Assessment of IcareONE rebound tonometer for self-measuring intraocular pressure

被引:1
|
作者
Parodi, Maurizio Battaglia [1 ]
Iacono, Pierluigi [2 ]
Campa, Claudio [3 ]
La Spina, Carlo [1 ]
Triolo, Giacinto [1 ]
Lattanzio, Rosangela [1 ]
Bandello, Francesco [1 ]
机构
[1] Univ Vita Salute, San Raffaele Sci Inst, Dept Ophthalmol, Milan, Italy
[2] GB Bietti Fdn, IRCCS, I-00198 Rome, Italy
[3] Ferrara Univ Hosp, Arcispedale St Anna, Eye Clin, Ferrara, Italy
关键词
intraocular pressure; rebound tonometer; IcareONE; self-measurements;
D O I
10.1111/aos.12108
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Abstract. Purpose: To evaluate the precision of the IcareONE rebound tonometer, which was developed for self-measuring intraocular pressure (IOP) and to compare IcareONE measurement with Goldmann applanation tonometry (GAT). Methods: Twenty-four healthy eyes and 81 glaucomatous eyes were enrolled. IOP measurements (three times per session) with IcareONE were made in a random order by an ophthalmologist (Icare(O)) and by the subject (Icare(S)). Intraclass correlation coefficients (CCs), kappa values and mean values of IOP were compared among the two types of Icare recordings and GAT. Bland-Altman analysis was used to assess agreement between methods. Multiple regression analysis was performed to identify the subject factors that influenced the discordant measurements between IcareONE and GAT. Results: The mean value of Icare(O) and Icare(S) measurements was 13.5 +/- 5.2 and 13.5 +/- 5.4 mmHg, respectively, neither of which was significantly different from GAT (13.8 +/- 4.4). The intrarater CC of Icare(O) and Icare(S) was 0.968 and 0.885, respectively. The intermethod CC and weighted kappa between Icare(O) and Icare(S) were 0.907 and 0.684, respectively. All pairwise correlations between the two types of IOP measurement showed coefficients of determination >0.8. Bland-Altman analysis did not show any proportional biases. Multiple regression analysis revealed that the differences between GAT and Icare(O) or Icare(S) were positively correlated with central corneal thickness (CCT) and negatively correlated with age. Conclusions: Intraocular pressure measurements with IcareONE by a physician and by the subject showed excellent agreement with GAT measurements; IcareONE measurements between a physician and the subject had high intrarater reliability, and good agreement thicker CCT led IcareONE measurement to overestimate IOP, while higher age caused it to underestimate IOP compared with GAT.
引用
收藏
页码:E243 / E244
页数:8
相关论文
共 50 条
  • [1] Evaluation of a new transpalpebral tonometer for self-measuring intraocular pressure
    Kobashi, Hidenaga
    PLOS ONE, 2024, 19 (05):
  • [2] Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects
    Halkiadakis, Ioannis
    Stratos, Aimilianos
    Stergiopoulos, George
    Patsea, Eleni
    Skouriotis, Sotiris
    Mitropoulos, Panagiotis
    Papaconstantinou, Dimitrios
    Georgopoulos, Gerasimos
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 250 (08) : 1207 - 1211
  • [3] Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects
    Ioannis Halkiadakis
    Aimilianos Stratos
    George Stergiopoulos
    Eleni Patsea
    Sotiris Skouriotis
    Panagiotis Mitropoulos
    Dimitrios Papaconstantinou
    Gerasimos Georgopoulos
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2012, 250 : 1207 - 1211
  • [4] Evaluation of a rebound tonometer for measuring intraocular pressure in dogs and horses
    Knollinger, AM
    La Croix, NC
    Barrett, PM
    Miller, PE
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2005, 227 (02): : 244 - 248
  • [5] Comparison of a rebound and an applanation tonometer for measuring intraocular pressure in normal rabbits
    Pereira, Fabiana Quartiero
    Bercht, Bernardo Stefano
    Soares, Mirian Gomes
    Bittencourt da Mota, Mauricio Grillo
    Tadeu Pigatto, Joao Antonio
    VETERINARY OPHTHALMOLOGY, 2011, 14 (05) : 321 - 326
  • [6] Evaluation of Rebound Tonometer for Measuring Intraocular Pressure at Deviated Angle and Position
    Takenaka, Joji
    Mochizuki, Hideki
    Kunihara, Eriko
    Tanaka, Junko
    Kiuchi, Yoshiaki
    CURRENT EYE RESEARCH, 2011, 36 (05) : 422 - 428
  • [7] Evaluation of a new rebound tonometer for self-measurement of intraocular pressure
    Dabasia, Priya L.
    Lawrenson, John G.
    Murdoch, Ian E.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (08) : 1139 - 1143
  • [8] Rebound Tonometer: Ideal Tonometer for Measurement of Accurate Intraocular Pressure
    Suman, Suwarna
    Agrawal, Ajai
    Pal, Virendra K.
    Pratap, Vir B.
    JOURNAL OF GLAUCOMA, 2014, 23 (09) : 633 - 637
  • [9] Evaluation of monkey intraocular pressure by rebound tonometer
    Yu, Wenhan
    Cao, Guiqun
    Qiu, Jinghua
    Liu, Xuyang
    Ma, Jia
    Li, Ni
    Yu, Man
    Yan, Naihong
    Chen, Lei
    Pang, Iok-Hou
    MOLECULAR VISION, 2009, 15 (235-36): : 2196 - 2201
  • [10] Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers
    Nakamura, Makoto
    Darhad, Urtogtah
    Tatsumi, Yasuko
    Fujioka, Miyuki
    Kusuhara, Azusa
    Maeda, Hidetaka
    Negi, Akira
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (02) : 332 - 334