Interventional MR imaging at 1.5 T: Quantification of sound exposure

被引:17
|
作者
Moelker, A
Maas, RAJJ
Lethimonnier, F
Pattynama, PMT
机构
[1] Erasmus Univ, Med Ctr Rotterdam, Dept Radiol, NL-3000 DR Rotterdam, Netherlands
[2] Erasmus Univ, Med Ctr Rotterdam, Dept Audiophys, NL-3000 DR Rotterdam, Netherlands
关键词
magnetic resonance (MR); biological effects; quality assurance; safety; technology; radiology and radiologists;
D O I
10.1148/radiol.2243010978
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Sound pressure levels (SPLs) during interventional magnetic resonance (MR) imaging may create an occupational hazard for the interventional radiologist (ie, the potential risk of hearing impairment). Therefore, A-weighted and linear continuous-equivalent SPLs were measured at the entrance of a 1.5-T MR imager during cardiovascular and real-time pulse sequences. The SPLs ranged from 81.5 to 99.3 dB (A-weighted scale), and frequencies were from I to 3 kHz. SPLs for the interventional radiologist exceeded a safe SPL of 80 dB (A-weighted scale) for all sequences; therefore, hearing protection is recommended. (C) RSNA, 2002.
引用
收藏
页码:889 / 895
页数:7
相关论文
共 50 条
  • [1] Hepatic iron overload:: Quantification with MR imaging at 1.5 T
    Ernst, O
    Rose, C
    Sergent, G
    L'Herminé, C
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 172 (04) : 1141 - 1142
  • [2] PROSTATIC DISORDERS - MR IMAGING AT 1.5 T
    PHILLIPS, ME
    KRESSEL, HY
    SPRITZER, CE
    ARGER, PH
    WEIN, AJ
    MARINELLI, D
    AXEL, L
    GEFTER, WB
    POLLACK, HM
    RADIOLOGY, 1987, 164 (02) : 386 - 392
  • [3] MR IMAGING OF THE LARYNX AT 1.5-T
    SAKAI, F
    GAMSU, G
    DILLON, WP
    LYNCH, DA
    GILBERT, TJ
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (01) : 60 - 71
  • [4] TESTES AND SCROTUM - MR IMAGING AT 1.5 T
    SEIDENWURM, D
    SMATHERS, RL
    LO, RK
    CARROL, CL
    BASSETT, J
    HOFFMAN, AR
    RADIOLOGY, 1987, 164 (02) : 393 - 398
  • [5] Quantification of low fat contents: a comparison of MR imaging and spectroscopy methods at 1.5 and 3 T
    Mansson, Sven
    Peterson, Pernilla
    Johansson, Edvin
    MAGNETIC RESONANCE IMAGING, 2012, 30 (10) : 1461 - 1467
  • [6] FEASIBILITY OF FAST MR IMAGING OF THE LIVER AT 1.5 T
    RAVAL, B
    MEHTA, S
    NARAYANA, P
    KULKARNI, M
    MAGNETIC RESONANCE IMAGING, 1989, 7 (02) : 203 - 210
  • [7] MR imaging of the chest: A practical approach at 1.5 T
    Puderbach, M.
    Hintze, C.
    Ley, S.
    Eichinger, M.
    Kauczor, H. -U.
    Biederer, J.
    EUROPEAN JOURNAL OF RADIOLOGY, 2007, 64 (03) : 345 - 355
  • [8] Interventional MR imaging
    Hagspiel, KD
    Kandarpa, K
    Jolesz, FA
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1997, 8 (05) : 745 - 758
  • [9] MR imaging of the knee at 0.2 T and 1.5 T: Correlation with arthroscopy
    Cotten, A
    Lapegue, F
    Boukhelifa, M
    Guiot, S
    Fossati, P
    Gougeon, F
    RADIOLOGY, 1998, 209P : 192 - 192
  • [10] 3.0-T MR Imaging of the Abdomen: Comparison with 1.5 T
    Chang, Kevin J.
    Kamel, Ihab R.
    Macura, Katarzyna J.
    Bluemke, David A.
    RADIOGRAPHICS, 2008, 28 (07) : 1983 - 1998