Fixed-flexion knee radiography using a new positioning device produced highly repeatable measurements of joint space width: ELSA-Brasil Musculoskeletal Study (ELSA-Brasil MSK)

被引:0
|
作者
Telles, Rosa Weiss [1 ]
Costa-Silva, Luciana [1 ]
Machado, Luciana A. C. [1 ]
Padilha dos Reis, Rodrigo Citton [1 ]
Barreto, Sandhi Maria [1 ]
机构
[1] Univ Fed Minas Gerais, Ctr Invest ELSA Brasil Minas Gerais, Fac Med, Belo Horizonte, MG, Brazil
关键词
Fixed-flexion radiography; Joint space width; Repeatability; Knee; Osteoarthritis; TEST-RETEST RELIABILITY; OSTEOARTHRITIS; SENSITIVITY; REPRODUCIBILITY; PERFORMANCE; PROGRESSION; AGREEMENT; ALIGNMENT; HEALTH; SCALE;
D O I
10.1016/j.rbr.2016.10.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To describe the performance of a non-fluoroscopic fixed-flexion PA radiographic protocol with a new positioning device, developed for the assessment of knee osteoarthritis (OA) in ELSA-Brasil MSK. Material and methods: A test-retest design including 19 adults (38 knee images) was conducted. Feasibility of the radiographic protocol was assessed by image quality parameters and presence of radioanatomic alignment according to intermargin distance (IMD) values. Repeatability was assessed for IMD and joint space width (JSW) measured at three different locations. Results: Approximately 90% of knee images presented excellent quality. Frequencies of nearly perfect radioanatomic alignment (IMD <1 mm) ranged from 29% to 50%, and satisfactory alignment was found in up to 71% and 76% of the images (IMD <= 1.5 mm and <= 1.7 mm, respectively). Repeatability analyses yielded the following results: IMD [SD of mean difference = 1.08; coefficient of variation (%CV) = 54.68%; intraclass correlation coefficient (ICC) (95%0) = 0.59 (0.34-0.77)1; JSW [SD of mean difference = 0.34-0.61; %CV = 4.48% - 9.80%; ICC (95%CI)= 0.74 (0.55-0.85) -0.94 (0.87-0.97)]. Adequately reproducible measurements of IMD and JSW were found in 68% and 87% of the images, respectively. Conclusions: Despite the difficulty in achieving consistent radioanatomic alignment between subsequent radiographs in terms of IMD, the protocol produced highly repeatable JSW measurements when these were taken at midpoint and 10 mm from the medial extremity of the medial tibial plateu. Therefore, measurements of JSW at these locations can be considered adequate for the assessment of knee OA in ELSA-Brasil MSK. (C) 2016 Published by Elsevier Editora Ltda.
引用
收藏
页码:154 / 161
页数:8
相关论文
共 2 条
  • [1] REPEATABILITY AND CORRELATIONS OF TIBIAL PLATEAU ALIGNMENT AND JOINT SPACE WIDTH IN NON-FLUOROSCOPIC FIXED-FLEXION KNEE RADIOGRAPHS: THE ELSA-BRASIL MUSCULOSKELETAL STUDY (ELSA-BRASIL MSK)
    Telles, R. W.
    Silva, L. C.
    Machado, L. A.
    Reis, R. C.
    Barreto, S. M.
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2016, 24 : S295 - S295
  • [2] EXTERNAL VALIDATION OF A PRE-TRAINED CONVOLUTIONAL NEURAL NETWORK FOR OSTEOARTHRITIS DIAGNOSIS IN KNEE RADIOGRAPHS OF THE BRAZILIAN LONGITUDINAL STUDY OF ADULT HEALTH MUSCULOSKELETAL - ELSA-BRASIL MSK
    Domingues, Julio Guerra
    Costa-Silva, Luciana
    Machado, Alexei M.
    Machado, Luciana A.
    Barreto, Sandhi M.
    Telles, Rosa Weiss
    [J]. OSTEOARTHRITIS AND CARTILAGE, 2024, 32 : S70 - S71