An automated microfluidic system for efficient capture of rare cells and rapid flow-free stimulation

被引:12
|
作者
Dettinger, Philip [1 ]
Wang, Weijia [1 ]
Ahmed, Nouraiz [1 ]
Zhang, Yang [1 ]
Loeffler, Dirk [1 ]
Kull, Tobias [1 ]
Etzrodt, Martin [1 ]
Lengerke, Claudia [2 ]
Schroeder, Timm [1 ]
机构
[1] Swiss Fed Inst Technol, Dept Biosyst Sci & Engn, Basel, Switzerland
[2] Univ Basel, Dept Biomed, Basel, Switzerland
基金
瑞士国家科学基金会;
关键词
HEMATOPOIETIC STEM; LINEAGE CHOICE; CULTURE; PROLIFERATION; ACTIVATION; GENERATION; TRACKING; PU.1;
D O I
10.1039/d0lc00687d
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Cell fates are controlled by environmental stimuli that rapidly change the activity of intracellular signaling. Studying these processes requires rapid manipulations of micro-environmental conditions while continuously observing single cells over long periods of time. Current microfluidic devices are unable to simultaneously i) efficiently capture and concentrate rare cells, ii) conduct automated rapid media exchanges via diffusion without displacing non-adherent cells, and iii) allow sensitive high-throughput long-term time-lapse microscopy. Hematopoietic stem and progenitor cells pose a particular challenge for these types of experiments as they are impossible to obtain in very large numbers and are displaced by the fluid flow usually used to change culture media, thus preventing cell tracking. Here, we developed a programmable automated system composed of a novel microfluidic device for efficient capture of rare cells in independently addressable culture chambers, a custom incubation system, and user-friendly control software. The chip's culture chambers are optimized for efficient and sensitive fluorescence microscopy and their media can be individually and quickly changed by diffusion without non-adherent cell displacement. The chip allows efficient capture, stimulation, and sensitive high-frequency time-lapse observation of rare and sensitive murine and human primary hematopoietic stem cells. Our 3D-printed humidification and incubation system minimizes gas consumption, facilitates chip setup, and maintains stable humidity and gas composition during long-term cell culture. This approach now enables the required continuous long-term single-cell quantification of rare non-adherent cells with rapid environmental manipulations, e.g. of rapid signaling dynamics and the later stem cell fate choices they control.
引用
收藏
页码:4246 / 4254
页数:9
相关论文
共 50 条
  • [1] Development of a flow-free magnetic actuation platform for an automated microfluidic ELISA
    Coarsey, Chad
    Coleman, Benjamin
    Kabir, Md Alamgir
    Sher, Mazhar
    Asghar, Waseem
    RSC ADVANCES, 2019, 9 (15): : 8159 - 8168
  • [2] Microfluidic device to study flow-free chemotaxis of swimming cells
    Garcia-Seyda, Nicolas
    Aoun, Laurene
    Tishkova, Victoria
    Seveau, Valentine
    Biarnes-Pelicot, Martine
    Bajenoff, Marc
    Valignat, Marie-Pierre
    Theodoly, Olivier
    LAB ON A CHIP, 2020, 20 (09) : 1639 - 1647
  • [3] Chemotactic Responses of Jurkat Cells in Microfluidic Flow-Free Gradient Chambers
    Sonmez, Utku M.
    Wood, Adam
    Justus, Kyle
    Jiang, Weijian
    Syed-Picard, Fatima
    LeDuc, Philip R.
    Kalinski, Pawel
    Davidson, Lance A.
    MICROMACHINES, 2020, 11 (04)
  • [4] Chemotaxis of Immune Cells in Microfluidic Flow-Free Concentration Gradient Generator
    Sonmez, Utku M.
    LeDuc, Philip R.
    Kalinski, Pawel
    Davidson, Lance A.
    BIOPHYSICAL JOURNAL, 2018, 114 (03) : 217A - 217A
  • [5] Flow-Free Microfluidic Device for Quantifying Chemotaxis in Spermatozoa
    Berendsen, Johanna T. W.
    Kruit, Stella A.
    Atak, Nihan
    Willink, Ellen
    Segerink, Loes I.
    ANALYTICAL CHEMISTRY, 2020, 92 (04) : 3302 - 3306
  • [6] Microfluidic flow-free generation of chemical concentration gradients
    Zhou, Yao
    Lin, Qiao
    SENSORS AND ACTUATORS B-CHEMICAL, 2014, 190 : 334 - 341
  • [7] VERSATILE MICROFLUIDIC GENERATION OF CHEMICAL CONCENTRATION GRADIENTS IN A FLOW-FREE SETTING
    Zhou, Y.
    Lin, Q.
    2011 IEEE 24TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS (MEMS), 2011, : 988 - 991
  • [8] Bacterial chemotaxis in chemical gradients created in a flow-free microfluidic device
    Nagy, K.
    Haja, O.
    Kerenyi, A.
    Valkai, S.
    Ormos, P.
    Galajda, P.
    EUROPEAN BIOPHYSICS JOURNAL WITH BIOPHYSICS LETTERS, 2013, 42 : S50 - S50
  • [9] Automated Microfluidic System for Dynamic Stimulation and Tracking of Single Cells
    Dettinger, Philip
    Frank, Tino
    Etzrodt, Martin
    Ahmed, Nouraiz
    Reimann, Andreas
    Trenzinger, Christoph
    Loeffler, Dirk
    Kokkaliaris, Konstantinos D.
    Schroeder, Timm
    Tay, Savas
    ANALYTICAL CHEMISTRY, 2018, 90 (18) : 10695 - 10700
  • [10] RAPID CAPTURE OF RARE CANCER CELLS USING A HIGH-PERFORMANCE MICROFLUIDIC CHIP
    Sheng, Weian
    Chen, Tao
    Tan, Weihong
    Fan, Hugh
    PROCEEDINGS OF THE ASME INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, 2013, VOL 10, 2014,