magnetic resonance (MR);
biological effects;
quality assurance;
safety;
technology;
radiology and radiologists;
D O I:
10.1148/radiol.2243010978
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Sound pressure levels (SPLs) during interventional magnetic resonance (MR) imaging may create an occupational hazard for the interventional radiologist (ie, the potential risk of hearing impairment). Therefore, A-weighted and linear continuous-equivalent SPLs were measured at the entrance of a 1.5-T MR imager during cardiovascular and real-time pulse sequences. The SPLs ranged from 81.5 to 99.3 dB (A-weighted scale), and frequencies were from I to 3 kHz. SPLs for the interventional radiologist exceeded a safe SPL of 80 dB (A-weighted scale) for all sequences; therefore, hearing protection is recommended. (C) RSNA, 2002.