) : 609 - 609
  • [2] In Vivo Evaluation of a Novel Radiofrequency Ablation Electrode in Pig Livers
    Cho, Sung-Hyun
    Oh, Dongwook
    Huh, Gunn
    Song, Tae-Jun
    Seo, Dong-Wan
    [J]. APPLIED SCIENCES-BASEL, 2022, 12 (13):
  • [3] Radiofrequency tissue ablation of the kidney: Ex vivo comparison of four different devices
    Haecker, A
    Vallo, S
    Thomas, K
    Wendt-Nordahl, G
    Alken, P
    Michel, MS
    [J]. JOURNAL OF UROLOGY, 2005, 173 (04): : 267 - 267
  • [4] Radiofrequency ablation for hepatocellular carcinoma: A prospective comparison of four radiofrequency devices
    Lin, Shi-Ming
    Lin, Chen-Chun
    Chen, Wei-Ting
    Chen, Yi-Chen
    Hsu, Chao-Wei
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2007, 18 (09) : 1118 - 1125
  • [5] EUS-guided pancreatic radiofrequency ablation: preclinical comparison of two currently available devices in a pig model
    Barret, Maximilien
    Leblanc, Sarah
    Rouquette, Alexandre
    Chaussade, Stanislas
    Terris, Benoit
    Prat, Frederic
    [J]. ENDOSCOPY INTERNATIONAL OPEN, 2019, 7 (02) : E138 - E143
  • [6] A comparison of microwave ablation and bipolar radiofrequency ablation both with an internally cooled probe: Results in ex vivo and in vivo porcine livers
    Yu, Jie
    Liang, Ping
    Yu, Xiaoling
    Liu, Fangyi
    Chen, Lei
    Wang, Yang
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2011, 79 (01) : 124 - 130
  • [7] Comparison of two commercially available cryoballoon ablation systems
    Diofano, F.
    Pflieger, C.
    Teumer, Y.
    Bothner, C.
    Rottbauer, W.
    Dahme, T.
    Weinmann, K.
    [J]. EUROPEAN HEART JOURNAL, 2023, 44
  • [8] Objective Comparison of Commercially Available Breast Implant Devices
    Henderson, Peter W.
    Nash, David
    Laskowski, Marta
    Grant, Robert T.
    [J]. AESTHETIC PLASTIC SURGERY, 2015, 39 (05) : 724 - 732
  • [9] Objective Comparison of Commercially Available Breast Implant Devices
    Peter W. Henderson
    David Nash
    Marta Laskowski
    Robert T. Grant
    [J]. Aesthetic Plastic Surgery, 2015, 39 : 724 - 732
  • [10] Comparison of the accuracy of contact force measurement in four commercially available force-sensing ablation catheters
    Kueffer, T.
    Haeberlin, A.
    Knecht, S.
    Baldinger, S. H.
    Servatius, H.
    Madaffari, A.
    Seiler, J.
    Muhl, A.
    Franzeck, F.
    Asatryan, B.
    Noti, F.
    Tanner, H.
    Roten, L.
    Reichlin, T.
    [J]. SWISS MEDICAL WEEKLY, 2022, 152 : 16S - 16S