Fabrication of a flexible wireless pressure sensor for intravascular blood pressure monitoring

被引:27
|
作者
Yeh, Chao-Chi [1 ]
Lo, Shao-Hsiang [1 ]
Xu, Ming-Xin [1 ]
Yang, Yao-Joe [1 ,2 ]
机构
[1] Natl Taiwan Univ, Dept Mech Engn, Taipei, Taiwan
[2] Natl Appl Res Labs, Taiwan Instrument Res Inst, Taipei, Taiwan
关键词
Capacitor pressure sensor; Wireless sensing; L-C resonator; Flexible sensor; Blood pressure monitoring; PARYLENE-C;
D O I
10.1016/j.mee.2019.04.009
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This work presents the development of a wireless passive pressure sensor for cardiovascular pressure monitoring. The device is designed to be inserted into a stent that can be deployed within a blood vessel using a catheter-based delivery system. By utilizing micromachining techniques, a variable capacitor and an inductor were implemented in the sensor as an L-C resonator for wireless pressure sensing and signal retrieving. The sensor employs biocompatible polymeric material, parylene-C, as the substrate material. The pressure responses of the sensor, such as the sensitivity, linearity, and hysteresis, were measured and discussed. The sensitivity was approximately 6.19 x 10(-2)/kPa (8.25 x 10(-3)/mmHg), and the linearity over the ranges of 0-6 kPa (0-50 mmHg) and 6-33 kPa (50-250 mmHg) were approximately 91.5% and 87.6%, respectively. The wireless pressure sensing performance of the sensor was evaluated using the phase-dip technique. Additionally, the measured results indicate that the resonant frequency of the resonator shifted from 211 MHz to 182 MHz, as the pressure varied from 0 to 33 kPa (250 mmHg). The maximum standard error of the mean is 0.86 MHz. Finally, the influence of coil thickness on the wireless readout performance was investigated.
引用
收藏
页码:55 / 61
页数:7
相关论文
共 50 条
  • [1] Fabrication of an Integrated, Flexible, Wireless Pressure Sensor Array for the Monitoring of Ventricular Pressure
    Hernandez-Sebastian, Natiely
    Diaz-Alonso, Daniela
    Barrientos-Garcia, Bernardino
    Renero-Carrillo, Francisco Javier
    Calleja-Arriaga, Wilfrido
    MICROMACHINES, 2024, 15 (12)
  • [2] Flexible and wearable capacitive pressure sensor for blood pressure monitoring
    Bijender
    Kumar, Ashok
    SENSING AND BIO-SENSING RESEARCH, 2021, 33
  • [3] Flexible mesostructured capacitive pressure sensor for blood pressure monitoring
    Kumar, Shubham
    Yadav, Sanjay
    Kumar, Ashok
    2023 IEEE SENSORS, 2023,
  • [4] Design and Fabrication of a New Wearable Pressure Sensor for Blood Pressure Monitoring
    Ion, Marian
    Dinulescu, Silviu
    Firtat, Bogdan
    Savin, Mihaela
    Ionescu, Octavian N.
    Moldovan, Carmen
    SENSORS, 2021, 21 (06) : 1 - 19
  • [5] Flexible LC Sensor Array for Wireless Multizone Pressure Monitoring
    Zhang, Yunfan
    Tang, Wuyang
    Chen, Hongbin
    Li, Hui
    Xue, Zifan
    Shen, Shengnan
    IEEE SENSORS JOURNAL, 2024, 24 (03) : 2628 - 2636
  • [6] FOLDABLE POLYMER STENT INTEGRATED WITH WIRELESS PRESSURE SENSOR FOR BLOOD PRESSURE MONITORING
    Oyunbaatar, Nomin-Erdene
    Lee, Dong-Weon
    2023 IEEE 36TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, MEMS, 2023, : 61 - 64
  • [7] Fabrication of a Flexible Graphene Pressure Sensor
    Jung, Hyojin
    Chun, Sungwoo
    Kim, Youngjun
    Oh, Hyeong Sik
    Bae, Gi Yoon
    Bae, Giyeol
    Park, Wanjun
    JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2015, 15 (11) : 9020 - 9024
  • [8] Flexible wireless pressure sensor module
    Shin, KH
    Moon, CR
    Lee, TH
    Lim, CH
    Kim, YJ
    SENSORS AND ACTUATORS A-PHYSICAL, 2005, 123-24 : 30 - 35
  • [9] Thin-Film Flexible Wireless Pressure Sensor for Continuous Pressure Monitoring in Medical Applications
    Farooq, Muhammad
    Iqbal, Talha
    Vazquez, Patricia
    Farid, Nazar
    Thampi, Sudhin
    Wijns, William
    Shahzad, Atif
    SENSORS, 2020, 20 (22) : 1 - 22
  • [10] GRAPHENE-BASED WIRELESS INLINE PRESSURE SENSOR FOR IN VIVO BLOOD PRESSURE MONITORING
    Inoue, Nagisa
    Onoe, Hiroaki
    2019 IEEE 32ND INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2019, : 720 - 722