Novel internal analysis of metal irrigation/aspiration tips could explain mechanisms of posterior capsule rupture

被引:2
|
作者
Lockington, David [1 ]
Macente, Alice [2 ]
Marrocco, Tiziana [3 ]
Sillars, Fiona [3 ]
机构
[1] Gartnavel Royal Hosp, Tennent Inst Ophthalmol, 1053 Great Western Rd, Glasgow G12 0YN, Lanark, Scotland
[2] Univ Glasgow, Sch Geog & Earth Sci, Glasgow G12 8RZ, Lanark, Scotland
[3] Univ Strathclyde, Dept Mech & Aerosp Engn, Adv Mat Res Lab, Glasgow G1 1XJ, Lanark, Scotland
关键词
CATARACT-SURGERY; ASPIRATION HANDPIECE; BIMANUAL TECHNIQUE; PHACOEMULSIFICATION; COMPLICATIONS; STRATIFICATION; IRRIGATION; MANAGEMENT; OUTCOMES; REMOVAL;
D O I
10.1038/s41433-020-01181-7
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Introduction Posterior capsule rupture (PCR) rates are used to measure cataract surgeons' quality. We wished to evaluate the internal non-visible surfaces of metal irrigation/aspiration (I/A) tips to identify potential mechanisms for PCR via novel metallographic imaging. Methods Ten metal I/A instruments underwent metallographic preparation by fine sectioning to expose inner surfaces near the aspiration opening. Analysis of inner bore, lumen, and opening aperture of metal aspiration tips was performed by optical microscopy, scanning electron microscopy (SEM), and 3D volume X-ray computational tomography (XCT). Distances from external aperture to first sharp metal surface were obtained and compared with a silicone-tipped instrument. Results We identified metal burrs near the aspiration apertures and manufacturing defects within all tips. XCT confirmed optical and SEM findings of significant defects and metal irregularities within aspiration tips. Samples also showed variation in lumen size/thickness, rough surfaces and material inhomogeneity, most pronounced at the internal tip. Median distance from outer aperture opening to first metal burr was 30 microns (range 10-120) and to internal tip irregularity (manufacturing flaw) was 250 microns (range 100-350). By comparison, distance to metal from the silicone outer aperture opening was 850 microns. Conclusions We have demonstrated the hidden sharp metallic irregularities within commonly used metal I/A tips. If an aspirated capsule encounters these sharp metal flaws, PCR could result. Minimising this risk would require lengthening potential distance between capsule and bare metal (as with polymer/silicone tips). Our study provides unique imaging evidence endorsing this principle and illustrates a hidden mechanism contributing to PCR.
引用
收藏
页码:1915 / 1921
页数:7
相关论文
共 2 条
  • [1] Novel internal analysis of metal irrigation/aspiration tips could explain mechanisms of posterior capsule rupture
    David Lockington
    Alice Macente
    Tiziana Marrocco
    Fiona Sillars
    Eye, 2021, 35 : 1915 - 1921
  • [2] Electron microscopy of silicone irrigation/aspiration tips involved in posterior capsule rupture
    Hart, John C., Jr.
    Tanner, Michael
    Rooney, David M.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2018, 44 (12): : 1517 - 1520